Cryopush logoAO2-P-001 Cryopush Cold Compression DeviceUser Manual
Cryopush Cold Compression Device
Model No.:AO2-P-001
Version: A/3

Intended Use

The Cr yopush Cold Compression Device is indicated for the temporary relief of minor muscle aches and pains.
The device is indicated for temporar y increase in circulation of the treated areas in people who are in good health, and simulates kneading and stroking of tissues using by an inflatable wrap.
The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessar y or desirable.

Contraindications

Do NOT use the device if any of the below condition applies:

  • Suspected, active or untreated: deep vein thrombosis, ischemic vascular disease, severe arteriosclerosis, pulmonaryedema, congestive heart failure, thrombophlebitis or an active infection;
  • On a limb where wraps would inter fere with the following conditions: vein ligation, gangrene, dermatitis, open wounds, a recent skin graft, massive edema or extreme deformity of the af fected area;
  • Those with Neuropathy;
  • Those who are extremely insensitive to pain;
  • Where increased circulation is undesirable;
  • Those who have significant vascular impairment in the af fected area such as prior frostbite, diabetes and ischemia;
  • Those who have known hematological dyscrasias which af fect thrombosis such as paroxysmal cold hemoglobinuria, cryoglobulinemia, sickle-cell disease and serum cold agglutinins.

Product Introduction

The Cryopush Cold Compression Device consists of a main unit and wraps.
Main unit inflates the wrap to a specified pressure of 20 mmHg, 40 mmHg, 60 mmHg, 80 mmHg or 100 mmHg (set by the user), and once the pressure reached the targeted level, it will hold for 10 seconds before releasing, then decompresses and the air pressure drops. The cycle begins again. Default working time is 30 minutes. Built-in rechargeable lithium batteries allow the system to be completely por table for use anytime and anywhere.AO2-P-001 Cryopush Cold Compression Device - Overview

How To Use

If you want compression treatment with cold therapy, follow steps 1.1 and 1.2 to prepare the cold pack; For compression therapy alone, start from step 1.3.

1.Prepare Wrap
1.1 Freeze the gel pack for around 6 hours in a freezer or a refrigerator (temperature-18℃±2℃).
1.2 Place the external wrap flat with Velcro facing up, attach gel pack to wrap, see Pic 1 & 2 below.
1.3 Secure wrap on the knee, adjust to snug fit through Velcro and straps bottom up, Pic 3 &4.
Note: Do not use the cold wrap on exposed skin, apply over clothing or barrier cloth. For replacement gel pack, please contact the after-sale service.
1.4 Connect the air hose of wrap to main unit, Pic 5.AO2-P-001 Cryopush Cold Compression Device - Connect the air hose2. Operate the device
2.1 Turn It On: Long press the power button to turn it on, LCD display will light up as Pic 6.
If not responding, connect the main unit with power supply to charge; It is advised not to use it when charging.
2.2 Set Pressure: Press “PORT I” or “PORT II” button (according to P1/P2 port wrap is connecting with) to set desired pressure as Pic 7. three seconds after wards without fur ther action, it will start working. Pressure is adjustable when it is in operation.
2.3 Set Time: Press “Timer +” or “Timer -” button to switch inter face to set working time as Pic 8, 3 seconds after wards without fur ther action, it will star t working. The default setting is to automatically shut down after 30 minutes operation.
2.4 Turn it Off: When LCD display is on, long press power button to turn it off.AO2-P-001 Cryopush Cold Compression Device - LCD displayNotice:

  1.  When main unit is working, the pressure displayed is real time value.
  2. When using, ensure the wrap is properly connected with port P1/P2; If targeted pressure level is not reached in several minutes, the pump unit will alarm and corresponding port indicator flashes. Pump will stop working and please check if wrap is properly connected or if the wrap is leaking.

Cautions

  1. Indoor use only.
  2. Main unit contains small parts, only professionals are allowed to disassemble.
  3. Avoid sharp objects contacting the wrap in case of damage or gel leakage.
  4. Main unit is only applicable to Cryopush inflatable wraps.
  5. Charge the main unit every 2 or 3 months to maintain batter y integrity if you do not use it for a long time.
  6. Use this device only at its intended body par t and according to instructions.
  7. Do not use the product in high temperature environment.
  8. Keep away from water, do not use if water or other liquid permeated main unit.
  9. Do not use cold wrap on exposed skin, use the wrap only over barrier cloth.
  10. Allow at least one hour inter val before repeating cold therapy.

Warnings

  1. Contains built-in lithium battery, dispose properly.
  2. Do not throw the device in fire in case of explosion.
  3. Use only the power adaptor provided by the manufacturer.
  4. If pulsations, numbness, or throbbing occurs, the wrap may be wrapped too tightly, loosen immediately.
  5. Ensure the main control unit is turned of f and unplugged from the wall outlet prior to and while cleaning.
  6. To avoid the risk of electrical shock, do not remove any panels from the main unit.
  7. Stop using device if swelling occurs; consult physician.

Performance Characteristics

ModelA02- P-001
Main unit Size173*82*40 mm
Main unit Weight340g
Adaptor Input100-240V-50/60Hz
Adaptor Output12V 2A
Pressure Levels20 mmHg, 40 mmHg, 60 mmHg, 80 mmHg, 100 mmHg (±15mmHg)
Working Time10min, 20 min, 30 min, 40 min, 50 min, 60 min, 70 min, 80 min, 90 min, 100 min, 110 min, 120 min
Battery Capacity2600mAh
Charge TimeAround 4 hours
Battery LifeAround 10 hours
Noise Level-.C55dB
Pressure holding time10 seconds
Pressure release time20 seconds

NOTE:

  • Default time setting is 30 minutes.
  • Anti kink Luer connector is used for quick connection/disconnection and avoid kinked tube.

Troubleshooting

Device cannot turn onBattery is low, connect the main unit to power source to charge
Cannot feel pressure1.Check if hose is properly connected to main unit.
2.Wrap is not snug fit, adjust wrap fit.
3.Check if wrap is leaking or damaged.
Pressure abnormalIn normal use, the display shows the pressure value has reached target level while the main unit keeps inflating; the pressure sensor malfunctions and stop using immediately.
Return the device for repair or exchange.
PORT 1″ or “PORT 2” flashes with buzzer sound1.”PORT 1″ flash means the PORT 1 wrap pressure cannot reach targeted level; Short press PORT I button to stop alarm and check the connection between main unit and wrap.
2.”PORT 2″ flash means the PORT 2 wrap pressure cannot reach targeted level; Short press PORT II button to stop alarm and check the connection between main unit and wrap.
Device shuts down during chargingBattery temperature protection: when battery temperature is higher than set value during charging, device will shut down. It will back to normal when battery temperature lowers to normal level.
Device shuts down during working1.Load over current protection: When load increases and the current exceed the set protection threshold, device will shut down.
Use the device when current is back to normal; Return the device for repair.
2.Battery temperature is too high, wait and use till it lowers to normal level.

Symbol Index

warning 2Caution
Indicates that impor tant cautionar y information, such as warnings and precautions, cannot be presented on the medical device itself and are Presented in the instructions for use.
Thermometers PC868. Infrared Thermometer - symbol 1Refer to instruction manual
HOTDOG B107 Head Warming Wrap icon 17Type BF applied par t
AO2-P-001 Cryopush Cold Compression Device - icon 1Unique Device Identifier, Indicates a carrier that contains Unique Device
Identifier information
ICONManufacturer
SMART METER SMPO1000 US iPulseOx Pulse Oximeter - icon 5Date of manufacturer
HOTDOG B107 Head Warming Wrap icon 18Batch code
SMART METER SMPO1000 US iPulseOx Pulse Oximeter - icon 1Serial number
milwaukee M12 SLED Spot Ligh - Icon 1For indoor use only
EGO ST1400E ST 56 Volt Lithium Ion Cordless Line Trimmer - Icon 6Direct current
WEE-Disposal-icon.pngDisposal of product according to local regulations.
IP21Ingress Protection Marking of IP21: Protected from touch by hands greater than 12 millimeters, Protected from condensation.
AO2-P-001 Cryopush Cold Compression Device - icon 2This side up
AO2-P-001 Cryopush Cold Compression Device - icon 4Fragile
HOTDOG B107 Head Warming Wrap icon 12Keep Dr y
AO2-P-001 Cryopush Cold Compression Device - icon 3Maximum number of packages that may be stacked is 6

Standards Compliance

This device adheres to the following standards: IEC 60601-1; ANSI AAMI ES60601-1; EN 60601-1; IEC 60601-1-2; AAMI/IEC60601-1-2; EN 60601-1-2; IEC 60601-1-11; ANSI/AAMI HA60601-1-11; EN 60601-1-11;

Electromagnetic Compatibility

The device satisfies the EMC requirements of the international standard IEC 60601-1-2. The requirements are satisfied under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use.
Table 1 for Emissions
GUIDANCE AND MANUFACTURER’S DECLARATION LECTROMAGNETIC EMISSIONS
The product is intended for use in the electromagnetic environment specified below, The customer or the user of the product should assure that it is used in such an environment.

EMISSIONS TESTCOMPLIANCEELECTROMAGNETIC ENVIRONMENT GUIDANCE
RFemissionsCISPR11Group1The product is only for its internal function. There fore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RFemissionsCISPR11Class BThe product is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC61000-3-2Class A
Voltage fluctuations/flicker emissions IEC61000-3-3Complies

Table 2 for RF Immunity
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE PRODUCT
The product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the product can help prevent electromagnetic inter ference by maintaining a minimum distance between por table and mobile RF communications equipment (transmitters) and the product as recommended below, according to the maximum output power of the communications equipment.

RATED MAXIMUMSEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER
OUTPUT POWER OF TRANSMITTERW150KHz TO 80MHz D= 1.2AO2-P-001 Cryopush Cold Compression Device - icon 5P80MHz TO 800MHz D= 1.2 AO2-P-001 Cryopush Cold Compression Device - icon 5P800MHz TO 2 .5GHz D= 2.3AO2-P-001 Cryopush Cold Compression Device - icon 5 P
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters(m)can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts(W) according to the transmitter manufacturer.
NOTE1: At 80MHz and 800MHz,the separation distance for the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is af fected by absorption and reflection from structures, objects and people.

Table 3 for Transient Electromagnetic Immunity
GUIDANCE AND MANUFACTURER’S DECLARATION LECTROMAGNETIC IMMUNITY
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment.

IMMUNITY TESTIEC60601TEST LEVELCOMPLIANCE LEVELELECTROMAGNETIC ENVIRONMENT-GUIDANCE
Electrostatic discharge (ESD) IEC61000-4-2±8kV contact +2, 4, 8, 15kV air±8kV contact +2, 4, 8, 15kV airFloors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative Humidity should be at least 30%.
Electrical fast transient/ burst IEC61000-4-4+2kV for Power supply lines±2kV for power supply linesMains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±0.5, 1kV line(s) to
1ine(s)
±0.5, 1kV 1ine(s) to
1ine(s)
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions And voltage
variations on power supply Input lines IEC61000-4-11
<5% UT (>95%dip in UT) for
0.5cycle
40% UT (60%dip in UT) for 5cycles 70% UT (30% dip in UT)for 25
cycles <5% UT (>95%dip in UT) for 300 cycles
<5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60%dip in UT) for 5cycles 70% UT (30%dip in UT) for 25cycles< 5% UT (>95%dip in UT) for 5s
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the product requires
continued operation during power mains interruptions, it is commended that the product be powered from an uninterruptible power supply or a battery.
Power frequency (50/60Hz) magnetic field IEC61000-4-830A/m30A/mPower frequency magnetic fields should beat levels characteristic of a typical location in a typical commercial or hospital environment.
Proximity fields from RF wireless communications equipment
IEC 61000-4-3
CompliesComplies/

NOTE UT is the AC mains voltage prior to application of the test level.

Table 4 for RF Electromagnetic Immunity
GUIDANCE AND MANUFACTURER’S DECLARATION LECTROMAGNETIC IMMUNITY
The product is intended for use in the electromagnetic environment specified below.
The customer or the user of the product should assure that it is used in such an environment.

IMMUNITY TESTIEC60601 TEST LEVELCOMPLIANCE LEVELELECTROMAGNETIC ENVIRONMENT‒ GUIDANCE
Conducted
RF IEC610004-6
3Vrms
150kHz to 80MHz
3VrmsPortable and mobile RF communications equipment should be used no closer to any part of the product, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1 .2 √  P
d = 1 .2 √ P
d = 2.3 √ P 800MHzto2 .7GHz
where P is the maximum output Power rating of the transmitter in watts(W)
according to the transmitter manufacturer and dis the recommended Separation distance in meters(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site sur vey, a should be less than the compliance level in
each frequency range. Inter ference may occur in the vicinity of equipment marked with the following symbol:AO2-P-001 Cryopush Cold Compression Device - icon 6
6 Vrms
150 kHz  to 80MHz
ISM frequency band
6Vrms
Radiated RF
IEC610004-3
80MHz
To 2.7GHz
10V/m

NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is af fected by absorption and reflection from structures,objects and people.

FCC Statement

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

Caution: Any changes or modifications to this device not explicitly approved by manufacturer could void your authority to operate this equipment.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:

  1. This device may not cause harmful interference, and
  2. this device must accept any interference received, including interference that may cause undesired operation.

RF Exposure Information
The device has been evaluated to meet general RF exposure r equirement.
The device can be used in portable exposure condition, compliance with exposure requirements.

Main Materials

Main Unit: Acr ylonitrile Butadiene Styrene;
Wrap: Polyvinyl chloride and nylon composite
Cold pack: Thermoplastic polyurethanes, the gel is made from Glycerin, resin and water.

Environment Requirements

Operating:
Temperature: 5℃〜40℃(41℉〜104℉)
Relative humidity: 10%〜90%
Atmospheric pressure: 700~1060hpa
Transport & Storage:
Temperature: -25℃~55℃(-13℉〜131℉)
Relative humidity: 10% 〜90%
Atmospheric pressure: 700~1060hpa
Cleaning
Caution: Main unit must be turned of f and disconnected from power mains (wall outlet) prior to cleaning or disinfection.
Caution: No modification of this equipment is allowed.
* Use soft cloth or towel to wipe to clean.

Software Version

KRP-A02-P-001-A. V1

Service Life

Main unit: 1000 hours
Inflatable wrap: 250 hours

Disposal

Main unit contains rechargeable batteries. Do not discard it in regular waste; The device and packaging must be disposed of in accordance with respective local or national environmental regulations.

Warranty Information

Cr yo-Push warrants that A02-P-001, if properly used, will be free from defects in materials or workmanship for a limited period, main unit is warranted for one (1) year and wrap 6 months after the date of original purchase.
If the product, which is the subject of this Limited Warranty, malfunctions during the warranty period for reasons covered by this Limited Warranty, Cr yo-Push, at its options, may elect to replace or repair the returned product with either a new or reconditioned product. This limited warranty exists only to the original purchaser of the product and are non-transferable.
This limited warranty does not cover damages due to external causes, including, without limitation, accident, usage not in accordance with product instructions, misuse, neglect, alteration, normal wear and tear or unauthorized repairs.

Warranty Registration Card

(Please fill out and return upon purchase of Cr yo-Push Product.)
Your Name:………………………………….
Phone No.:……………………………..
Email Address:……………………
Company/Organization:……………….
Purchase From:………………………………
Purchase Date:……………………………….
Model No.:……………………………….
Series No.:………………………….

Statement
MedWatch is the Food and Drug Administration’s (FDA) program for repor ting serious reactions, product quality problems, therapeutic inequivalence/failure, and product use errors with human medical products, including drugs, biologic products, medical devices, dietar y supplements, infant formula, and cosmetics. If you think you or someone in your family has experienced a serious reaction to a medical product, you are encouraged to take the repor ting form to your doctor.
Your health care provider can provide clinical information based on your medical record that can help FDA evaluate your report.
However, we understand that for a variety of reasons, you may not wish to have the form filled out by your health care provider, or your health care provider may choose not to complete the form. Your health care provider is not required to repor t to the FDA. In these situations, you may complete the Online Repor ting Form yourself.
You will receive an acknowledgement from FDA when your report is received. Repor ts are reviewed by FDA staff.
You will be personally contacted only if we need additional information.
Submitting Adverse Event Repor ts to FDA 6 of 16 Use one of the methods below to submit voluntar y adverse event repor ts to the FDA: Repor t Online at www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=reporting.home 
Consumer Repor ting Form FDA 3500B.
Follow the instructions on the form to either fax or mail it in for submission.
For help filling out the form, see MedWatchLearn. The form is available at www.fda.gov/downloads/aboutFDA/reportsmanualsforms/forms/ucm349464.pdf
Call FDA at 1-800-FDA-1088 to report by telephone.
Repor ting Form FDA 3500 commonly used by health professionals. The form is available at www.fda.gov/downloads/aboutFDA/reportmanualsforms/forms/ucm163919.pdf

Cryopush logoChengdu Cryo-Push Medical Technology Co., Ltd
AO2-P-001 Cryopush Cold Compression Device - icon 7 2, Zone 9 and 102,105, Zone 20, Huayin Industrial Por t, No.618, Kexing
Road (West), Wenjiang District, Chengdu 611137 Sichuan P.R. China
AO2-P-001 Cryopush Cold Compression Device - icon 8 +86 28 85039149 +86 28 65483285
AO2-P-001 Cryopush Cold Compression Device - icon 9 www.cryopush.com
AO2-P-001 Cryopush Cold Compression Device - icon 10 info@cryopush.com

Documents / Resources

CRYOPUSH AO2-P-001 Cryopush Cold Compression Device [pdf] User Manual
AO2-P-001 Cryopush Cold Compression Device, AO2-P-001, Cryopush Cold Compression Device, Cold Compression Device, Compression Device, Device