1. Introduction
This manual provides comprehensive instructions for the setup, operation, maintenance, and troubleshooting of your Generic PH-W3988 Water Quality Monitor. This device is a multiparameter 6-in-1 analyzer designed for accurate measurement of pH, Electrical Conductivity (EC), Total Dissolved Solids (TDS), Salinity (SALT), Specific Gravity (S.G), and Temperature in various water samples. It features Wi-Fi connectivity for remote monitoring via a dedicated application.
2. Package Contents
Please verify that all items listed below are present in your package:
- PH-W3988 Water Quality Monitor Unit
- Power Adapter (specific to region, e.g., AU Plug)
- pH Electrode
- EC/TDS/SALT/S.G/Temp Electrode
- User Manual (this document)
- Calibration Solutions (pH 4.00, pH 6.86, pH 9.18, EC 1413 µS/cm, EC 12.88 mS/cm - may vary by package)
3. Technical Specifications
| Parameter | Range |
|---|---|
| pH | 0.00 – 14.00 pH |
| EC | 0 – 9000 µS/cm, 0 – 199.0 mS/cm |
| TDS | 0 – 19990 ppm, 0 – 199.0 ppt |
| SALT | 0 – 199.9 ppt, 0% – 20% |
| S.G (Specific Gravity) | 0.990 – 1.400 |
| Temperature | 0°C – 50°C (32°F – 122°F) |
- Connectivity: Wi-Fi (2.4 GHz)
- Dimensions: Approximately 131mm x 78mm
- Item Weight: Approximately 1.76 ounces (50 grams)
- Power Supply: AC/DC Adapter (Input: 100-240V, Output: 5V DC)

4. Setup Instructions
- Unpack the Device: Carefully remove all components from the packaging.
- Connect Electrodes:
- Connect the pH electrode to the designated port (usually marked "pH").
- Connect the EC/TDS/SALT/S.G/Temp electrode to its designated port (usually marked "EC/TDS/SALT").
- Ensure connections are secure to prevent inaccurate readings.
- Power Connection: Insert the power adapter into the monitor's power input port, then plug the adapter into a suitable electrical outlet. The device will power on automatically.
- Initial Rinse: Before first use, rinse the electrodes with distilled or deionized water to remove any storage solution or contaminants. Gently blot dry with a lint-free cloth.
- Download App: Search for the "Smart Life" or "Tuya Smart" app (refer to packaging or quick start guide for specific app name) on your smartphone's app store (iOS or Android) and install it.
5. Operating the Monitor
5.1 Basic Operation
- Power On: The device powers on automatically when connected to power.
- Mode Selection: Press the MODE button to cycle through the different measurement parameters displayed on the screen (e.g., TDS/EC/SALT/S.G/TEMP and pH/TEMP).
- Temperature Unit: Press the TEMP CAL/+ button briefly to switch between Celsius (°C) and Fahrenheit (°F) temperature units.
- Measurement: Immerse the electrodes into the water sample you wish to test. Ensure the electrode tips are fully submerged and free of air bubbles. Allow a few moments for the readings to stabilize on the display.
5.2 APP Remote Functionality
- Wi-Fi Connection:
- Ensure your monitor is powered on and within range of a 2.4 GHz Wi-Fi network.
- Open the installed "Smart Life" or "Tuya Smart" app (refer to packaging or quick start guide for specific app name) on your smartphone.
- Follow the in-app instructions to add a new device. You may need to put the monitor into pairing mode (often by pressing and holding the Wi-Fi button or a combination of buttons until the Wi-Fi indicator blinks rapidly).
- Once connected, the Wi-Fi icon on the monitor's display will become solid.
- Remote Monitoring: Through the app, you can view real-time readings, set alarms for specific parameter thresholds, and access historical data.
6. Calibration
Regular calibration is essential for maintaining the accuracy of your water quality monitor. Always use fresh, standard calibration solutions.
6.1 pH Calibration
- Preparation: Rinse the pH electrode with distilled water and gently blot dry. Prepare pH 6.86, pH 4.00, and pH 9.18 standard buffer solutions.
- First Point (pH 6.86): Immerse the pH electrode into the pH 6.86 buffer solution. Press and hold the CAL button until the display shows "CAL" or starts blinking. The device will automatically recognize the buffer and calibrate. Wait for the reading to stabilize and the calibration to complete (usually indicated by a stable reading or a confirmation message).
- Second Point (pH 4.00): Rinse the electrode again. Immerse it into the pH 4.00 buffer solution. Press and hold the CAL button. The device will calibrate to the second point.
- Third Point (pH 9.18 - Optional): For higher accuracy, especially in alkaline ranges, repeat the process with pH 9.18 buffer solution.
- Completion: After calibration, rinse the electrode and store it properly.
6.2 EC/TDS/SALT/S.G Calibration
- Preparation: Rinse the EC electrode with distilled water and gently blot dry. Prepare a standard EC calibration solution (e.g., 1413 µS/cm or 12.88 mS/cm).
- Calibration: Immerse the EC electrode into the standard solution. Press and hold the CAL button (or the TEMP CAL/+ button if it doubles as EC calibration, refer to specific device markings). The device will automatically calibrate. Wait for the reading to stabilize.
- Completion: After calibration, rinse the electrode and store it properly.
Note: TDS, SALT, and S.G values are typically derived from the EC measurement and temperature compensation. Calibrating EC usually calibrates these parameters simultaneously.
7. Maintenance
- Electrode Cleaning: Regularly rinse electrodes with distilled water after each use. If electrodes become dirty or readings are erratic, clean them gently with a soft brush and a mild cleaning solution (e.g., diluted detergent or specific electrode cleaning solution), then rinse thoroughly.
- Electrode Storage: Always store pH electrodes in a suitable storage solution (e.g., 3M KCl solution or pH 4 buffer). Do not store pH electrodes dry. EC electrodes can typically be stored dry after rinsing.
- Device Cleaning: Wipe the main unit with a soft, damp cloth. Do not use abrasive cleaners or immerse the main unit in water.
- Electrode Replacement: Electrodes have a limited lifespan and may need replacement if they no longer calibrate correctly or provide stable readings.
8. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| No power/Display off | Power adapter disconnected or faulty. | Check power connections. Ensure adapter is working. |
| Inaccurate or unstable readings |
|
|
| Wi-Fi connection failure |
|
|
| "Err" or "---" displayed | Measurement out of range or electrode fault. | Check if the sample is within the device's measurement range. Clean or replace electrodes if necessary. |
9. Warranty Information
This Generic PH-W3988 Water Quality Monitor is covered by a standard manufacturer's warranty against defects in materials and workmanship for a period of one year from the date of purchase. This warranty does not cover damage caused by misuse, accident, unauthorized modification, neglect, or normal wear and tear of consumable parts like electrodes. Please retain your proof of purchase for warranty claims.
10. Customer Support
For technical assistance, troubleshooting, or warranty inquiries, please contact your retailer or the manufacturer's customer support. Refer to the product packaging or the retailer's website for specific contact details.
Please have your model number (PH-W3988) and purchase date available when contacting support.