1. Introduction

This user manual provides essential information for the safe and effective operation of your LungSank Oxygen Concentrator. Please read this manual thoroughly before using the device and keep it for future reference.

LungSank Oxygen Concentrator in a home setting with a person sleeping

The LungSank Oxygen Concentrator is designed for home use, providing high-purity oxygen.

2. Safety Information

Important Safety Precautions:

  • Do not smoke or allow open flames near the oxygen concentrator or oxygen tubing. Oxygen supports combustion.
  • Keep the device away from heat sources and direct sunlight.
  • Ensure proper ventilation around the device. Do not block air inlets or outlets.
  • Use only distilled or purified water in the humidifier bottle.
  • Do not immerse the device in water or other liquids.
  • Unplug the device before cleaning or maintenance.
  • This device is for supplemental oxygen therapy and is not intended for life support. Consult a medical professional for appropriate oxygen therapy settings.

3. Package Contents

Please check the package contents upon unboxing:

  • Oxygen Concentrator Unit
  • Power Cord
  • Remote Control
  • Nasal Cannula (Disposable)
  • Humidifying Water Cup
  • Atomization Package (Nebulizer Kit)
  • Filter
  • User Manual
Diagram showing all components included in the LungSank Oxygen Concentrator package

All components included with your LungSank Oxygen Concentrator.

4. Product Overview

Familiarize yourself with the main parts and controls of your oxygen concentrator.

Close-up view of the control panel of the LungSank Oxygen Concentrator with labeled buttons and display

Control panel and display features.

Control Panel Features:

  1. Normal Working State Indicator
  2. Machine Failure Indicator
  3. Low Oxygen Concentration Indicator (below 82%)
  4. Voice Indicator Light
  5. Negative Ion Indicator Light
  6. Oxygen Concentration Display
  7. Flow Display
  8. Countdown and Time Display
  9. SOS One-Button Help Button
  10. Voice Switch Button
  11. Oxygen Making Switch Button
  12. Flow Regulation Button
  13. Oxygen Making Timing Setting Button

5. Setup

5.1 Initial Placement

Place the oxygen concentrator on a flat, stable surface. Ensure there is at least 10 cm of clear space around all sides for adequate ventilation.

5.2 Power Connection

Connect the power cord to the device and then plug it into a grounded electrical outlet.

5.3 Humidifier Bottle Setup (Optional for Humidified Oxygen)

  1. Unscrew the humidifying water cup from the device.
  2. Fill the cup with distilled or purified water up to the "MAX" line. Do not overfill.
  3. Screw the humidifying water cup back onto the device securely.
  4. Connect the nasal cannula to the oxygen outlet on the humidifier bottle.
Close-up of the humidifying water cup being filled with water

Filling the humidifying water cup.

Note: You can also use the oxygen function directly without adding water if humidification is not desired.

6. Operating Instructions

6.1 Power On/Off

Press the "Oxygen Making Switch Button" on the control panel or the remote control to turn the device on or off.

6.2 Adjusting Oxygen Flow

Use the "Flow Regulation Button" on the control panel or the remote control to adjust the oxygen flow rate from 2 to 9 L/min. The current flow rate will be displayed on the screen.

Chart showing oxygen concentration levels at different flow rates from 2L to 9L

Oxygen concentration varies with flow rate. At 2L/min, concentration reaches 96%.

6.3 Setting the Timer

Press the "Oxygen Making Timing Setting Button" to set the operating time. Each press increases the duration by 30 minutes, up to a maximum of 240 minutes. The remaining time will be displayed.

Image showing the oxygen concentrator in a bedroom setting with a 240-minute timer graphic

The device features a timer for convenient use, allowing for continuous operation for up to 240 minutes.

6.4 Nebulization Function

  1. Locate the atomization port on the back of the device.
  2. Unscrew the protective cap from the atomization port.
  3. Connect the nebulizer kit tubing to the atomization port.
  4. Add the prescribed medication to the nebulizer cup.
  5. Turn on the device and select the nebulization function (if applicable, or it may activate automatically when connected).
Image showing the nebulization port and the nebulizer kit components

The device supports nebulization for medication delivery.

6.5 Remote Control Usage

The included remote control allows for convenient operation from a distance. All main functions (power, flow adjustment, timer) can be controlled via the remote.

This video demonstrates the remote control functionality and oxygen concentration adjustment of the LungSank Oxygen Concentrator.

7. Maintenance

Regular maintenance ensures optimal performance and longevity of your device.

7.1 Cleaning the Exterior

Wipe the exterior of the device with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the device is unplugged before cleaning.

7.2 Humidifier Bottle Cleaning

Clean the humidifier bottle daily with mild soap and water. Rinse thoroughly and allow to air dry completely before reattaching.

7.3 Filter Replacement

The device features an 8-stage filtration system. Regularly check and replace filters as indicated by the device or as per the recommended schedule in the full manual.

Diagram illustrating the 8-stage filtration system of the oxygen concentrator

The 8-stage filtration system ensures clean and safe oxygen.

8. Troubleshooting

ProblemPossible CauseSolution
Device does not power onPower cord not connected; Power outlet faultyCheck power cord connection; Try a different outlet
Low oxygen concentration alarmFilter clogged; Device malfunctionClean or replace filter; Contact customer support
Unusual noiseDevice not on a flat surface; Internal issueEnsure stable placement; Contact customer support
No mist from nebulizerNebulizer cup empty; Tubing kinkedAdd medication; Check tubing for kinks

If the problem persists after attempting these solutions, please contact LungSank customer support.

9. Specifications

FeatureDetail
BrandLungSank
ModelY102W
Oxygen ConcentrationUp to 96% (at 2L/min)
Adjustable Flow Rate2-9 L/min
Noise Level42 dB
Timer FunctionUp to 240 minutes (30-minute increments)
Dimensions (Packaging)37.1 x 30.8 x 23.9 cm
Weight (Packaging)6.34 kg
FeaturesLCD Screen, Remote Control, SOS Function, Voice Broadcast, Nebulization Function, 8-Stage Filtration, Negative Ion Generator
Country of OriginChina

10. Warranty and Support

Your LungSank Oxygen Concentrator comes with a standard manufacturer's warranty. For details regarding warranty coverage and duration, please refer to the warranty card included in your package or visit the official LungSank website.

For technical support, troubleshooting assistance, or replacement parts, please contact LungSank customer service through the retailer where you purchased the product or via the contact information provided on the LungSank brand store page.

Visit the LungSank Brand Store

© 2024 LungSank. All rights reserved.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.