Introduction

The DLAB Hipette Single-Channel Adjustable Volume Micropipette is a precision instrument designed for accurate and reproducible liquid handling in laboratory settings. This model, with a volume range of 0.1-2.5µl, features a lightweight, ergonomic design to minimize user fatigue and prevent repetitive strain injuries. It incorporates a clear digital display for precise volume setting and a volume lock function for enhanced accuracy.

Each Hipette micropipette is calibrated in accordance with ISO8655 standards and supplied with an individual test certificate, ensuring reliability and performance. Constructed from high-quality, UV-resistant materials, the pipette is fully autoclavable for easy sterilization and maintenance.

Key Features:

  • Premium, UV-resistant, lightweight, and ergonomic design.
  • 4-digit digital display for clear volume setting.
  • Volume lock function for higher precision and accuracy.
  • Calibrated in accordance with ISO8655, with an individual test certificate included.
  • Fully autoclavable material for easy sterilization.

Product Overview

DLAB Hipette Single-Channel Adjustable Volume Micropipette

Figure 1: DLAB Hipette Single-Channel Adjustable Volume Micropipette. This image shows the complete pipette with its light blue and white body, clear digital display, and a transparent tip attached.

Setup and Initial Use

  1. Unpacking: Carefully remove the micropipette from its packaging. Inspect for any visible damage.
  2. Review Certificate: Locate and review the individual test certificate provided with your pipette. This document confirms its calibration and adherence to ISO8655 standards.
  3. Tip Attachment: Select a suitable pipette tip for the 0.1-2.5µl volume range. Firmly press the pipette shaft into the tip opening with a slight twisting motion to ensure a secure, airtight seal. Do not overtighten.
  4. Initial Volume Check: Before first use, set the desired volume and perform a few aspiration/dispensing cycles with distilled water to prime the pipette and ensure proper function.

Operating Instructions

1. Setting the Volume:

  • Rotate the volume adjustment knob (located at the top of the pipette) to set the desired volume.
  • Read the volume clearly on the digital display. Ensure the volume is within the pipette's specified range (0.1-2.5µl).
  • Once the desired volume is set, engage the volume lock function (if available on your model) to prevent accidental changes during pipetting.

2. Aspiration (Drawing Liquid):

  1. Press the operating button down to the first stop (the point of resistance).
  2. Immerse the pipette tip into the liquid to be aspirated, ensuring it is just below the surface.
  3. Slowly release the operating button to allow the liquid to be drawn into the tip. Wait a moment for the liquid to fully enter.
  4. Remove the tip from the liquid, touching it against the edge of the container to remove any excess droplets.

3. Dispensing (Releasing Liquid):

  1. Place the pipette tip against the inner wall of the receiving vessel at a slight angle.
  2. Press the operating button down to the first stop to dispense the main volume of liquid.
  3. Continue pressing the operating button down to the second stop (blow-out) to expel any remaining liquid from the tip.
  4. While holding the button at the second stop, withdraw the pipette from the vessel.

4. Tip Ejection:

  • After dispensing, point the pipette into a waste container.
  • Press the tip ejector button (if present) to release the used tip.
  • Dispose of the tip according to laboratory safety protocols.

Maintenance

Cleaning:

  • The DLAB Hipette is designed with fully autoclavable materials. For sterilization, the entire pipette can be autoclaved at 121°C (250°F) for 20 minutes.
  • After autoclaving, allow the pipette to cool completely before use.
  • For routine cleaning, wipe the exterior of the pipette with a soft cloth dampened with mild detergent or ethanol. Avoid harsh solvents.

Calibration:

  • Regular calibration checks are recommended to ensure continued accuracy, especially after extensive use or autoclaving.
  • Refer to ISO8655 guidelines or consult a qualified calibration service for detailed calibration procedures.

Storage:

  • Store the pipette vertically in a pipette stand when not in use to prevent contamination and damage to the tip cone.
  • Store in a clean, dry environment, away from direct sunlight and extreme temperatures.

Troubleshooting

ProblemPossible CauseSolution
Inaccurate Volume / Leaking
  • Loose or damaged tip
  • Contaminated tip cone
  • Internal seal wear
  • Ensure tip is securely attached. Replace if damaged.
  • Clean the tip cone.
  • Contact DLAB support for service.
Operating Button Sticking
  • Accumulation of dust or debris
  • Lack of lubrication
  • Clean the operating button area.
  • Contact DLAB support for maintenance.
Volume Adjustment Difficulty
  • Debris in mechanism
  • Internal component issue
  • Gently try to rotate. Do not force.
  • Contact DLAB support for service.

If the problem persists after attempting these solutions, please contact DLAB customer support or your local distributor for further assistance.

Specifications

AttributeDetail
Product NameHipette Single-Channel Adjustable Volume Micropipette
BrandDLAB
Model Number7030511001
Volume Range0.1 - 2.5 µl
CalibrationISO8655 compliant, individual test certificate included
AutoclavableYes, fully autoclavable
Package Dimensions11 x 4.15 x 2.57 inches
Item Weight1.4 Pounds

Warranty and Support

For specific warranty terms and conditions, please refer to the documentation included with your purchase or contact the authorized DLAB distributor or seller from whom you purchased this product. Warranty coverage typically addresses manufacturing defects and material flaws under normal use.

For technical support, service, or inquiries regarding spare parts and accessories, please contact your local DLAB representative or the customer service department of the seller. When contacting support, please have your product model number (7030511001) and purchase details readily available.

Manufacturer: DLAB Product

© 2024 DLAB Product. All rights reserved.

This manual is for informational purposes only. Specifications are subject to change without notice.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.