FLUKE 1673 FC Multifunction Tester

Product Specifications
- Accessories…………………………………………….. IEC 61010-031
- TP165X Remote Probe with cap ……….. CAT IV 600 V, CAT III 1000 V, 10 A
- TP165X Remote Probe without cap….. CAT II 1000 V, 10 A
- TL-L1, TL-L2, TL-L3 Test Leads …………. CAT IV 600 V, CAT III 1000 V, 10 A
- Test Probes with cap…………………………… CAT IV 600 V, CAT III 1000 V, 10 A
- Test Probes without cap …………………….. CAT II 1000 V, 10 A
- AC285 Alligator Clip ……………………………. CAT IV 600 V, CAT III 1000 V, 10 A
- Country-Specific Mains Cord…………….. CAT II 250 V, 1000 V dc
- Ingress Protection………………………………….. IEC 60529: IP40
Electromagnetic Compatibility (EMC)
Some mobile devices (for example, handheld transceivers) that transmit RF energy can transmit levels that exceed 3 V/m and can damage sensitive electronic circuits. To ensure the best performance, do not allow a device that transmits RF energy in >3 V/m to be within 30 cm of the Product while in use.
International ……………………………………………. IEC 61326-1: Portable CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object.
Wireless Radio with Adapter
Frequency Range……………………………………. 2400 MHz to 2483.5 MHz
Output Power ………………………………………….. <100 mW
SIMPLIFIED EU DECLARATION OF CONFORMITY
Hereby, Fluke declares that the radio equipment contained in this Product is in compliance with Directive 2014/53/EU. The full text of the EU declaration is available at the following Internet address: www.fluke.com/red
Electrical Measurement Specifications
The specification for accuracy is defined as ±(% reading +digit counts) at 23 °C ±5 °C, ≤80 % RH. Between 0 °C and 18 °C and between 28 °C and 40 °C, accuracy specifications may degrade by 0.1 x (accuracy specification) per °C. The specifications apply up to one year from the date of calibration.
Note
Accuracy indicates the intrinsic uncertainty according to IEC 61557.
Voltage Measurement (V)
| Range | Resolution | Accuracy 45 Hz to 66 Hz | Input Impedance | Overload Protection |
| 600 V | 0.1 V | 0.8 % + 3 | 320 kW | 660 V rms |
| Note: The display shows if dc voltage is detected. | ||||
Insulation Resistance Measurement (RISO)
| Test Voltage | Accuracy | |
| Model 1672 | Model 1673 FC/1674 FC | |
| 100-250-500-1000 V | 50-100-250-500-1000 V | +10 %, -0 % |
| Test Voltage | Insulation Resistance Range | Resolution | Test Current | Accuracy |
| 50 V | 10 kW to 50 MW | 0.01 MW | 1 mA @ 50 kW | ±(3 % + 3 digits) |
| 100 V | 10 kW to 20 MW | 0.01 MW | 1 mA @ 100 kW | ±(3 % + 3 digits) |
| 20 MW to 100 MW | 0.1 MW | ±(3 % + 3 digits) | ||
| 250 V | 10 kW to 20 MW | 0.01 MW | 1 mA @ 250 kW | ±(1.5 % + 3 digits) |
| 20 MW to 200 MW | 0.1 MW | ±(1.5 % + 3 digits) | ||
|
500 V |
10 kW to 20 MW | 0.01 MW |
1 mA @ 500 kW |
±(1.5 % + 3 digits) |
| 20 MW to 200 MW | 0.1 MW | ±(1.5 % + 3 digits) | ||
| 200 MW to 500 MW | 1 MW | ±10 % | ||
| 1000 V | 100 kW to 200 MW | 0.1 MW | 1 mA @ 1 MW | ±(1.5 % + 3 digits) |
| 200 MW to 1000 MW | 1 MW | ±10 % | ||
| Note: The number of insulation tests with full-charge battery is >2000. | ||||
| Auto Discharge | Discharge time constant <0.5 second for C = 1 mF or less. |
| Live Circuit Detection | Inhibits test if terminal voltage >30 V prior to initiation of test. |
| Maximum Capacitive Load | Operable with up to 5 mF load. |
| Insulation Safety Pretest (1674 FC) | Connections from the Tester to L, N, and PE are required. |
SPD (Surge Protection Device) Insulation RAMP Test (Varistor Test) IEC 61643-11
| Test Voltage | Voltage Range | Resolution | Test Current | Accuracy |
| 500 V | Step Ramp 0 V to 500 V | 1 V | 1 mA | ±(1.5 % + 3 digits) |
| 1000 V | Step Ramp 0 V to 1000 V | 1 V | 1 mA | ±(1.5 % + 3 digits) |
Continuity Test (RLO)
| Range (Autoranging) | Resolution | Open Circuit Voltage | Accuracy |
| 20 W | 0.01 W | >4 V | ±(1.5 % + 3 digits)[1] |
| 200 W | 0.1 W | >4 V | ±(3 % + 3 digits) |
| 2000 W | 1 W | >4 V | ±(3 % + 3 digits) |
| Note: The number of 250 mA @ 1 W continuity tests with full-charge battery is >1500. [1] For 10 mA, add 3 digits. | |||
| Range Setting | Display Range | Test Current[1] |
|
250 mA |
0.2 W to 2.0 W | 250 mA |
| 2 W to 160 W | 250 mA to 50 mA | |
| 160 W to 800 W | 10 mA | |
| 800 W to 2000 W | 2 mA | |
| 10 mA | 0 W to 800 W | 10 mA |
| 800 W to 2000 W | 2 mA | |
| [1] All test currents ±10 %. | ||
| Test Probe Zeroing | Select ZERO to zero the test probe.
Can subtract up to 3 W of lead resistance. Error message for >3 W. |
| Live Circuit Detection | Inhibits test if terminal voltage >10 V ac detected before initiation of test. Does not apply in point-to-point continuous mode. |
Mains Wiring Indicator
Icons indicate if L-PE or L-N terminals are reversed.

Loop and RCD tests are inhibited and an error message is generated if the input voltage is not between 100 V and 500 V. The UK Loop and RCD tests are inhibited if the L-PE or the L-N terminals are reversed.
Loop and Line Impedance (ZI No Trip and Hi Current)
| Mains Input Voltage Range | 50 V ac to 600 V ac (45 Hz to 66 Hz) |
| Input Connection (softkey selection) | Loop Impedance: phase-to-earth |
| Line Impedance: phase-to-neutral | |
| Limit on Consecutive Tests | Auto-shutdown when the temperature of internal components is too high. |
| Maximum Test Current @ 600 V | 30 A sinusoidal for 10 ms |
| Maximum Test Current @ 400 V | 20 A sinusoidal for 10 ms |
| Maximum Test Current @ 230 V | 12 A sinusoidal for 10 ms |
| Range Setting | Resolution | Accuracy[1] |
| 10 Ω [2] | 0.001 Ω | Hi Current mΩ mode: ±(2 % + 35 digits) |
| 20 Ω | 0.01 Ω | No Trip mode: ±(3 % + 6 digits) |
| Hi Current mode: ±(2 % + 4 digits) | ||
| 200 Ω | 0.1 Ω | No Trip mode: ±(3 %) |
| Hi Current mode: ±(2 %) | ||
| 2000 Ω | 1 Ω | ±6 %[3] |
| [1] Valid for resistance of neutral circuit <20 Ω and up to a system phase angle of 30 °. Test leads must be zeroed before testing. [2] 1674 FC model only. [3] Valid for mains voltage >200 V. | ||
Voltage Drop (ΔV)
| Measurement Range | Resolution | Accuracy |
| 0.0 % to 99.9 % | 0.1 % | Consider accuracy of line impedance measurements. |
Prospective Earth Fault Current (PEFC) Prospective Short Circuit Current (PSC)
|
Computation |
Prospective Earth Fault Current (PEFC/IK) or Prospective Short Circuit Current (PSC/IK) determined by dividing measured mains voltage by measured loop (L-PE) resistance or line (L-N) resistance, respectively. | |
| Range | 0 kA to 50 kA | |
|
Resolution |
Range | Resolution |
| IK <1000 A | 1 A | |
| IK ³1000 A | 0.1 kA | |
| Accuracy | Determined by accuracy of loop resistance and mains voltage measurements. | |
RCD Tests
RCD Types Tested
Limit on consecutive tests: Automatic shutdown for RCD tests when the temperature of internal components is too high.
| RCD Type[1] | 1672 | 1673 FC | 1674 FC | |
| AC[2] | G[3] | |||
| AC | S[4] | |||
| A[5], F[6] | G | |||
| A, F | S | |||
| B, B+[7] | G | |||
| B, B+ | S | |||
| RDC-DD, A/EV, B/Mi[8] | G | |||
| GFCI | G | |||
| [1] RCD test inhibited for V >265 ac RCD tests permitted only if the selected current, multiplied by earthing resistance, is <50 V [2] AC – Responds to ac [3] G – General, no delay [4] S – Time delay [5] A – Responds to ac and pulsed signal [6] F – responds to ac, pulsed, and high frequency [7] B, B+ – Responds to ac, pulsed, high frequency, and smooth dc [8] RDC-DD, A/EV, B/Mi – Responds to dc residual currents >6 mA |
||||
Test Signals
| Type | Test Signal Description |
|
RCD type AC (sinusoidal) |
The waveform is a sinewave starting at zero crossing, polarity determined by phase selection (0 ° phase starts with low to high zero crossing, 180 ° phase starts with high to low zero crossing). The magnitude of the test current is IΔN x Multiplier for all tests. |
| A
(half-wave) |
The waveform is a half wave rectified sinewave starting at zero, polarity determined by phase selection (0 ° phase starts with low to high zero crossing, 180 ° phase starts with high to low zero crossing). The magnitude of the test current is 0.7 x IΔN (rms) x Multiplier for all tests where the multiplier is x0.5 (x1/2). The magnitude of the test current is 2.0 x IΔN (rms) x Multiplier for all tests where both the multiplier is ≥ x1 and IΔN = 0.01A. The magnitude of the test current is 1.4 x IΔN (rms) x Multiplier for all tests for all other settings. |
| B (DC) | This is a smooth dc current in accordance to EN61557-6 Annex A |
RCD Tripping Indicator
The RCD symbol switches on as a good test indicator when testing the RCD trip time or RCD trip current if the trip time meets these conditions:
| RCD Type | IΔN | Trip Time Limits |
| G | x1 | <300 ms |
| S | x1 | >130 ms and <500 ms |
| G | x5 | <40 ms |
| S | x5 | >50 ms and <150 ms |
RCD Tripping Time (ΔT)
| Test Function | RCD Current Selection | |||||||
| 6 mA | 10 mA | 30 mA | 100 mA[1] | 300 mA[1] | 500 mA[1] | 1000 mA[2] | Var[3] | |
| x ½, 1 | ||||||||
| x 5 | ||||||||
| Ramp | ||||||||
| Auto | ||||||||
| Mains voltage 100 V – 265 V ac, 45/66 Hz [1] Type B RCDs require mains voltage range of 195 V to 265 V. [2] Type AC RCDs only. [3] Type A RCDs are limited to 700 mA, not available for Type B RCDs. | ||||||||
| Current Multiplier | RCD Type | Measurement Range | Accuracy | |
| Europe | UK | |||
| x ½ | G | 310 ms | 2000 ms | ±(1 % Reading + 1 ms) |
| x ½ | S | 510 ms | 2000 ms | ±(1 % Reading + 1 ms) |
| x 1 | G | 310 ms | 310 ms | ±(1 % Reading + 1 ms) |
| x 1 | S | 510 ms | 510 ms | ±(1 % Reading + 1 ms) |
| x 5 | G | 50 ms | 50 ms | ±(1 % Reading + 1 ms) |
| x 5 | S | 160 ms | 160 ms | ±(1 % Reading + 1 ms) |
| x 1 | RDC-DD, A/EV, B/Mi | 10 s | 10 s | ±(1 % Reading + 1 ms) |
| 350 mA | A | 10 ms | 10 ms | ±(1 % Reading + 1 ms) |
| [1] G – General, no delay / S – Time delay | ||||
RCD Tripping Current (IΔN) Measurement/Ramp Test
| Current Range | Step Size | Dwell Time | Measurement Accuracy | |
| Type G | Type S | |||
| 30 % to 110 % of
RCD-rated current[1] |
10 % of IDN[2] | 300 ms/step | 500 ms/step | ±5 % |
| [1] Specified trip current ranges (IEC 61008-1): 30 % to 150 % for Type A IΔ N >10 mA
30 % to 210 % for Type A IΔ N = 10 mA 20 % to 210 % for Type B 50 % to 100 % for Type AC 35 % to 140 % for Type A (>10 mA) 35 % to 200 % for Type A (≤10 mA) 50 % to 200 % for Type B [2] 5 % for Type B |
||||
Insulation Monitoring Devices (IMD) IEC 61557-8
| Range | Resolution | Note |
| 1 kΩ to 10 kΩ | 1 kΩ |
>1 MΩ only available with voltages >100 V |
| 10 kΩ to 100 kΩ | 10 kΩ | |
| 100 kΩ to 3 MΩ | 100 kΩ |
| Maximum Time[1] | Resolution | Note |
| 600.0 s | 0.1 s | Manual timing with TEST button Test |
| [1] The maximum test time duration may be reduced due to heat development when low resistance values are used in combination with high voltages. | ||
Phase Sequence Test
| Icon | |
|
Display of Phase Sequence |
Displays “1-2-3” in digital display field for correct sequence. Displays “3-2-1” for incorrect phase. Dashes in place of a number indicate a valid determination could not be made. |
| Mains Input Voltage Range (phase-to-phase) | 185 V to 600 V |
Earth Resistance Test (RE)
1673 FC and 1674 FC only
| Range | Resolution | Accuracy |
| 200 W | 0.1 W | ±(2 % + 5 digits) |
| 2000 W | 1 W | ±(3.5 % + 10 digits) |
| [1]
Range: RE + RPROBE |
Test Current |
| 2200 W | 3.5 mA |
| 16 000 W | 500 mA |
| 52 000 W | 150 mA |
| [1] Without external voltages. | |
| Freqency | Output Voltage |
| 128 Hz | 25 V |
| Live Circuit Detection | Inhibits test if terminal voltage >10 V ac is detected before the start of the test. |
Auto Test Sequence
Models 1673 FC and 1674 FC only.
Meets the specifications of the individual tests.
Operating Ranges and Uncertainties (IEC 61557)
The maximum permissible error shows as a percent of reading plus a number of digits when all uncertainty influence factors E1-E10 are counted.
|
Function |
Display Range |
IEC 61557
Measurement Range Operating Uncertainty |
Nominal Values |
| V | 0.0 V ac to 600 V ac | 50 V ac to 600 V ac
±(2 % + 2 digits) |
UN = 230/400 V ac
f = 50/60 Hz |
| RLO
IEC 61557-4 |
0.00 W to 2000 W | 0.2 W to 2000 W
±(10 % + 2 digits) |
4.0 V dc <UQ <24 V dc RLO ≤2.00 W IN ≥200 mA |
|
RISO IEC 61557-2 |
0.00 MW to 1000 MW |
1 MW to 200 MW
±(10 % + 2 digits) 200 MW to 1000 MW ±(15 % + 2 digits) |
UN = 50 / 100 / 250 / 500 /
1000 V dc IN = 1.0 mA |
|
ZI IEC 61557-3 |
ZI (No Trip)
0.00 W to 2000 W |
0.4 W to 2000 W
±(15 % + 6 digits) |
UN = 230/400 V ac f = 50/60 Hz IK= 0 A to 10.0 kA |
| ZI (Hi Current)
0.00 W to 2000 W |
0.2 W to 200 W
±(10 % + 4 digits) |
||
| ZI (Hi Current, Hi Res) 0 mW to 9999 mW | 100 mW to 9999 mW
±(8 % + 20 digits) |
||
| RE
0.00 W to 2000 W |
10 W to 1000 W
±(10 % + 2 digits) |
||
|
DT, IDN IEC 61557-6 |
DT 0.0 ms to 2000 ms | 25 ms to 2000 ms
±(10 % + 1 digits) |
DT @ 10 / 30 / 100 / 300 /
500 / 1000 / Var mA |
| IΔN 3 mA to 550 mA
(Var 3 mA to 700 mA) |
3 mA to 550 mA
±(10 % + 1 digits) |
IΔN = 10 / 30 / 100 / 300 /
500 / 1000 / Var mA |
|
| RE
IEC 61557-5 |
0.0 W to 2000 W | 10 W to 2000 W
±(10 % + 2 digits) |
f = 128 Hz |
| Phase
IEC 61557-7 |
— | — | 1 : 2 : 3 |
| IMD
IEC 61557-8 |
0 s to 600 s | — | user-defined manual timer |
Operating Uncertainties (IEC 61557)
The Operating Uncertainty shows the maximum permissible uncertainty as a percent of reading when all uncertainty influence factors E1-E10 are counted.
| Volts | RLO[1] | RISO[2] | ZI[3] | DT[4] | IDN[4] | RE[5] | |
| Intrinsic Uncertainty A | 0.80 % | 1.50 % | 10.00 % | 6.00 % | 1.00 % | 5.00 % | 3.50 % |
| [1] IEC 61557-4 [4] IEC 61557-6 [2] IEC 61557-2 [5] IEC 61557-5 [3] IEC 61557-3 | |||||||
| Influence Quantity | Volts | RLO[1] | R [2] ISO | ZI[3] | DT[4] | IDN[4] | RE[5] |
| E1
Position |
0.00 % | 0.00 % | 0.00 % | 0.00 % | 0.00 % | 0.00 % | 0.00 % |
| E2
Supply voltage |
0.50 % | 3.00 % | 3.00 % | 3.00 % | 3.00 % | 2.75 % | 2.00 % |
| E3
Temperature |
0.50 % | 3.00 % | 3.00 % | 3.00 % | 3.00 % | 2.25 % | 1.50 % |
| E4
Series interferences voltage |
— |
— |
— |
— |
— |
— |
2.00 % |
| E5
Resistance of the probes and auxiliary earth electrodes |
— |
— |
— |
— |
— |
— |
4.60 % |
| E6.2
System phase angle |
— | — | — | 1.00 % | — | — | — |
| E7
System frequency |
0.50 % | — | — | 2.50 % | — | — | 0.00 % |
| E8
System voltage |
— | — | — | 2.50 % | 2.50 % | 2.50 % | 0.00 % |
| E9
Harmonics |
— | — | — | 2.50 % | — | — | — |
| E10
D.C. quantity |
— | — | — | 2.50 % | — | — | — |
| [1] IEC 61557-4 [4] IEC 61557-6 [2] IEC 61557-2 [5] IEC 61557-5 [3] IEC 61557-3 | |||||||
Maximum Display Values
The following tables can be used for the determination of maximum or minimum display values considering maximum instrument operating uncertainty per EN61557-1.
Insulation Resistance (RISO)
| 50 V | 100 V | 250 V | 500 V | 1000 V | |||||
| Limit Value | Maximum Display Value | Limit Value | Maximum Display Value | Limit Value | Maximum Display Value | Limit Value | Maximum Display Value | Limit Value | Maximum Display Value |
| 1 | 1.12 | 1 | 1.12 | 1 | 1.12 | 1 | 1.12 | 1 | 1.3 |
| 2 | 2.22 | 2 | 2.22 | 2 | 2.22 | 2 | 2.22 | 2 | 2.4 |
| 3 | 3.32 | 3 | 3.32 | 3 | 3.32 | 3 | 3.32 | 3 | 3.5 |
| 4 | 4.42 | 4 | 4.42 | 4 | 4.42 | 4 | 4.42 | 4 | 4.6 |
| 5 | 5.52 | 5 | 5.52 | 5 | 5.52 | 5 | 5.52 | 5 | 5.7 |
| 6 | 6.62 | 6 | 6.62 | 6 | 6.62 | 6 | 6.62 | 6 | 6.8 |
| 7 | 7.72 | 7 | 7.72 | 7 | 7.72 | 7 | 7.72 | 7 | 7.9 |
| 8 | 8.82 | 8 | 8.82 | 8 | 8.82 | 8 | 8.82 | 8 | 9.0 |
| 9 | 9.92 | 9 | 9.92 | 9 | 9.92 | 9 | 9.92 | 9 | 10.1 |
| 10 | 11.02 | 10 | 11.02 | 10 | 11.02 | 10 | 11.02 | 10 | 11.2 |
| 20 | 22.02 | 20 | 22.2 | 20 | 22.2 | 20 | 22.2 | 20 | 22.2 |
| 30 | 33.02 | 30 | 33.2 | 30 | 33.2 | 30 | 33.2 | 30 | 33.2 |
| 40 | 44.02 | 40 | 44.2 | 40 | 44.2 | 40 | 44.2 | 40 | 44.2 |
| 45 | 49.52 | 50 | 55.2 | 50 | 55.2 | 50 | 55.2 | 50 | 55.2 |
| 60 | 66.2 | 60 | 66.2 | 60 | 66.2 | 60 | 66.2 | ||
| 70 | 77.2 | 70 | 77.2 | 70 | 77.2 | 70 | 77.2 | ||
| 80 | 88.2 | 80 | 88.2 | 80 | 88.2 | 80 | 88.2 | ||
| 90 | 99.2 | 90 | 99.2 | 90 | 99.2 | 90 | 99.2 | ||
| 100 | 110.2 | 100 | 110.2 | 100 | 110.2 | ||||
| 180 | 198.2 | 200 | 232 | 200 | 232 | ||||
| 300 | 347 | 300 | 347 | ||||||
| 400 | 462 | 400 | 462 | ||||||
| 430 | 497 | 500 | 577 | ||||||
| 600 | 692 | ||||||||
| 700 | 807 | ||||||||
| 800 | 922 | ||||||||
| 850 | 980 | ||||||||
Continuity (RLO)
| Limit Value | Maximum Display Value |
| 0.2 | 0.16 |
| 0.3 | 0.25 |
| 0.4 | 0.34 |
| 0.5 | 0.43 |
| 0.6 | 0.52 |
| 0.7 | 0.61 |
| 0.8 | 0.70 |
| 0.9 | 0.79 |
| 1 | 0.88 |
| 2 | 1.78 |
| Limit Value | Maximum Display Value |
| 3 | 2.68 |
| 4 | 3.58 |
| 5 | 4.48 |
| 6 | 5.38 |
| 7 | 6.28 |
| 8 | 7.18 |
| 9 | 8.08 |
| 10 | 8.98 |
| 20 | 17.98 |
| 30 | 26.8 |
Loop Tests (ZI)
| Loop ZI Hi Current | Loop ZI No Trip | Loop ZI Hi Res | |||
| Limit Value | Maximum Display Value (Ω) | Limit Value | Maximum Display Value (Ω) | Limit Value | Maximum Display Value (mΩ) |
| 0.20 | 0.14 | – | – | 0.20 | 180 |
| 0.30 | 0.23 | – | – | 0.30 | 280 |
| 0.40 | 0.32 | 0.40 | 0.28 | 0.40 | 370 |
| 0.50 | 0.41 | 0.50 | 0.37 | 0.50 | 460 |
| 0.60 | 0.50 | 0.60 | 0.45 | 0.60 | 550 |
| 0.70 | 0.59 | 0.70 | 0.54 | 0.70 | 640 |
| 0.80 | 0.68 | 0.80 | 0.62 | 0.80 | 740 |
| 0.90 | 0.77 | 0.90 | 0.71 | 0.90 | 830 |
| 1.00 | 0.86 | 1.00 | 0.79 | 1.00 | 920 |
| 1.10 | 0.95 | 1.10 | 0.88 | 1.10 | 1010 |
| 1.20 | 1.04 | 1.20 | 0.96 | 1.20 | 1100 |
| 1.30 | 1.13 | 1.30 | 1.05 | 1.30 | 1200 |
| 1.40 | 1.22 | 1.40 | 1.13 | 1.40 | 1290 |
| 1.50 | 1.31 | 1.50 | 1.22 | 1.50 | 1380 |
| 1.60 | 1.40 | 1.60 | 1.30 | 1.60 | 1470 |
| 1.70 | 1.49 | 1.70 | 1.39 | 1.70 | 1560 |
| 1.80 | 1.58 | 1.80 | 1.47 | 1.80 | 1660 |
| 1.90 | 1.67 | 1.90 | 1.56 | 1.90 | 1750 |
| 2.00 | 1.76 | 2.00 | 1.64 | 2.00 | 1840 |
| 3.00 | 2.66 | 3.00 | 2.49 | 3.00 | 2760 |
| 4.00 | 3.56 | 4.00 | 3.34 | 4.00 | 3680 |
| 5.00 | 4.46 | 5.00 | 4.19 | 5.00 | 4600 |
| 10.00 | 8.96 | 10.00 | 8.44 | 10.00 | 9200 |
RCD/FI (ΔT, IΔN)
| RCD/FI Time | RCD/FI Current | ||
| Limit Value | Maximum Display Value | Limit Value | Maximum Display Value |
| 25 | 22.4 | 0.5 | 0.49 |
| 30 | 26.9 | 0.6 | 0.49 |
| 40 | 35.9 | 0.7 | 0.59 |
| 50 | 44.9 | 0.8 | 0.69 |
| 60 | 53.9 | 0.9 | 0.79 |
| 70 | 62.9 | 1 | 0.89 |
| 80 | 71.9 | 2 | 1.79 |
| 90 | 80.9 | 3 | 2.69 |
| 100 | 89.9 | 4 | 3.59 |
| 200 | 179.9 | 5 | 4.49 |
| 300 | 269 | 6 | 5.39 |
| 400 | 359 | 7 | 6.29 |
| 500 | 449 | 8 | 7.19 |
| 600 | 539 | 9 | 8.09 |
| 700 | 629 | 10 | 8.99 |
| 800 | 719 | 20 | 17.99 |
| 900 | 809 | 30 | 26.9 |
| 1000 | 899 | 40 | 35.9 |
| 2000 | 1799 | 50 | 44.9 |
| 60 | 53.9 | ||
| 70 | 62.9 | ||
| 80 | 71.9 | ||
| 90 | 80.9 | ||
| 100 | 89.9 | ||
| 200 | 179.9 | ||
| 300 | 269.9 | ||
| 400 | 359.9 | ||
| 500 | 449.9 | ||
Earth Tests (RE)
| Limit Value | Maximum Display Value |
| 10 | 8.8 |
| 20 | 17.8 |
| 30 | 26.8 |
| 40 | 35.8 |
| 50 | 44.8 |
| 60 | 53.8 |
| 70 | 62.8 |
| 80 | 71.8 |
| 90 | 80.8 |
| 100 | 89.8 |
| Limit Value | Maximum Display Value |
| 200 | 179.8 |
| 300 | 268 |
| 400 | 358 |
| 500 | 448 |
| 600 | 538 |
| 700 | 628 |
| 800 | 718 |
| 900 | 808 |
| 1000 | 898 |
| 2000 | 1798 |