ℹ️ Document Conversion Notice: This page was converted from the original file for easier reading. Diagrams/images may appear only in the original PDF below.
Document preview
File info: application/pdf · 8 pages · 565.19KB

MaxO2+ Industrial Oxygen Analyzer

Instructions for Use

Brand: Maxtec

Model: MaxO2+

Overview

Base Unit Description

  • Provides unparalleled performance and reliability due to an advanced design.
  • Features an extra-life oxygen sensor with approximately 1,500,000 O2 percent hours (2-year warranty).
  • Durable, compact design for comfortable, hand-held operation and easy cleaning.
  • Operates using two AA Alkaline batteries (2 x 1.5 Volts) for approximately 5000 hours of continuous use. AA Lithium batteries may be used for extended life.
  • Oxygen-specific, galvanic sensor achieves 90% of final value in approximately 15 seconds at room temperature.
  • Large, easy-to-read, 3 1/2-digit LCD display for readings in the 0-100% range.
  • Simple operation with easy one-key calibration.
  • Includes self-diagnostic check of analog and microprocessor circuitry.
  • Features low battery indication.
  • Equipped with a calibration reminder timer that alerts the operator via a calibration icon on the LCD display to perform a unit calibration.

Component Identification

Figure 1 illustrates the components of the MaxO2+ analyzer:

  • 1. 3-DIGIT LCD DISPLAY: Displays oxygen concentrations from 0 – 105.0%. Also shows error codes and calibration codes.
  • 2. LOW BATTERY INDICATOR: Located at the top of the display, activated when battery voltage is below normal operating level.
  • 3. "%" SYMBOL: Appears next to the concentration number during normal operation.
  • 4. CALIBRATION SYMBOL: Located at the bottom of the display, activates when calibration is necessary.
  • 5. ON/OFF KEY: Used to turn the device on or off.
  • 6. CALIBRATION KEY: Used to calibrate the device. Holding for over three seconds initiates calibration mode.
  • 7. SAMPLE INLET CONNECTION: Port for connecting to determine oxygen concentration.

Operating Instructions

Getting Started

Protect Tape

? NOTE: Prior to turning on the unit, remove the protective film covering the threaded sensor face. Wait approximately 20 minutes for the sensor to reach equilibrium after removing the film.

Automatic Calibration

After turning on the unit, it will automatically calibrate to room air. The display should stabilize and read 20.9%.

ℹ️ CAUTION: The device assumes a percent oxygen concentration during calibration. Ensure 100% oxygen or ambient air concentration is applied during calibration for correct results.

Figure 2 shows the sample gas connection setup.

Checking Oxygen Concentration

(After the unit has been calibrated):

  1. Connect the Tygon tubing to the bottom of the analyzer by threading the barbed adapter onto the oxygen sensor (Figure 2, B).
  2. Attach the other end of the sample hose to the sample gas source and initiate flow at a rate of 1-10 liters per minute (2 liters per minute recommended).
  3. Ensure the unit is powered ON using the "ON/OFF" key.
  4. Allow the oxygen reading to stabilize (approximately 30 seconds or more).

Calibrating the MAXO2+ Oxygen Analyzer

? NOTE: Maxtec recommends using medical grade USP or >99% purity oxygen for calibration.

The MAXO2+ Analyzer should be calibrated upon initial power-up and recommended weekly thereafter. A one-week timer starts with each calibration, displaying a reminder icon "CAL" when calibration is due. Calibration is advised if the last procedure date is unknown or if measurement values are in question. Start calibration by pressing the Calibration key for over 3 seconds. The MAXO2+ automatically detects if calibrating with 100% oxygen or 20.9% oxygen (normal air).

? DO NOT attempt to calibrate to any other concentration.

Calibration Requirements

For ID testing or optimum accuracy, a new calibration is required when:

  • The measured O2 percentage in 100% O2 is below 99.0%.
  • The measured O2 percentage in 100% O2 is above 101.0%.
  • The CAL reminder icon is blinking on the LCD.

If unsure about the displayed O2 percentage, refer to "Factors Influencing Accurate Readings".

A simple calibration can be made with the sensor open to ambient air. For optimum accuracy, Maxtec recommends a closed-loop circuit with gas flow across the sensor. Calibrate using the same circuit and flow rate as for measurements.

In Line Calibration (Flow Diverter – Tee Adapter)

  1. Attach the diverter to the MAXO2+ by threading it onto the bottom of the sensor.
  2. Insert the MAXO2+ into the center position of the tee adapter (Figure 2, A).
  3. Attach an open-ended reservoir to the end of the tee adapter. Start calibration flow of oxygen at two liters per minute. (Six to 10 inches of corrugated tubing works well as a reservoir).
  4. Allow the oxygen to saturate the sensor. While a stable value is usually observed within 30 seconds, allow at least two minutes for complete saturation.
  5. If the MAXO2+ is not already ON, press the "ON" button.
  6. Press the Cal button until "CAL" appears on the display (approx. 3 seconds). The analyzer will seek a stable sensor signal and good reading, then display the calibration gas.

? NOTE: Analyzer will display "Cal Err St" if the sample gas has not stabilized.

Direct Flow Calibration (Barb)

  1. Attach the Barbed Adapter to the MAXO2+ by threading it onto the bottom of the sensor.
  2. Connect the Tygon tube to the barbed adapter (Figure 2, B).
  3. Attach the other end of the sampling tube to a source of oxygen with a known concentration value. Initiate flow of the calibration gas at two liters per minute.
  4. Allow the oxygen to saturate the sensor. While a stable value is usually observed within 30 seconds, allow at least two minutes for complete saturation.
  5. If the MAXO2+ is not already ON, press the "ON" button.
  6. Press the Cal button until "CAL" appears on the display (approx. 3 seconds). The analyzer will seek a stable sensor signal and good reading, then display the calibration gas.

Factors Influencing Accurate Readings

Elevation/Pressure Changes

  1. Changes in elevation cause a reading error of approximately 1% per 250 feet.
  2. Calibration should be performed when elevation changes by more than 500 feet.
  3. This device does not automatically compensate for barometric pressure or altitude changes. Recalibrate if moved to a different altitude.

Temperature Effects

The MAXO2+ maintains calibration and reads correctly within ±3% when at thermal equilibrium within the operating temperature range. The device must be thermally stable when calibrated and allowed to stabilize after temperature changes for accurate readings.

  • For best results, perform calibration at a temperature close to the analysis environment.
  • Allow adequate time for the sensor to equilibrate to a new ambient temperature.

ℹ️ CAUTION: "CAL Err St" may result from a sensor not at thermal equilibrium.

Pressure Effects

Readings are proportional to the partial pressure of oxygen (concentration x absolute pressure). Readings are proportional to concentration if pressure is constant.

  • Calibrate the MAXO2+ at the same pressure as the sample gas.
  • If using tubing for sample gases, use the same apparatus and flow rates for calibration as for measurement.

Humidity Effects

Humidity (non-condensing) dilutes the gas, potentially reducing oxygen concentration by up to 4%. The device responds to actual oxygen concentration, not dry concentration. Avoid environments where condensation may occur, as it can obstruct the sensing surface, leading to erroneous readings and slower response times.

  • Avoid usage in environments with greater than 95% relative humidity.

HELPFUL HINT: Dry the sensor by lightly shaking moisture out, or flow dry gas at two liters per minute across the sensor membrane.

Calibration Errors and Error Codes

The MAXO2+ analyzers feature self-testing to detect faulty calibrations, sensor failures, and low voltage. Listed below are error codes and possible actions:

E02: No sensor attached

  • MaxO2+A: Open unit, disconnect and reconnect sensor. Unit should auto-calibrate and read 20.9%. If not, contact Maxtec Customer Service.
  • MaxO2+AE: Disconnect and reconnect external sensor. Unit should auto-calibrate and read 20.9%. If not, contact Maxtec Customer Service.

E03: No valid calibration data available

  • Ensure unit has reached thermal equilibrium. Press and hold the Calibration Button for three seconds to manually force a new calibration.

E04: Battery below minimum operating voltage

  • Replace batteries.

CAL ERR ST: O2 Sensor reading not stable

  • Wait for the displayed oxygen reading to stabilize when calibrating with 100% oxygen.
  • Wait for the unit to reach thermal equilibrium (up to one hour if stored outside the specified operating temperature range).

CAL ERR LO: Sensor voltage too low

  • Press and hold the Calibration Button for three seconds to manually force a new calibration. If the error repeats more than three times, contact Maxtec Customer Service.

CAL ERR HI: Sensor voltage too high

  • Press and hold the Calibration Button for three seconds to manually force a new calibration. If the error repeats more than three times, contact Maxtec Customer Service.

CAL ERR BAT: Battery voltage too low to recalibrate

  • Replace batteries.

Changing the Batteries

Batteries should be changed by service personnel.

  • Use only brand name batteries.
  • Replace with two AA batteries, observing the orientation marked on the device.

The device indicates a battery change is needed in two ways:

  • The battery icon (?) on the display will flash. The unit will continue to function normally for approximately 200 hours.
  • If a very low battery level is detected, error code "E04" will appear, and the unit will not function until batteries are changed.

To change batteries:

  1. Remove the three screws from the back of the device using a #1 Phillips screwdriver.
  2. Gently separate the two halves of the device.
  3. Replace the batteries, ensuring correct polarity as indicated on the back case.

? NOTE: Incorrect battery installation will prevent contact and operation.

  1. Carefully bring the two halves of the case together, ensuring wires are not pinched. The gasket will be captured on the back case half.
  2. Reinsert and tighten the three screws until snug (Figure 3).

The device will automatically perform a calibration and display the % of oxygen.

HELPFUL HINT: If the unit does not function, verify screws are tight for proper electrical connection.

HELPFUL HINT: Before closing the case halves, ensure the keyed slot on the coiled cable assembly engages the tab on the back case for correct orientation and to prevent rotation, which could hinder closing and operation.

Changing the Oxygen Sensor

MAXO2+AE Model

If the oxygen sensor needs changing, the device will display "Cal Err lo".

  1. Unthread the sensor from the cable by rotating the thumbscrew connector counterclockwise.
  2. Pull the sensor from the connection.
  3. Replace the new sensor by inserting its electrical plug into the receptacle on the oxygen sensor.
  4. Rotate the thumbscrew clockwise until snug. The device will automatically calibrate and display the % of oxygen.

Cleaning and Maintenance

Store the MAXO2+ analyzer at a temperature similar to its ambient environment of daily use.

Instrument Cleaning

When cleaning or disinfecting the exterior, take care to prevent solutions from entering the instrument.

? DO NOT immerse the unit in fluids.

  • Clean the surface using a mild detergent and a moist cloth.
  • The MAXO2+ analyzer is not intended for steam, ethylene oxide, or radiation sterilization.

Oxygen Sensor

⚠️ WARNING: Never install the sensor where it may be exposed to patient's exhaled breath or secretions, unless disposal of the sensor, flow diverter, and tee adapter is intended after use.

  • Clean the sensor with a cloth moistened with isopropyl alcohol (65% alcohol/water solution).
  • Maxtec does not recommend spray disinfectants as salts can accumulate and impair readings.
  • The oxygen sensor is not intended for steam, ethylene oxide, or radiation sterilization.

Accessories

The flow diverter and tee adapter may be disinfected by washing with isopropyl alcohol. Ensure parts are thoroughly dry before use.

Specifications

Base Unit Specifications

MeasurementValue
Measurement Range0-100%
Resolution0.1%
Accuracy and Linearity1% of full scale at constant temperature, R.H. and pressure (when calibrated at full scale)
Total Accuracy±3% actual oxygen level over full operating temp range
Response Time90% of final value in approximately 15 seconds at 23°C
Warm-up TimeNone required
Operating Temperature15°C - 40°C (59°F - 104°F)
Storage Temperature-15°C - 50°C (5°F - 122°F)
Atmospheric Pressure800-1013 mBars
Humidity0-95% (non-condensing)
Power Requirements2, AA Alkaline batteries (2 x 1.5 Volts)
Battery LifeApproximately 5000 hours (continuous use)
Low Battery Indication"BAT" icon (?) displayed on LCD
Sensor TypeMaxtec MAX-250 series galvanic fuel cell
Expected Sensor Life> 1,500,000 O2 percent hours minimum (2-year in typical medical applications)
Dimensions (A Model)3.0"(W) x 4.0"(H) x 1.5"(D) [76mm x 102mm x 38mm]
Weight (A Model)0.4 lbs. (170g)
Dimensions (AE Model)3.0"(W) x 36.0"(H) x 1.5"(D) [76mm x 914mm x38mm] (Height includes external cable length retracted)
Weight (AE Model)0.6 lbs. (285g)
Drift of Measurement< +/-1% of full scale at constant temperature, pressure and humidity

Sensor Specifications

TypeValue
Galvanic fuel sensor (0-100%)0-100%
Life2-years in typical applications

Spare Parts and Accessories

Included With Your Unit

PART NUMBERITEM
R217M72User's Guide and Operating Instructions
RP76P06Lanyard
R110P10-001Flow Diverter
RP16P02Blue Tee Adapter
R217P35Dovetail Bracket
R125P03-004MAX-250E Oxygen Sensor
R217P08Gasket
RP06P25#4-40 Pan Head Stainless Steel Screw
R217P16-001Front Assembly (Includes Board & LCD)
R217P11-002Back Assembly
R217P09-001Overlay

Optional Accessories

Optional Adapters

PART NUMBERITEM
RP16P02Blue Tee Adapter
R103P90Perfusion Tee Adapter
RP16P12Long-Neck Tee Adapter
RP16P05Pediatric Tee Adapter
RP16P10MAX-Quick Connect
R207P17Threaded Adapter with Tygon Tubing

Mounting Options (requires dovetail R217P23)

PART NUMBERITEM
R206P75Pole Mount
R205P86Wall Mount
R100P10Rail Mount
R213P31Swivel Mount

Carrying Options

PART NUMBERITEM
R217P22Belt Clip and Pin
R213P02Zipper Carrying Case with Shoulder Strap
R213P56Deluxe Carrying Case, Water Tight
R217P32Soft Case, Tight Fit Carrying Case

? NOTE: Repair of this equipment must be performed by a qualified service technician experienced in repair of portable hand held medical equipment.

Equipment in need of repair shall be sent to:

Maxtec, Service Department, 2305 South 1070 West, Salt Lake City, Ut 84119 (Include RMA number issued by customer service)

Electromagnetic Compatibility

Electromagnetic Emissions

This equipment is intended for use in the specified electromagnetic environment. The user should assure it is used in such an environment.

EMISSIONSCOMPLIANCE ACCORDING TOELECTROMAGNETIC ENVIRONMENT
RF Emissions (CISPR 11)Group 1The MaxO2+ uses RF energy only for its internal function. Its RF emissions are very low and unlikely to cause interference in nearby electronic equipment.
CISPR Emissions ClassificationClass AThe MaxO2+ is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. ? NOTE: Suitable for industrial areas and hospitals (CISPR 11 class A). In residential environments (CISPR 11 class B normally required), adequate protection may not be offered to radio-frequency communication services. Mitigation measures like relocating or re-orienting may be necessary.
Harmonic Emissions (IEC 61000-3-2)Class A
Voltage FluctuationsComplies

Recommended separation distances between portable and mobile RF communications equipment and the MaxO2+:

RATED MAXIMUM OUTPUT POWER OF TRANSMITTER (W)Separation distance according to frequency of transmitters (meters)
150 kHz to 80 MHz80 MHz to 800 MHz800MHz to 2.5 GHz
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223

For transmitters not listed, separation distance 'd' (meters) can be estimated using the equation applicable to the transmitter's frequency, where 'P' is the maximum output power in watts (W). ? NOTE 1: At 80 MHz and 800 MHz, use the distance for the higher frequency range. ? NOTE 2: Guidelines may not apply in all situations due to absorption and reflection from structures.

Electromagnetic Immunity

This equipment is intended for use in the specified electromagnetic environment. The user should assure it is used in such an environment.

IMMUNITY AGAINSTIEC 60601-1-2: (4TH EDITION) TEST LEVELELECTROMAGNETIC ENVIRONMENT
Electrostatic discharge, ESD (IEC 61000-4-2)Contact discharge: ±8 kV Air discharge: ±2 kV, ±4 kV, ±8 kV, ±15 kVFloors should be wood, concrete, or ceramic tile. If covered with synthetic material, relative humidity should be kept low to reduce electrostatic charge.
Electrical fast transients/bursts (IEC 61000-4-4)Power supply lines: ±2 kV Longer input/output lines: ±1 kVMains power quality should be typical of a commercial or hospital environment.
Surges on AC mains lines (IEC 61000-4-5)Common mode: ±2 kV Differential mode: ±1 kVEquipment emitting high levels of power line magnetic fields (in excess of 30A/m) should be kept at a distance to reduce interference. If continued operation during power mains interruptions is required, ensure batteries are installed and charged, or provide an uninterruptible power source.
3 A/m power frequency magnetic field 50/60 Hz (IEC 61000-4-8)30 A/m 50 Hz or 60 Hz
Voltage dips and short interruptions on AC mains input lines (IEC 61000-4-11)Dip>95%, 0.5 periods Dip 60%, 5 periods Dip 30%, 25 periods Dip >95%, 5 seconds

This equipment is intended for use in the specified electromagnetic environment. The customer or user should assure it is used in such an environment.

IMMUNITY TESTELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Conducted RF coupled into lines (IEC 61000-4-6)3V (0.15 - 80 MHz) 6V (ISM bands)Portable and mobile RF communications equipment (including cables) should be used no closer than the recommended separation distance calculated from the equation applicable to the transmitter's frequency. Recommended separation distance: d=1.2 √P (80 MHz to 800 MHz) and d=2.3 √P (800 MHz to 2.7 GHz), where P is the transmitter's maximum output power in watts. Field strengths from fixed RF transmitters should be less than compliance levels. Interference may occur near equipment marked with the symbol ((())). The ISM bands are specified. Field strengths from fixed transmitters cannot be predicted theoretically; an electromagnetic site survey may be considered. If abnormal performance is observed, measures like re-orienting or relocating may be necessary.
Radiated RF immunity (IEC 61000-4-3)3 V/m 80 MHz - 2.7 GHz 80% @ 1 KHz AM Modulation

Maxtec | 2305 South 1070 West, Salt Lake City, Utah 84119 | (800) 748-5355 | www.maxtec.com

Models: MaxO2, Oxygen Analysis

Original Document

If the viewer doesn’t load, open the PDF directly.