FLB FORELIBRA FLBNJ-AL-PAN

FLB FORELIBRA Moisture Analyzer User Manual

Model: FLBNJ-AL-PAN

1. Introduction

The FLB FORELIBRA Moisture Analyzer is a precision instrument designed for accurate determination of moisture content in various samples. Featuring a 5-inch touch screen, 3 drying modes, and RS232 interface, it offers reliable and efficient analysis for laboratory, industrial, and research applications. This manual provides essential information for the safe and effective operation of your moisture analyzer.

FLB FORELIBRA Moisture Analyzer, closed, showing the touch screen and top vent.

Figure 1: FLB FORELIBRA Moisture Analyzer (Closed View)

2. Safety Information

Please read all safety instructions carefully before operating the device to prevent injury or damage to the instrument.

  • Ensure the power supply is 110V as specified.
  • Do not touch the heating element during or immediately after operation, as it will be hot.
  • Operate the analyzer in a well-ventilated area.
  • Avoid spilling liquids on the device, especially near electrical components.
  • Only use the provided accessories and sample pans.
  • Keep the device away from flammable materials.

3. Package Contents

Verify that all items are present in the package:

  • Moisture Analyzer Unit
  • Plastic Cover
  • Aluminum Sample Pans
  • Load Cell (integrated)
  • Touch Screen (integrated)
  • Power Cable
  • User Manual

4. Product Overview

The FLB FORELIBRA Moisture Analyzer integrates a precision balance with a heating unit to measure moisture content by drying a sample and calculating the weight loss. Key components include the heating chamber, sample pan, load cell, and the intuitive touch screen interface.

FLB FORELIBRA Moisture Analyzer with the lid open, showing the heating element and a sample pan with coffee beans.

Figure 2: Moisture Analyzer with Heating Chamber Open

Four views of the FLB FORELIBRA Moisture Analyzer: front, back, side, and angled, highlighting various ports and controls.

Figure 3: Multiple Views of the Analyzer

5. Setup

  1. Unpacking: Carefully remove the moisture analyzer from its packaging. Retain packaging for future transport or storage.
  2. Placement: Place the analyzer on a stable, level surface, away from direct sunlight, drafts, and vibrations. Ensure adequate space for ventilation.
  3. Leveling: Use the adjustable feet at the bottom of the unit to level the analyzer. Check the bubble level (if present) to ensure precise leveling.
  4. Power Connection: Connect the power cable to the analyzer's power input and then to a grounded 110V AC outlet.
  5. Initial Power On: Press the power button. The touch screen will illuminate and display the startup sequence.
FLB FORELIBRA Moisture Analyzer with key dimensions labeled in inches.

Figure 4: Product Dimensions

6. Operating Instructions

The moisture analyzer features a user-friendly 5-inch touch screen interface for easy operation.

Close-up of the FLB FORELIBRA Moisture Analyzer's touch screen displaying drying parameters and the top vent.

Figure 5: Touch Screen Interface

6.1 Basic Operation

  1. Prepare Sample: Place a clean, dry aluminum sample pan on the pan support.
  2. Tare: Close the heating chamber lid and press the 'Tare' button on the touch screen to zero the balance.
  3. Add Sample: Open the lid and carefully place your sample onto the sample pan. Ensure the sample is evenly distributed.
  4. Close Lid: Close the heating chamber lid.
  5. Select Program: The analyzer has 34 memory programs. Select an appropriate program for your sample type or configure new settings.
  6. Set Drying Parameters: Adjust the drying temperature (40℃ to 180℃) and drying mode (Automatic or Timing) via the touch screen. The analyzer can control drying speed (soft or fast).
  7. Start Analysis: Press the 'Start' button to begin the drying process. The screen will display real-time drying data, including drying time, current temperature, and moisture percentage.
  8. End of Analysis: The analysis will automatically stop when the selected drying mode criteria are met. The final moisture content will be displayed.
  9. Record Data: Data can be saved to memory or exported via the RS232 interface to a computer or printer.

6.2 Drying Modes

  • Automatic Mode: The analyzer automatically detects when the sample has reached a constant weight, indicating complete drying.
  • Timing Mode: The drying process runs for a pre-set duration. This mode is suitable for samples with known drying characteristics.

7. Maintenance

Regular maintenance ensures the longevity and accuracy of your moisture analyzer.

  • Cleaning: Disconnect the power before cleaning. Use a soft, damp cloth to wipe the exterior. For the heating chamber and sample pan, ensure they are cool before cleaning. Do not use abrasive cleaners or solvents.
  • Calibration: Periodically calibrate the balance using certified calibration weights to maintain accuracy. Refer to the on-screen menu for calibration procedures.
  • Sample Pan Replacement: Replace aluminum sample pans after each use.
  • Storage: When not in use, keep the plastic cover on the analyzer to protect it from dust and debris. Store in a dry, stable environment.

8. Troubleshooting

ProblemPossible CauseSolution
Analyzer does not power onNo power supply; loose cableCheck power cable connection; ensure outlet is functional.
Inaccurate readingsNot leveled; calibration needed; drafts/vibrationsLevel the analyzer; perform calibration; move to a stable environment.
Heating element not workingTemperature not set; internal faultVerify temperature settings; contact support if issue persists.
Touch screen unresponsiveTemporary software glitchRestart the device.

9. Specifications

FeatureDetail
Model NumberFLBNJ-AL-PAN
Capacity50g / 110g (depending on variant)
Readability0.001g (0.01%)
Display TypeLCD Touch Screen (5-inch)
Drying ModesAutomatic, Timing
Heating Temperature Range40℃ to 180℃
Power Supply110V
InterfaceRS232
Memory Programs34
Product Dimensions14.57 x 8.46 x 7.68 inches
Weight10.85 Pounds
CertificationsCE Certified

10. Warranty and Support

FLB FORELIBRA products are manufactured to high-quality standards and undergo rigorous testing. Each scale is tested and calibrated before shipment.

For warranty information, technical support, or service inquiries, please refer to the contact information provided with your purchase documentation or visit the official FLB FORELIBRA website. Please have your model number (FLBNJ-AL-PAN) and purchase date available when contacting support.

© 2023 FLB FORELIBRA. All rights reserved.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.