INFURIDER YF-36K Digital Multimeter User Manual

Model: YF-36K

1. Introduction

This user manual provides detailed instructions for the safe and effective operation of the INFURIDER YF-36K Digital Multimeter. Please read this manual thoroughly before using the device to ensure proper functionality and to prevent potential hazards.

Two INFURIDER YF-36K Digital Multimeters

Figure 1: INFURIDER YF-36K Digital Multimeter (Two units shown)

The image displays two INFURIDER YF-36K digital multimeters, side-by-side. Each multimeter is red and black, featuring a large LCD screen, a central rotary switch, and input jacks at the bottom. The display shows "000.0" and "AUTO DC".

2. Safety Information

Always adhere to the following safety precautions to avoid personal injury or damage to the multimeter or equipment under test.

  • Do not exceed the maximum input values for any range.
  • Use caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC. These voltages pose a shock hazard.
  • Disconnect the test leads from the circuit before changing functions.
  • Ensure the correct function and range are selected for the measurement.
  • Do not operate the multimeter if it appears damaged or if the case is open.
  • Replace the battery when the low battery indicator appears to ensure accurate readings.
  • This device is rated CAT III 600V.

3. Product Features

The INFURIDER YF-36K Digital Multimeter offers a range of features designed for versatile electrical testing:

  • Max 4000 counts display for precise readings.
  • Auto-ranging capability for ease of use.
  • Measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, and Temperature.
  • Includes Frequency, hFE test, Diode, and Continuity test functions.
  • LCD Display with Backlight for visibility in dimly lit areas.
  • Auto Power Off function to conserve battery life.
  • Low Battery Indication.
  • Data Hold function for convenient reading.
  • Overload Protection on all ranges.
  • Comfortable hand-holding size with anti-slip rubberized case.
  • Built-in kickstand for angled viewing.
Labeled diagram of INFURIDER YF-36K Multimeter components

Figure 2: Multimeter Components Overview

This diagram labels key components of the INFURIDER YF-36K multimeter, including the LCD Display, Function Selector Switch, Rotary Switch, 10A Input Jack, COM Input Jack, and VΩmA Hz Input Jack. It also highlights the Data-hold/Back Light Switch and provides icons for DC&AC Voltage, DC&AC Current, Capacitance, Temperature, and Backlight functions.

4. Package Contents

Verify that all items are present in your package:

  • 2 x INFURIDER YF-36K Digital Multimeter
  • 2 x User Manual
  • 2 x Temperature Probe
  • 2 x Test Lead (2 Pairs)
  • 2 x 9V (6F22) Battery
  • 2 x Protective Bag
Contents of the INFURIDER YF-36K Multimeter package

Figure 3: Package Contents

The image shows the complete contents of the INFURIDER YF-36K package, including two YF-36K multimeters, two user manuals, two temperature probes, two pairs of test leads, two 9V batteries, and a package box.

5. Setup

5.1. Battery Installation

  1. Locate the battery compartment cover on the back of the multimeter.
  2. Use a screwdriver to open the battery compartment.
  3. Insert the included 9V (6F22) battery, observing the correct polarity (+ and -).
  4. Replace the battery compartment cover and secure it with the screw.

5.2. Test Lead Connection

  • For most measurements (Voltage, Resistance, Capacitance, Frequency, Diode, Continuity, Temperature, and mA/µA Current), insert the red test lead into the "VΩmA Hz" jack and the black test lead into the "COM" jack.
  • For 10A current measurements, insert the red test lead into the "10A" jack and the black test lead into the "COM" jack.

6. Operating Instructions

To begin operation, turn the rotary switch to the desired measurement function. The multimeter will automatically select the appropriate range (auto-ranging).

6.1. Power On/Off

  • To power on, rotate the function switch from "OFF" to any desired measurement function.
  • To power off, rotate the function switch back to "OFF".
  • The multimeter features an Auto Power Off function, which will turn off the device after approximately 15 minutes of inactivity to save battery life.

6.2. Backlight and Data Hold

  • Press the HOLD button briefly to activate the Data Hold function, which freezes the current reading on the display. Press again to release.
  • Press and hold the HOLD button for over 2 seconds to turn the LCD backlight on or off.

6.3. Function Selection (SELECT Button)

In certain rotary switch positions, the SELECT button allows you to toggle between different measurement modes within that function. For example, in the voltage range, you might switch between AC and DC voltage if not auto-detected, or between Diode and Continuity test modes.

6.4. Measurement Functions

6.4.1. DC & AC Voltage Measurement

  1. Set the rotary switch to the V~ (AC Voltage) or V- (DC Voltage) position. The multimeter will auto-range.
  2. Connect the red test lead to the positive (+) side of the circuit and the black test lead to the negative (-) side or ground.
  3. Read the voltage value on the LCD display.
Demonstration of DC & AC Voltage Measurement with INFURIDER YF-36K

Figure 4: DC & AC Voltage Measurement

This image illustrates the process of measuring DC and AC voltage using the INFURIDER YF-36K multimeter. One multimeter is shown measuring the voltage of a 9V battery, displaying "5.40 V". Another multimeter is shown measuring AC voltage from a wall outlet, displaying "21.30 V".

6.4.2. DC & AC Current Measurement

  1. Set the rotary switch to the A~ (AC Current) or A- (DC Current) position, or the mA/µA position.
  2. For 10A measurements, ensure the red test lead is in the "10A" jack. For mA/µA measurements, ensure the red test lead is in the "VΩmA Hz" jack.
  3. Open the circuit where current is to be measured and connect the multimeter in series with the load.
  4. Read the current value on the LCD display.

6.4.3. Resistance Measurement

  1. Set the rotary switch to the Ω (Resistance) position.
  2. Ensure the circuit is de-energized before measuring resistance.
  3. Connect the test leads across the component to be measured.
  4. Read the resistance value on the LCD display.

6.4.4. Capacitance Measurement

  1. Set the rotary switch to the Cap (Capacitance) position.
  2. Ensure the capacitor is fully discharged before measurement to prevent damage to the multimeter.
  3. Connect the test leads across the capacitor terminals.
  4. Read the capacitance value on the LCD display.

6.4.5. Frequency Measurement

  1. Set the rotary switch to the Hz (Frequency) position.
  2. Connect the test leads across the signal source.
  3. Read the frequency value on the LCD display.

6.4.6. Temperature Measurement

  1. Set the rotary switch to the Temp (Temperature) position.
  2. Insert the temperature probe into the "VΩmA Hz" and "COM" jacks, observing polarity.
  3. Place the tip of the temperature probe on or in the object whose temperature is to be measured.
  4. Read the temperature value on the LCD display.
Demonstration of Temperature Measurement with INFURIDER YF-36K using a K-type thermocouple

Figure 5: Temperature Measurement

This image shows the INFURIDER YF-36K multimeter being used to measure temperature. A K-type thermocouple is connected to the multimeter and submerged in a glass of water. The multimeter's display shows "0030" degrees Celsius.

6.4.7. Diode Test

  1. Set the rotary switch to the Diode position.
  2. Connect the red test lead to the anode and the black test lead to the cathode of the diode.
  3. The display will show the forward voltage drop. Reverse the leads; an open circuit (OL) reading indicates a good diode.

6.4.8. Continuity Test

  1. Set the rotary switch to the Continuity position.
  2. Connect the test leads across the circuit or component.
  3. A continuous beep indicates a complete circuit (low resistance). The display will show the resistance value.

6.4.9. hFE Test (Transistor Test)

  1. Set the rotary switch to the hFE position.
  2. Insert the transistor's emitter, base, and collector leads into the corresponding holes in the hFE socket on the multimeter, ensuring correct NPN or PNP type.
  3. The display will show the hFE value (DC current gain).

7. Maintenance

7.1. Cleaning

Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Keep the input terminals free of dirt or debris.

7.2. Battery Replacement

When the low battery indicator appears on the display, replace the 9V battery as described in Section 5.1. Prompt battery replacement ensures continued accuracy and proper operation.

7.3. Fuse Replacement

If the current measurement function fails, the fuse may need replacement. Refer to the internal diagram in the full user manual for fuse specifications and replacement procedures. This procedure should only be performed by qualified personnel.

8. Troubleshooting

ProblemPossible CauseSolution
No display or dim displayLow or dead battery; Auto Power Off activated.Replace battery; Turn rotary switch to ON position.
"OL" (Overload) displayedInput value exceeds selected range; Open circuit in resistance/continuity.Select a higher range (if manual ranging); Check circuit connection.
Incorrect readingsIncorrect function/range selected; Poor test lead connection; Low battery.Verify settings; Ensure secure connections; Replace battery.
Current measurement not workingBlown fuse.Replace fuse (refer to manual for details).

9. Specifications

ParameterRange / ValueAccuracy
Display4000 Counts LCD with Backlight
DC Voltage400mV/4V/40V/400V/1000V+/-0.8%+5
AC Voltage4V/40V/400V/750V+/-1.2%+8
DC Current400uA/4000uA/40mA/400mA/10A+/-1.0%+5
AC Current400uA/4000uA/40mA/400mA/10A+/-1.2%+8
Resistance400Ω/4KΩ/40KΩ/400KΩ/4MΩ/40MΩ+/-1.2%+5
Capacitance4nF/40nF/400nF/4uF/40uF/400uF/4000uF+/-2.5%+5
Frequency9.999Hz to 9.999MHz+/-0.5%+15
Temperature-20℃ to 1000℃ / -4℉ to 1832℉+/-1℃
Diode TestYes
Continuity TestYes
hFE TestYes
Power Source9V (6F22) Battery
Safety RatingIEC 1010, CAT III 600V, CE standard
INFURIDER YF-36K Multimeter in use on a circuit board

Figure 6: Versatile Digital Multimeter in Use

This image shows the INFURIDER YF-36K digital multimeter being used to test components on a circuit board. The multimeter is placed on a green cutting mat alongside various electronic components and tools, demonstrating its application in troubleshooting electrical issues.

10. Warranty and Support

INFURIDER is committed to providing high-quality measuring tools. For any questions or support regarding your YF-36K Digital Multimeter, please refer to the contact information provided in the original product packaging or visit the official INFURIDER website.

The product comes with a standard warranty. Please retain your proof of purchase for warranty claims. Details of the warranty period and terms are typically included in the physical user manual provided with the product.

© 2024 INFURIDER. All rights reserved.