CSPRING Mk2 Automatic CPAP Instruction Manual

CSPRING logo

CSpring MK2

Instruction for:
ResMed Airsense 10
ResMed Aircurve 10
ResMed Airsense 11

CSPRING Mk2 Automatic CPAP

PARTS LIST

CSPRING Mk2 Automatic CPAP - a1

  1. Lid
  2. Water funnel
  3. Pump hose bend support
  4. Anti-siphon & primer pinhole valve / sponge
  5. Water pump
  6. Water level
  7. Touch on/off button
  8. Water supply hose
  9. Sensor cable
  10. Sensor holder clip x 2
  11. Power adapter
  12. Water tank
  13. Pump hose
1 Replacing the Water Chamber / Installing the sensors

CSPRING - Important 1 Most CSpring devices come with a replacement water chamber and pre-installed sensors.

CSPRING Mk2 Automatic CPAP - a2

  1. 1- Replace your CPAP water chamber with the one included with your CSpring one®.
    (The sensor harness should come attached to the water chamber clips)
    2- Align the the sensor harness plug into the socket pins, Insert and tighten until it is secure.
  2. Clip: Back / Front
  3. Clip holding sensor
  4. Anti-spill Sensor (52)
  5. Water level sensor (S1)
  6. Back

Airsense & Aircurve 10

CSPRING Mk2 Automatic CPAP - b1    CSPRING Mk2 Automatic CPAP - b2

CSPRING Mk2 Automatic CPAP - b3 CSPRING Mk2 Automatic CPAP - b4 CSPRING Mk2 Automatic CPAP - b5

Airsense 11

CSPRING Mk2 Automatic CPAP - b6 CSPRING Mk2 Automatic CPAP - b7

CSPRING Mk2 Automatic CPAP - b8 CSPRING Mk2 Automatic CPAP - b9

2 Connecting water supply hose

CSPRING Mk2 Automatic CPAP - c1

  1. 1• Connect each end of the water supply hose into the CSpring One® connection valve and the CPAP water chamber’s adapter top valve.
    2• Press the panel connect valve button, to connect hose or release.
  2. Release button
  3. Panel connect valve / button
  4. Water supply hose
  5. Male hose valve
  6. Front
3 How to use

Follow these steps after the CPAP water chamber has been replaced with the one included with your CSpring Mk2.

CSPRING Mk2 Automatic CPAP - d1 CSPRING Mk2 Automatic CPAP - d2 CSPRING Mk2 Automatic CPAP - d3

1- Take the lid off, initially add 1L of distilled water.
The optimal location for testing is next to a sink on a firm surface.

2- Press the on/off button once to power on the device. The ambiance light will become visible. Tap the on/off button a second time to turn off the ambiance light. The device will remain on, with only the UVC light active. The water pump will start automatically.

3- After a few moments, water will begin flowing through the hose into the CPAP water chamber. The water will stop, once it reaches the first sensor.

4- Confirm that the CSpring Mk2® is filling the CPAP water chamber and stopping promptly, once the water reaches the first sensor.

5- SUCCESS! It’s now ready for use.

CSPRING Mk2 Automatic CPAP - d4

  1. Anti-spill Sensor (S2)
  2. Water level sensor (S1)
  3. Tap switch on/off button
4 Internal pump placement

CSPRING Mk2 Automatic CPAP - e1

  1. Bend support part slides into place
  2. **Primer pinhole valve.
    Water is ejected during operation.
  3. Place the rubber feet of the pump on the tank floor and press tightly, ensuring they are positioned anywhere but the center hole.
CSPRING - Important 1 Common Troubleshooting

If you are having problems with your CSpring One, see if one of the solutions below can resolve the problem.

Problem Solution
CSpring is not powering on. Make sure that the power adapter is properly connected, both to wall electric receptacle and into the CSpring power adapter input.
Water is not coming out from the hose into the water chamber Verify that both ends of the hose are properly connected into the Cspring and the water chamber intake valve.
Make sure that the hose is not bent or kinked.
Water is coming out very slowly Clean the waterpump filter, verify that the hose is not bent or kinked,
Water spraying our from the hose This is normal. there is a valve with a pin hole that allows some water out from the hose to provide proper hose priming.  
Noise when first powering on When the CSpring is turned on for the first time, the pump will make a noise while the hose gets primed with water and gets rid of any air in the line.

If you are still having trouble with your CSpring Device, please contact out support team: Email: support@cspring.us

CSPRING logo

Documents / Resources

PDF thumbnailMk2 Automatic CPAP
Instruction Manual · Mk2 Automatic CPAP, Mk2, Automatic CPAP, CPAP

References

Questions and answers

Helpful product questions tied to this manual. Each item links to the full Q&A page.

Questions & Answers

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

  1. After every use do I turn off the MK2 or leave it on?

    Royal Abbit asks After every use do I turn off the MK2 or leave it on?
    Manuals+ clarified question
    Question

    Should I turn off the CSpring Mk2 after every use or leave it on?

    Answer

    Based on the CSpring Mk2 manual, the device is designed to automate the refilling process. While the manual does not explicitly state you must turn it off after every use, it provides the following information regarding the power button states:

    • Powering On: Press the on/off button once to power on the device. This activates the ambiance light and the automatic water pump.
    • Standby/UVC Mode: Tapping the on/off button a second time turns off the ambiance light, but the device remains on with the UVC light active to maintain water hygiene.

    Because the CSpring Mk2 uses sensors to monitor your CPAP water level and a UVC light to keep the water clean, many users leave the device on (in the second mode with the ambiance light off) so it can continue to monitor and sanitize the water. If you prefer to turn it off completely, you can use the touch on/off button, but you will need to turn it back on before your next sleep session to ensure your CPAP chamber is refilled.