Ultrasonic Thickness Gauge AR850 User Manual

Model: AR850 | Brand: KSKSK

1. Introduction

The KSKSK Ultrasonic Thickness Gauge AR850 is a precision instrument designed for measuring the thickness of various hard materials. It utilizes ultrasonic technology to provide accurate and non-destructive measurements, making it suitable for applications involving steel, cast iron, aluminum, copper, zinc, quartz glass, polyethylene, PVC, gray cast iron, and ductile iron. This manual provides detailed instructions for the proper setup, operation, and maintenance of your device to ensure optimal performance and longevity.

2. Safety Information

Please read and understand all safety instructions before operating the device. Failure to do so may result in injury or damage to the instrument.

3. Package Contents

Upon opening the package, please verify that all items listed below are present and in good condition:

Contents of the Ultrasonic Thickness Gauge AR850 package, including the main unit, probe, coupling agent, calibration certificate, and other accessories.

Figure 3.1: Complete Package Contents

This image displays the full contents of the AR850 package, including the main gauge unit, the ultrasonic probe, a bottle of coupling agent, a calibration certificate, and a small metal block used for calibration. Please note that while batteries are shown, they are typically not included with the product.

4. Product Overview

Familiarize yourself with the components of the Ultrasonic Thickness Gauge AR850.

Front view of the Ultrasonic Thickness Gauge AR850 with its connected probe.

Figure 4.1: Main Unit and Probe

This image shows the AR850 main unit with its display screen and control buttons, connected to the ultrasonic probe via a cable. The probe is designed to transmit and receive ultrasonic waves for thickness measurement.

Angled view of the Ultrasonic Thickness Gauge AR850, highlighting the display and control buttons.

Figure 4.2: Gauge Display and Controls

This image provides a closer look at the AR850's display, which shows thickness readings and other indicators, along with the various control buttons for operation, including power, measurement mode, and data functions.

Key Components:

5. Setup

5.1 Battery Installation

  1. Locate the battery compartment on the back of the main unit.
  2. Open the battery compartment cover.
  3. Insert three (3) AAA Ni-Cd or Ni-Hm batteries, ensuring correct polarity (+/-) as indicated inside the compartment.
  4. Close the battery compartment cover securely.

5.2 Probe Connection

  1. Connect the two connectors of the ultrasonic probe cable to the corresponding ports on the top of the main unit. Ensure a firm connection.

5.3 Initial Power On and Zero Adjustment

  1. Press the Power button to turn on the device. The display will illuminate.
  2. The device features an Automatic Zero Adjustment function. Allow a few moments for the device to stabilize upon power-on.
  3. For precise measurements, it is recommended to perform a calibration using the included calibration block before first use and periodically thereafter. Refer to the 'Operating Instructions' section for calibration details.

6. Operating Instructions

6.1 Powering On/Off

6.2 Preparing for Measurement

  1. Ensure the surface of the material to be measured is clean, smooth, and free of rust, paint, or other coatings that could interfere with the ultrasonic signal.
  2. Apply a small amount of coupling agent to the surface of the material where the measurement will be taken, or directly onto the probe face. The coupling agent is essential for transmitting the ultrasonic waves between the probe and the material.

6.3 Taking a Measurement

  1. Place the probe firmly and flatly onto the prepared surface of the material. Ensure there are no air gaps between the probe and the material.
  2. The device will automatically detect the thickness and display the reading on the LCD screen.
  3. Move the probe slightly to different spots on the material to take multiple readings and ensure consistency.

6.4 Switching Units (mm/m/s)

6.5 Data Storage Function

6.6 Sound Velocity Adjustment

The accuracy of thickness measurement depends on the correct sound velocity setting for the material being measured. The AR850 has a sound velocity range of 1000-9999 m/s.

6.7 Calibration

The AR850 features a high-precision linear compensation circuit (1% H + 0.1) and automatic zero adjustment. For optimal accuracy, periodic calibration is recommended.

  1. Ensure the probe is clean and free of coupling agent.
  2. Press the CAL button to enter calibration mode.
  3. Follow the on-screen prompts to calibrate the device using the provided calibration block. Typically, this involves placing the probe on the block and confirming the known thickness.
  4. Once calibration is complete, the device will return to measurement mode.

7. Maintenance

7.1 Cleaning

7.2 Storage

7.3 Battery Replacement

8. Troubleshooting

ProblemPossible CauseSolution
Device does not power onDead or incorrectly installed batteriesReplace batteries, ensuring correct polarity.
Inaccurate or unstable readings
  • Insufficient coupling agent
  • Rough or dirty surface
  • Incorrect sound velocity setting
  • Probe not flat on surface
  • Probe or cable damage
  • Apply more coupling agent
  • Clean and smooth the measurement surface
  • Adjust sound velocity for the material
  • Ensure firm, flat contact with the surface
  • Inspect probe/cable for damage; replace if necessary
Low Battery IndicationBatteries are lowReplace all three AAA batteries.
No display or partial displayDevice malfunctionTry resetting the device by removing and reinserting batteries. If problem persists, contact customer support.

9. Specifications

ParameterValue
Measuring Range1.2-225.0mm (Steel)
Accuracy+/-(1%+/-0.1)mm
Pipe Measuring Lower Limit20*3mm (Steel)
Sound Velocity Range1000-9999m/s
Working Frequency5MHz
Working Temperature Range0-45°C
Automatic Zero AdjustmentYes
Data Store FunctionUp to 10 test results
Auto Power Shut OffYes
Low Battery Indication3V+/-0.2V
Power Supply3*AAA Ni-Cd or Ni-Hm Battery (Not included)
Item Weight3.31 pounds
Package Dimensions1.18 x 0.79 x 0.39 inches
MaterialPlastic

10. Warranty Information

Specific warranty details for the KSKSK Ultrasonic Thickness Gauge AR850 are not provided in this manual. For information regarding warranty coverage, duration, and claims procedures, please refer to the warranty card included with your product, or contact the retailer or manufacturer directly. Please retain your proof of purchase for any warranty-related inquiries.

11. Support

If you encounter any issues or have questions regarding the operation, maintenance, or troubleshooting of your Ultrasonic Thickness Gauge AR850 that are not covered in this manual, please contact the seller or manufacturer for assistance. Provide your product model number (AR850) and a detailed description of the issue when seeking support.