coczow PM2116S

coczow PM2116S Full-Automatic Clamp Meter User Manual

Model: PM2116S

1. Introduction

This manual provides detailed instructions for the safe and effective operation of the coczow PM2116S Full-Automatic Clamp Meter. This device is a compact, auto-ranging digital multimeter designed for measuring AC/DC voltage, AC/DC current, resistance, continuity, and frequency. It features True RMS measurement for accuracy and a Non-Contact Voltage (NCV) detection function for enhanced safety. Please read this manual thoroughly before use and retain it for future reference.

2. Safety Information

WARNING: To avoid electric shock or personal injury, please read and understand all safety information before using this product.

  • Always adhere to local and national safety codes.
  • Do not use the meter if it appears damaged or if the test leads are damaged.
  • Do not apply more than the rated voltage, as marked on the meter, between the terminals or between any terminal and earth ground.
  • Use caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC. These voltages pose a shock hazard.
  • Keep your fingers behind the finger guards on the test probes during use.
  • Do not operate the meter around explosive gas, vapor, or dust.
  • Before measuring current, ensure the meter is in the correct function and the test leads are properly connected.
  • Always disconnect the test leads from the circuit before changing functions.
  • Replace the batteries immediately when the low battery indicator appears to ensure accurate readings.

3. Product Overview

The coczow PM2116S is a smart clamp meter designed for ease of use and safety. Its compact design allows for convenient portability and operation.

coczow PM2116S Full-Automatic Clamp Meter with test leads and carrying pouch

Figure 3.1: The coczow PM2116S Full-Automatic Clamp Meter, shown with its included test leads, two AAA batteries, and a protective carrying pouch.

3.1 Key Components

  • Clamp Jaw: Used for non-contact AC/DC current measurement.
  • Function Buttons: For power, mode selection, data hold, and backlight.
  • LCD Display: Shows measurement readings, units, and function indicators.
  • Input Jacks: For connecting test leads (COM and VΩHz).
  • NCV Sensor: Detects non-contact voltage.
Close-up view of the coczow PM2116S clamp meter showing the display, buttons, and input jacks.

Figure 3.2: A detailed view of the clamp meter's front panel, highlighting the display, function buttons, and test lead input terminals.

3.2 LCD Display Indicators

Close-up of the clamp meter's LCD display showing various icons and measurement units.

Figure 3.3: The LCD display showing indicators for AC/DC, auto-ranging, NCV, data hold, battery level, and units like V, A, Hz, Ω, kΩ.

  • AUTO: Indicates automatic measurement mode.
  • NCV: Non-Contact Voltage detection active.
  • AC/DC: Indicates Alternating Current or Direct Current measurement.
  • H: Data Hold function active.
  • Battery Icon: Indicates battery level.
  • Units: V (Volts), A (Amperes), Hz (Hertz), Ω (Ohms), kΩ (Kilo-Ohms).

4. Setup

4.1 Battery Installation

The meter requires two 1.5V AAA batteries for operation. Ensure correct polarity when installing.

  1. Locate the battery compartment cover on the back of the meter.
  2. Use a screwdriver to open the battery compartment.
  3. Insert two 1.5V AAA batteries, observing the polarity markings (+ and -).
  4. Replace the battery compartment cover and secure it with the screw.

4.2 Connecting Test Leads

For voltage, resistance, continuity, and frequency measurements, connect the test leads as follows:

  • Insert the black test lead into the COM (common) input jack.
  • Insert the red test lead into the VΩHz input jack.
Package contents including the clamp meter, test leads, batteries, and user manual.

Figure 4.1: The complete package contents, showing the clamp meter, red and black test leads, two AAA batteries, and the user manual.

5. Operating Instructions

The coczow PM2116S features full-automatic recognition, simplifying operation by automatically identifying the measurement type (voltage, resistance, continuity, frequency) when test leads are connected. For current measurement, the clamp function is used.

Clamp meter displaying 'Fully Automatic Recognition' feature with icons for AC/DC Voltage, AC/DC Current, Resistance, Continuity, Frequency, and Voltage Alarm.

Figure 5.1: The meter's 'Fully Automatic Recognition' capability, which eliminates the need for manual range selection for common measurements.

5.1 Power On/Off

Press the red power button () to turn the meter on. The meter will automatically enter auto-ranging mode. Press and hold the power button to turn the meter off. The meter also features an auto-shutdown function after 10 minutes of inactivity to conserve battery life.

5.2 AC/DC Voltage Measurement

  1. Ensure the meter is powered on.
  2. Connect the black test lead to the COM jack and the red test lead to the VΩHz jack.
  3. Touch the test probes to the circuit points where voltage is to be measured. The meter will automatically detect AC or DC voltage and display the reading.
Clamp meter measuring AC voltage from a wall outlet using test leads, and measuring AC current by clamping around a wire.

Figure 5.2: Demonstrates precise AC/DC voltage measurement using the test leads inserted into a wall socket.

5.3 AC/DC Current Measurement (Clamp Function)

The clamp function allows for non-contact measurement of AC/DC current up to 100A.

  1. Ensure the meter is powered on.
  2. Open the clamp jaw by pressing the trigger.
  3. Place the single conductor (wire) through the center of the clamp jaw. Ensure the jaw is fully closed.
  4. The meter will display the AC or DC current reading. Note: To measure current, the conductor must be isolated; clamping around a combined power cord (live and neutral) will result in a zero reading.
Clamp meter measuring current by clamping around a single electrical wire within a circuit breaker panel.

Figure 5.3: Illustrates the correct method for measuring AC/DC current by clamping the meter around a single conductor.

5.4 Resistance Measurement

  1. Ensure the meter is powered on.
  2. Connect the black test lead to the COM jack and the red test lead to the VΩHz jack.
  3. Touch the test probes across the component or circuit where resistance is to be measured. The meter will automatically detect and display the resistance reading in Ohms (Ω) or Kilo-Ohms (kΩ).
  4. Ensure the circuit is de-energized before measuring resistance.

5.5 Continuity Test

  1. Ensure the meter is powered on.
  2. Connect the black test lead to the COM jack and the red test lead to the VΩHz jack.
  3. Touch the test probes to the two points of the circuit to be tested for continuity.
  4. If continuity exists (resistance is low), the meter will emit an audible beep and display a low resistance value.
Clamp meter test leads touching, indicating a continuity test with a buzzing icon.

Figure 5.4: The continuity test function, indicated by the buzzing icon, confirms a complete circuit when the test leads are connected.

5.6 Frequency Measurement

  1. Ensure the meter is powered on.
  2. Connect the black test lead to the COM jack and the red test lead to the VΩHz jack.
  3. Touch the test probes to the circuit where frequency is to be measured. The meter will automatically detect and display the frequency in Hertz (Hz).

5.7 Non-Contact Voltage (NCV) Detection

The NCV function allows for detection of AC voltage without direct contact with conductors, enhancing safety.

  1. Press the SEL/NCV button to activate NCV mode. The NCV indicator will appear on the display.
  2. Move the top part of the meter (where the NCV sensor is located) close to the conductor or outlet.
  3. If AC voltage is detected, the meter will emit an audible beep and the NCV indicator will flash. The frequency of beeps and flashes increases with stronger voltage.
Clamp meter performing non-contact voltage detection near a wall outlet, with concentric circles indicating the detection field.

Figure 5.5: Demonstrates the Non-Contact Voltage (NCV) detection feature, providing a safety warning without direct electrical contact.

5.8 Data Hold Function

To freeze the current reading on the display:

  • During a measurement, press the H/* button. The 'H' indicator will appear on the display.
  • Press the button again to release the data hold function.

5.9 Backlight Function

To illuminate the display in low-light conditions:

  • Press and hold the H/* button to turn on the backlight.
  • Press and hold the button again to turn off the backlight.
Clamp meter display with backlight activated and 'H' icon for data hold.

Figure 5.6: The meter's display with the backlight active for improved visibility and the 'H' indicator showing the data hold function is engaged.

6. Maintenance

6.1 Cleaning

Wipe the meter with a damp cloth and mild detergent. Do not use abrasives or solvents. Keep the input terminals free of dirt and moisture.

6.2 Battery Replacement

When the battery icon appears on the display, replace the batteries immediately to ensure accurate measurements. Follow the instructions in Section 4.1.

7. Troubleshooting

ProblemPossible CauseSolution
Meter does not power on.Dead or incorrectly installed batteries.Check battery polarity or replace batteries.
Inaccurate readings.Low battery, damaged test leads, or incorrect measurement technique.Replace batteries, inspect test leads, ensure proper connection and measurement method.
No current reading with clamp.Clamping around a combined power cord (live and neutral).Ensure only a single conductor is placed within the clamp jaw.
NCV function is too sensitive or not detecting.Environmental interference or distance from source.Move away from other electrical fields; ensure close proximity to the conductor.

8. Specifications

  • Model: PM2116S
  • Measurement Type: Full-Automatic Clamp Multimeter
  • Display: 6000 Counts
  • AC/DC Voltage: Up to 600V
  • AC/DC Current: Up to 100A (Resolution 10mA)
  • Resistance: Up to 6000Ω
  • Frequency: Up to 1000Hz
  • Continuity: Yes, with audible buzzer
  • NCV (Non-Contact Voltage): Yes
  • True RMS: Yes, for accurate measurements on non-linear signals
  • Data Hold: Yes
  • Backlight: Yes
  • Auto Shut-off: Yes (after 10 minutes)
  • Power Source: 2 x 1.5V AAA Batteries
  • Item Weight: 270 Grams
  • Package Dimensions: 6.97 x 4.06 x 1.34 inches
  • Safety Rating: IEC 61010-1:2000-1, CAT III 600V

9. Warranty and Support

The coczow PM2116S Full-Automatic Clamp Meter comes with a 1-YEAR warranty from the date of purchase. This warranty covers manufacturing defects and malfunctions under normal use. It does not cover damage caused by misuse, accidents, unauthorized modifications, or improper maintenance.

For technical support, warranty claims, or any product-related inquiries, please contact your retailer or the manufacturer's customer service. Please have your purchase receipt and product model number (PM2116S) available when contacting support.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.