Baxtran logoABD COMPACT BALANCE WEIGHT CONTROL
User Manual

Baxtran ABD Measuring Scales

The manufacturer reserves the right to modify the specifications of its products in order to make technical improvements or comply with new regulations.

EXCITATION

ABD

Input230~240V
Output10V 600mA
Rechargeable Battery6V/4Ah

BEFORE ITS USE

  1. Use an independent electric source to prevent electronic disturbances.
  2. Don’t place any object on the platform when switching on the indicator.
  3. Please, warm-up the scale during 2-3 minutes before using it.
  4. Avoid sudden changes in temperature and draughts.
  5. Don’t overload the scale; do not exceed its maximum capacity.

CONSUMPTION

ABD

Battery life:Without back illumination, approx., 160 hours.
With back illumination, approx., 120 hours.

KEYBOARD DESCRIPTION

Baxtran ABD Measuring Scales - KEYBOARD DESCRIPTION

Baxtran ABD Measuring Scales - icon1First function; Press this key to switch on the scale.
Baxtran ABD Measuring Scales - icon2First function; Press and hold the key during 3 seconds to switch of the indicator.
Baxtran ABD Measuring Scales - icon3First function; to choose the unit of weight.
Second function; to exit from the programming mode.
Baxtran ABD Measuring Scales - icon4First function; : to remove (tare) the weight of a container.
Second function; to move to the right in the programming mode.
Baxtran ABD Measuring Scales - icon5First function; 0.
Second function; to move to the left.
Baxtran ABD Measuring Scales - icon6First function; to view the brute weight after that to do the tare (weight) the container.
Second function; To increase the values inside the programming.
Baxtran ABD Measuring Scales - icon7First function; to accumulate in memory the value of the weight that appears in the screen.
Second function; manual transmission of data through RS-232 port to a PC or printer. (No function ABD)
Third function; confirmation key in the programming mode..

ABD APPLICATIONS

5.1 NORMAL WEIGHING MODE
5.1.1 EQUIPMENT CONFIGURATION
5.1.2 FIRST CALIBRATION
5.1.3 USE
5.2 PIECE COUNTING MODE
5.2.1 EQUIPMENT CONFIGURATION
5.2.2 FIRST CALIBRATION
5.2.3 USE
5.3 LIMITS AND ALARM
5.4 UNIT RANGE/ MULTI RANGE/ MULTI INTERVAL
5.1 NORMAL WEIGHING MODE
5.1.1 CONFIGURATION OF THE EQUIPMENT
See section LF2 of the technical parameters
5.1.2 FIRST CALIBRATION
See section LF1 of the technical parameters
5.1.3 USE
Switch on the equipment when all the parameters have been correctly configured and the equipment has been calibrated

  • Make sure than the value of the indicator, without load on the platform, is 0. If this is not the case, pressBaxtran ABD Measuring Scales - icon5.
  • Place the weight on the platform and the platform will show the weight.
  • The accumulation and sending of data will depend on the mode chosen in the section UF-6

5.2 PIECE COUNTING MODE
5.2.1 CONFIGURATION OF THE EQUIPMENT
See section LF2 of the technical parameters
5.2.2 FIRST CALIBRATION
See section LF1 of the technical parameters
5.2.3 USE
Switch on the equipment when all the parameters have been correctly configured. Make sure that the value on the visor, with no load on the platform, is 0. If this is not the case press the keyBaxtran ABD Measuring Scales - icon5 .
STEPS TO FOLLOW;

  1. Press the keyBaxtran ABD Measuring Scales - icon3 until the symbol PCS appears on the screen.
  2. Press the keyBaxtran ABD Measuring Scales - icon6 successively to choose the quantity of pieces of the sample. On the screen will appear, successively, C10/C20/ C50/C100/C200.
  3. Place the sample on the platform, and wait until the sign of stability and press the keyBaxtran ABD Measuring Scales - icon7 .
  4. Place the product on the platform and the screen will show the number of pieces.

The accumulation and the sending of data will depend on the mode chosen in the section UF-6.

  • To turn to the normal weighing mode, press the keyBaxtran ABD Measuring Scales - icon3
  • If the user wants to go back to the piece counting mode, using the same sample of reference, press the keyBaxtran ABD Measuring Scales - icon3 again.
  • If the user want to change the sample of reference, the user must repeat the steps described above.

5.3 LIMITS AND ALARM
The user can configure the superior and inferior limits of the sample placed on the platform.
The display will show if the sample is lower to the inferior limit Lo, above the superior limit Hi or in the zone between the two limits OK The user can configure when he wants to make the alarm ring and the stability needed to make it happen.
All the procedure is described in the section UF-2.
If the user wants to define the limits in the normal weighing mode and wants to use the limits in piece counting mode, he must define the new limits for this mode, when it changes to weight mode again, the user will recover the limits he already has. The same  happens otherwise.
5.4 UNIT RANGE / MULTI-RANGE / MULTI-INTERVAL
The indicator can be configured with an only range, a maximum weight and a value of step. It can also be configured as multi range or multi interval, in such cases there is a maximum weight.
From 0 to the medium weight of these maximum weight it is used the value of the chosen step ( step 1) and from the half to the maximum weight it is used the next value in the step ( step 2).
The screen indicators R1 and R2 point out the range which the user is using at every moment.
In the multi interval mode the weight increases, in the range use it used step 1, when the user goes to range 2 the step 2 is used.
When the weight decreases and the user go back to range 1, the step 1 is used again. On the contrary, in mode multi range, when the weight decreases and the device go back to range 1, the device continues using step 2 until it reaches 0.
In the section LF2 of the technical parameters the user can choose the range mode.

SETTINGS

PARAMETER DESCRIPTION
Baxtran ABD Measuring Scales - icon8Internal calculation (A/D)
Limit Configuration of weight ( superior and inferior)
auto auto off
Baxtran ABD Measuring Scales - icon9Back illumination of display
Four modes of hold
RS-232 Output (PC/PRINT) (No function ABD)
Baxtran ABD Measuring Scales - icon10Configuration of the speed of the converter (A/D)
blind
Configuration of gravity
Baxtran ABD Measuring Scales - icon11Speed of filter
Stabilization speeds

PARAMETERS CONFIGURATION

7.1 INTERNAL COUNTING (A/D)
7.2 CONFIGURATION OF THE WEIGHT LIMITS (SUPERIOR AND INFERIOR)
7.3 AUTO SWITCH OFF
7.4 CONFIGURATION OF THE ILLUMINATION OF THE DISPLAY
7.5 HOLD FUNCTION
7.6 RS-232 DATA EXIT (NO FUNCTION ABD)
7.7 CONFIGURATION OF THE SPEED OF THE AD CONVERTER
7.8 BLIND
7.9 CONFIGURATION OF THE GRAVITY
7.10 SPEED OF FILTER
7.11 STABILIZATION SPEEDS

To access the configuration of parameters when the screen is in zero, the user must press at the same time the keysBaxtran ABD Measuring Scales - icon7 andBaxtran ABD Measuring Scales - icon4. Press the keyBaxtran ABD Measuring Scales - icon5 orBaxtran ABD Measuring Scales - icon4 to choose the desired character (UF-1 ~ UF-11) To go back to the previous mode press the keyBaxtran ABD Measuring Scales - icon3.
7.1 INTERNAL COUNTING (A/D) | Baxtran ABD Measuring Scales - icon12

  1. Press the keyBaxtran ABD Measuring Scales - icon7 to view the internal sums of the scale.
  2. To go to the next parameter, press the keyBaxtran ABD Measuring Scales - icon7 the screen will show the value of the voltage of the battery.
  3. To exit this mode and go back to the normal weighing, the user must press the keyBaxtran ABD Measuring Scales - icon7 orBaxtran ABD Measuring Scales - icon3.

7.2 CONFIGURATION OF THE LIMITS OF WEIGHT ( SUPERIOR AND LOWER) | Baxtran ABD Measuring Scales - icon13

  1. Press the keyBaxtran ABD Measuring Scales - icon7 to access the parameter.
  2. The display will show the message “ Baxtran ABD Measuring Scales - icon14” ( inferior limit)
  3. Use the keysBaxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon4 to move the cursor and the keyBaxtran ABD Measuring Scales - icon6 to choose the desired number.
  4. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.
  5. The display will show the message “Baxtran ABD Measuring Scales - icon15” (Superior limit= Hi)
  6. Use the keysBaxtran ABD Measuring Scales - icon4 andBaxtran ABD Measuring Scales - icon5 to move the cursor and press the keyBaxtran ABD Measuring Scales - icon6 to choose the desired number.
    Note: You can can stay over the last digit to modify if you wish.
  7. Press the Key Baxtran ABD Measuring Scales - icon7 to confirm.
  8. El display shows the value.
    A B C
    0 0 0
    (configuration of the alarm)
    DISPLAYVALUEESTABILITY
    A0There is no need to stabilize the alarm to make it ring
    1The alarm must be stabilized to ring
    B0Always 0
    C0
    1
    2
    Alarm switched off
    The alarm ring if it is place on the band OK
    The alarm rings if it is situated below the HiBaxtran ABD Measuring Scales - icon16
  9. Press the keysBaxtran ABD Measuring Scales - icon4 andBaxtran ABD Measuring Scales - icon5 to move the cursor and the keyBaxtran ABD Measuring Scales - icon6 to choose the desired number.
  10. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.

7.3 AUTO SWITCH OFF | Baxtran ABD Measuring Scales - icon17

MODES:

  • AoFF 00 – Auto switch off deactivated
  • AoFF 01 – Auto switch off activated in a minute. The scale is going to switch off automatically after 1 minute of not being used.
  • You can configure the value wished from 1 to 99 minutes.
  1. Press the keyBaxtran ABD Measuring Scales - icon7 to have an access to the parameter.
  2.  Press the keysBaxtran ABD Measuring Scales - icon4 andBaxtran ABD Measuring Scales - icon5 to move the cursor and the keyBaxtran ABD Measuring Scales - icon6 to choose the desired number.
  3. Press the key Baxtran ABD Measuring Scales - icon7to confirm.

7.4 DISPLAY BACKLIGHTING | Baxtran ABD Measuring Scales - icon18
MODES:

  • A: Automatic.
  • ON: Illumination Activated.
  • OFF: Illumination Deactivated
  1. Press the keyBaxtran ABD Measuring Scales - icon7 to have an access to the parameter.
  2. Press the keyBaxtran ABD Measuring Scales - icon6 to select the desired mode.
  3. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.

7.5 HOLD FUNCTION | Baxtran ABD Measuring Scales - icon19
(Once the object is retired from the plate, the display maintains the weight fixed until press the KeyBaxtran ABD Measuring Scales - icon3 ,This function is very useful for the weighing of animals). Note; if you activate the hold mode, not can use the function limits and accumulation.

  1. Press the keyBaxtran ABD Measuring Scales - icon7 to access to the parameter.
  2. Press the keyBaxtran ABD Measuring Scales - icon6 to choose the desired mode.
  3. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.

MODES:

  • HOLD 0 : Deactivated.
  • HOLD 1 : Animal in movement
  • HOLD 2 : Value of peak
  • HOLD 3 : Hold steady
  • HOLD 4 : Hold steady with self cancelling at zero.

HOLD 1.

  • When the user can access this parameter, the screen shows the message PCt,
  • Use the keysBaxtran ABD Measuring Scales - icon4 andBaxtran ABD Measuring Scales - icon5 to move the cursor and the keyBaxtran ABD Measuring Scales - icon6 to choose the desired value of the range of HOLD, you can choose a number from 001 to 100.
  • Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.
  • It will appear on the screen the message time 8, use the keysBaxtran ABD Measuring Scales - icon4 andBaxtran ABD Measuring Scales - icon5 to move the cursor and the keyBaxtran ABD Measuring Scales - icon6 to choose the number of times you want to repeat during the range of hold.
  • Press the keyBaxtran ABD Measuring Scales - icon7 to confirm. Example: PCt small and big time means more accuracy and longer stabilization..

7.6 ACCUMULATION|Baxtran ABD Measuring Scales - icon20 (NO FUNCTION ABD)

FORMAT OF THE DATA

232 1
232 2
232 3
NO FUNCTION ABD
232 4
232 5
232 6
NO FUNCTION ABD
232 7Connection sending and manual accumulation, passing by zero and pressing the keyBaxtran ABD Measuring Scales - icon7 . When the display is set to 0, press the key Baxtran ABD Measuring Scales - icon7 twice and removed from weighing memory.
232 8Connection sending and manual accumulation, passing by zero and pressing the keyBaxtran ABD Measuring Scales - icon7 . When the display is set to 0, press the keyBaxtran ABD Measuring Scales - icon7 twice and removed from weighing memory.
232 9Connection sending and manual accumulation, passing by zero and pressing the keyBaxtran ABD Measuring Scales - icon7 . When the display is set to 0, press the keyBaxtran ABD Measuring Scales - icon7 twice and removed from weighing memory.
232 10Connection sending and manual accumulation, passing by zero and pressing the keyBaxtran ABD Measuring Scales - icon7 . When the display is set to 0, press the key Baxtran ABD Measuring Scales - icon7 twice and removed from weighing memory.

7.6.1 TRANSMISSION SPEED

  1. Press the keyBaxtran ABD Measuring Scales - icon7 to access the parameter
  2. Press the keyBaxtran ABD Measuring Scales - icon6 to choose the data exit mode wanted.
  3. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.
    The screen shows the different possibility:
    SPEED OF TRANSMISSIOIN
    b12001200 baudios
    b24002400 baudios
    b48004800 baudios
    b96009600 baudios
    b1920019200 baudios
    b3840038400 baudios
  4. Press the keyBaxtran ABD Measuring Scales - icon6 to choose the speed of transmission needed.
  5. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.

7.7 CONFIGURATION OF THE SPEED OF THE CONVERTER|Baxtran ABD Measuring Scales - icon21

  1. Press the keyBaxtran ABD Measuring Scales - icon7 to access the parameter.
  2. Press the keyBaxtran ABD Measuring Scales - icon6 to choose the desired mode:
    • Mode 1 : Normal
    • Mode 2: Fast
    • Mode 3: Slow
  3. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.

7.8 BLIND|Baxtran ABD Measuring Scales - icon22

It appears 0 on screen until the next division is selected. It starts to show values from that division.
Example:
Scale with e =2g
Blind in 5 divisions
It will show 0 until it reaches 25= 10 g, the first value it will show will be 12g

  1. Press the keyBaxtran ABD Measuring Scales - icon7 to access the parameter
  2. Press the keyBaxtran ABD Measuring Scales - icon6 to select.
  3. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.

7.9 CONFIGURATION OF GRAVITY|Baxtran ABD Measuring Scales - icon23

  1. Press the keyBaxtran ABD Measuring Scales - icon7 to view the value of the actual gravity.
  2. To change the value, press the keyBaxtran ABD Measuring Scales - icon7 , next you must use the keyBaxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon4 to move the cursor and the keyBaxtran ABD Measuring Scales - icon6 to select the desired number.
  3. Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.

CONFIGURATION OF THE READJUSTMENTS IN CALIBRATION

  1. When the user is in the normal mode of weighing, he must press the keysBaxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon7 , the message ECF-1 is going to appear on the display.
  2. Press the keysBaxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon4 to select the desired function: ECF-1, ECF-2 or ECF-3

* ECF-1 CALIBRATION OF ZERO + WEIGHT

Press the keyBaxtran ABD Measuring Scales - icon7 , the display will show CALZ.
Press the keyBaxtran ABD Measuring Scales - icon7 , to put the reading of the display to zero.
Press the keysBaxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon4 to move the cursor.
Press the keyBaxtran ABD Measuring Scales - icon6 to introduce the value of the weight of calibration..
Place the weight of calibration on the platform and press the keyBaxtran ABD Measuring Scales - icon7 to do the calibration once the reading is steady.
* ECF-2 CALIBRATION OF ZERO
Press the keyBaxtran ABD Measuring Scales - icon7  , the display will show CALZ.
Press the keyBaxtran ABD Measuring Scales - icon7 , to put the reading of the display to zero.
Press the keyBaxtran ABD Measuring Scales - icon7 , to calibration.
* ECF-3 CALIBRATION OF WEIGHT (SPAN)
Press the keyBaxtran ABD Measuring Scales - icon7 , the display will show the value of the weight of calibration.
Press the keys Baxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon4 to move the cursor.
Press the keyBaxtran ABD Measuring Scales - icon6 to modify the value of the weight of calibration.
Press the keyBaxtran ABD Measuring Scales - icon7 to confirm.
Place the weight of calibration on the platform and press the keyBaxtran ABD Measuring Scales - icon7 to do the calibration once the reading is stable.

TECHNICAL PARAMETERS

DO NOT MODIFY THE TECHNICAL PARAMETERS IF IT IS NOT STRICTLY NEEDED. A BAD CONFIGURATION OF THIS SECTION CAN CAUSE A WRONG FUNCTIONING OF THE SCALE.
ENTRANCE AND EXIT OF THE CALIBRATION

DISPLAYDESCRIPTION AND SEQUENCE OF USE
Baxtran ABD Measuring Scales - icon98With the visor switched off, press and hold the keyBaxtran ABD Measuring Scales - icon1 andBaxtran ABD Measuring Scales - icon5 until the message Baxtran ABD Measuring Scales - icon84 , appears on the screen, then you can release the keys
With the keysBaxtran ABD Measuring Scales - icon6 , Baxtran ABD Measuring Scales - icon5andBaxtran ABD Measuring Scales - icon4 introduce the code Baxtran ABD Measuring Scales - icon85.
Press the keyBaxtran ABD Measuring Scales - icon7 to start or the keyBaxtran ABD Measuring Scales - icon3 to exit the menu and the indicator will begin again auto- magically.

CALIBRATION OF THE WEIGHT Baxtran ABD Measuring Scales - icon86

DISPLAYDESCRIPTION and SEQUENCE OF USE
Baxtran ABD Measuring Scales - icon89Baxtran ABD Measuring Scales - icon25 The calibration can be done with any weight, but the weight ca not be inferior to 1/100 of the maximum capacity and it must also never be exceeded.
• Press the keyBaxtran ABD Measuring Scales - icon7 to start the calibration of zero ( pressBaxtran ABD Measuring Scales - icon3 to exit the calibration and go back to the menu LF1)
• UseBaxtran ABD Measuring Scales - icon5 ,Baxtran ABD Measuring Scales - icon4 andBaxtran ABD Measuring Scales - icon6 , 0 ~ 9 and thenBaxtran ABD Measuring Scales - icon7 pto introduce the weight with which the calibration will be done. ( press the keyBaxtran ABD Measuring Scales - icon3to exit the calibration and go back to the menu LF1)
• Place the required weight on the scale as it is indicated in the display.
• Once everything is steady, press the keyBaxtran ABD Measuring Scales - icon7 to calibrate it ( press the key ESC to exit the calibration and go back to the menu LF1).
THE CALIBRATION IS GOING TO FINISH AND THE SCALE WILL GO VACK TO THE WEIGHING MODE AUTOMATICALLY.

CONFIGURATION Baxtran ABD Measuring Scales - icon87

FIRST STEP

DISPLAYDESCRIPTION AND SEQUENCE OF USE

Baxtran ABD Measuring Scales - icon32FIRST STEP

• DISPLAY OF THE INTERNAL COUNTING
SECOND STEP• POSSIBLE VALUES OF THE PARAMETERS

Baxtran ABD Measuring Scales - icon33A B C D E F

A: Metric system0:NO1: kg2:T3:g
B: American system0:NO1: lb2:lb oz3:VISS
C: other unities0:NO1: TW kg2:HK kg
D: PCS0:OFF1: ON
E: double range0:OFF1: multi interval2:multi range
F: units of calibration1:use metric2:use American unities

lb oz cannot be selected as unities of calibration.
Baxtran ABD Measuring Scales - icon25The scale won’t let us continue up to the next step if there is a mistake during the programming.

Baxtran ABD Measuring Scales - icon89• UseBaxtran ABD Measuring Scales - icon4 , Baxtran ABD Measuring Scales - icon5andBaxtran ABD Measuring Scales - icon6 , 0 ~ 9 and then pressBaxtran ABD Measuring Scales - icon7 to introduce THE MAXIMUM CAPACITY.
•Use the KeysBaxtran ABD Measuring Scales - icon5 ,Baxtran ABD Measuring Scales - icon4to change the position of the DECIMAL POINT.
Baxtran ABD Measuring Scales - icon36
• UseBaxtran ABD Measuring Scales - icon6 to select the DIVISION: Baxtran ABD Measuring Scales - icon38
Baxtran ABD Measuring Scales - icon90Baxtran ABD Measuring Scales - icon25 After introducing the parameters LF2, the indicator will show the last configuration saved. All the steps to follow must be completed, if they are not done the indicator will continue with the previous configuration.
Baxtran ABD Measuring Scales - icon25 Proceed with the calibration of weight after LF2.
• Press the keyBaxtran ABD Measuring Scales - icon7 to start and display the internal value. PressBaxtran ABD Measuring Scales - icon3 to exit the menu and the scale will automatically restart.
• Press the keyBaxtran ABD Measuring Scales - icon7 To continue with the configuration. (press the keyBaxtran ABD Measuring Scales - icon3 to exit from the configuration and go back to the menu LF2).
Use the keysBaxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon6 , 0 ~ 9 and then press the keyBaxtran ABD Measuring Scales - icon7 to choose the UNITS OF WEIGHT . (press the keyBaxtran ABD Measuring Scales - icon3 to exit from the configuration and go back to the menu LF2).
Baxtran ABD Measuring Scales - icon91• Use the keys Baxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon6 , 0 ~ 9 and then press the keyBaxtran ABD Measuring Scales - icon7 to choose the MAXI-MUM WEIGHT. press the key Baxtran ABD Measuring Scales - icon3to exit from the configuration and go back to the menu LF2).
• Use the keys Baxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5 and thenBaxtran ABD Measuring Scales - icon7 To move the decimal point ( press the key To exit from the configuration and go back to the menu LF2).
Baxtran ABD Measuring Scales - icon92• UseBaxtran ABD Measuring Scales - icon6 and thenBaxtran ABD Measuring Scales - icon7 the change the step.(Press the keyBaxtran ABD Measuring Scales - icon3 to exit from the configuration and go back to the menu LF2).
• Use Baxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon7 To continue with the other configuration or press the keyBaxtran ABD Measuring Scales - icon3 to exit from the menu and the scale will begin again automatically).

LINEAL CALIBRATION Baxtran ABD Measuring Scales - icon88
Up to six steps of calibration W0 ~W6
Make sure that the plate of the scale is empty before starting the calibration.
Press the keyBaxtran ABD Measuring Scales - icon5 or the first point of calibration.
The previous points of calibration will be removed and the message CLEAR is going to appear on the screen.
Press the keyBaxtran ABD Measuring Scales - icon5 to proceed with the next point of calibration.
Press the key Baxtran ABD Measuring Scales - icon5 to go back to the previous point of calibration.
Press the keyBaxtran ABD Measuring Scales - icon7 to save everything.
Press the keyBaxtran ABD Measuring Scales - icon3 to finish the calibration and go back to the menu LF3.

  DISPLAYDESCRIPTION AND SEQUENCE OF USE
Baxtran ABD Measuring Scales - icon93• Press the KeyBaxtran ABD Measuring Scales - icon7 to start or the keyBaxtran ABD Measuring Scales - icon3 to exit the menu and the scale will start again auto-magically.

• Press the KeyBaxtran ABD Measuring Scales - icon5 to calibrate the zero. ( press the keyBaxtran ABD Measuring Scales - icon3  to exit from the calibration and go back to the menu LF3)
• Place 1/3 of the maximum weight on the plate and pressBaxtran ABD Measuring Scales - icon5 to calibrate 2/3 th the capacity. (press the keyBaxtran ABD Measuring Scales - icon3 to exit from the calibration and go back to the menu LF3).

Baxtran ABD Measuring Scales - icon94• Place 2/3 of the maximum weight on the plate and pressBaxtran ABD Measuring Scales - icon5 to calibrate 2/3 th the capacity. (press the keyBaxtran ABD Measuring Scales - icon3 to exit from the calibration and go back to the menu LF3).
• Place the maximum weight on the plate and press the keyBaxtran ABD Measuring Scales - icon5 to calibrate the full capacity. (press the keyBaxtran ABD Measuring Scales - icon3 to exit from the calibration and go back to the menu LF3).
Baxtran ABD Measuring Scales - icon95• Press the keyBaxtran ABD Measuring Scales - icon7 to complete the lineal calibration. (press the keyBaxtran ABD Measuring Scales - icon3 to exit from the calibration and go back to the menu LF3)
• Use the keys Baxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5   and then the keyBaxtran ABD Measuring Scales - icon7 to continue with other adjustments and press the keyBaxtran ABD Measuring Scales - icon3 to exit from the menu and the scale will begin again automatically.

VELOCIDAD DEL CONVERTIDOR AD Baxtran ABD Measuring Scales - icon89

SPEED 1 standard speed 15Hz.
SPEED 2 high speed 30Hz.
SPEED 3 low speed 7.5Hz
*This function stays blocked when UF-5 is in mode HOLD 1.
*The value of factory is 1

DISPLAYDESCRIPTION AND SEQUENCE OF USE
Baxtran ABD Measuring Scales - icon96• Press the keyBaxtran ABD Measuring Scales - icon7 to start or the keyBaxtran ABD Measuring Scales - icon3 to exit from the menu and the scale is going to begin again automatically.
• Use the keyBaxtran ABD Measuring Scales - icon6 and then the keyBaxtran ABD Measuring Scales - icon7 to select the speed of the AD converter ( press the keyBaxtran ABD Measuring Scales - icon3 to exit from the configuration and go back to the menu LF4)
• Use the keysBaxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5 and thenBaxtran ABD Measuring Scales - icon7 . To continue with the other adjustment and press the key Baxtran ABD Measuring Scales - icon3to exit from the menu and the scale will start again automatically.

BLIND Baxtran ABD Measuring Scales - icon56

DISPLAYDESCRIPTION AND SEQUENCE OF USE
Baxtran ABD Measuring Scales - icon57ZP 0 OFF
ZP 1 One division will not show being at zero
ZP 2 Two division will not show being at zero
ZP 3 Three divisions will not show being at zero
ZP 4 Four divisions will not show being at zero
ZP 5 Five divisions will not show being at zero
*This function is going to be blocked when UF-5 is in mode HOLD 1
*The value of factory is ZP 0
Baxtran ABD Measuring Scales - icon97• Press the keyBaxtran ABD Measuring Scales - icon7 to start or the keyBaxtran ABD Measuring Scales - icon3 to exit from the menu and the scale will start again automatically.
• Use the keyBaxtran ABD Measuring Scales - icon6 and the keyBaxtran ABD Measuring Scales - icon7 to select how many divisions will show when in zero ( press ESC to exit from the configuration and go back to the menu LF4)
• Use the keys Baxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5 and then the keyBaxtran ABD Measuring Scales - icon7 to continue with other adjustment or press the keyBaxtran ABD Measuring Scales - icon3 to exit from the menu and the scale is going to start automatically.

  APPROVAL CONFIRMITY Baxtran ABD Measuring Scales - icon61

DISPLAYDESCRIPTION AND SEQUENCE OF USE
Baxtran ABD Measuring Scales - icon61warning 2nonE Version not approved
DON’T MODIFY, IN NO WAY, THIS PARAMETER. IT MUST ALWAYS BE CONFIGURED AT nonE.
The change of this parameter implies the blocking of some functionalities.

GRAVITY Baxtran ABD Measuring Scales - icon63
*Introduce the gravity of your zone before doing the first calibration.
*Introduce the gravity of destination after doing the calibration.
*The value of the gravity will be denied if it is bigger than 9.83217 ( gravity of the pole) or inferior to 9.78031 ( gravity of the equator).
Value of factory: 9.8035

DISPLAYDESCRIPTION AND SEQUENCE OF USE
Baxtran ABD Measuring Scales - icon64• Press the keyBaxtran ABD Measuring Scales - icon7 to continue or the keyBaxtran ABD Measuring Scales - icon3 to exit from the menu and the scale will start again automatically.
• The screen is going to show the number of pre-calibration during one second.
• Press the keyBaxtran ABD Measuring Scales - icon7 to continue.
• Use the keys Baxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5 andBaxtran ABD Measuring Scales - icon6 , 0 ~ 9 and then the keyBaxtran ABD Measuring Scales - icon7 to introduce the value of gravity ( press ESC to exit from the configuration and go back to the menu LF7)

INITIAL ZERO Baxtran ABD Measuring Scales - icon65

SETZY resets of the point of zero every time that the scale is begun again.
SETZN resets of the point zero OFF

DISPLAYDESCRIPTION AND SEQUENCE OF USE
Baxtran ABD Measuring Scales - icon66• Press the keyBaxtran ABD Measuring Scales - icon7 to start or the keyBaxtran ABD Measuring Scales - icon3 to exit from the menu and the scale will start again automatically.
• Use the keyBaxtran ABD Measuring Scales - icon6 and then press the keyBaxtran ABD Measuring Scales - icon7 to choose the initial zero mode. (Press theBaxtran ABD Measuring Scales - icon3 Key to exit of the configuration and go back to menu LF8)
• Use the keys Baxtran ABD Measuring Scales - icon4 ,Baxtran ABD Measuring Scales - icon5 and then the keyBaxtran ABD Measuring Scales - icon7 to continue with other adjustments  0 press theBaxtran ABD Measuring Scales - icon3 key to exit from the menu and the scale will start again automatically.

ERROR MESSAGES

MESSAGEDESCRIPTION
Baxtran ABD Measuring Scales - icon67lnitial zero too high (over 10% of max. cap).
Baxtran ABD Measuring Scales - icon68lnitial zero too high (over 10% of max. cap).
Baxtran ABD Measuring Scales - icon69Initial zero too high (approved model).
Baxtran ABD Measuring Scales - icon70Initial zero too low (less than 10% of max. cap).
Baxtran ABD Measuring Scales - icon71Initial zero too low (approved model).
Baxtran ABD Measuring Scales - icon72Unstable internal value.
Baxtran ABD Measuring Scales - icon73Internal code is not stable.
Baxtran ABD Measuring Scales - icon74Overloading
Baxtran ABD Measuring Scales - icon75Overloading
Baxtran ABD Measuring Scales - icon76Overloading
Baxtran ABD Measuring Scales - icon77Linearity correcting not well or linearity correcting cancel.
Baxtran ABD Measuring Scales - icon78Weight is too low.
Baxtran ABD Measuring Scales - icon79Internal code is too low.
Baxtran ABD Measuring Scales - icon80E2ROM abnormal.
Baxtran ABD Measuring Scales - icon81Total price is more than 999999.
Baxtran ABD Measuring Scales - icon82Battery volume is too low.
Baxtran ABD Measuring Scales - icon83Calibration failure, check the loadcell.

GUARANTEE

This scale has a warranty against all manufacture and material defects, for a period of a year starting with the delivery date.
During this period, GIROPES, will be in charge of the repairing of the scale.
This warranty does not include the damages done by overload or wrong use.
The warranty does not cover the delivery expenses necessary for the repair of the scale.

Baxtran logotrade mark propriety of:
Pol. Empordà Internacional C/ Molló, 3
17469 VILAMALLA – (Girona) SPAIN
T. (34) 972 527 212 – F. (34) 972 527 211Baxtran logo1Baxtran ABD Measuring Scales - icon Hosteleria10.com

Documents / Resources