Reliefband Reliefband Classic Wristband
Reliefband Classic Wristband Instruction Manual
Model: Reliefband Classic Wristband
Brand: Reliefband
Introduction
The Reliefband Classic Wristband is an FDA-cleared, drug-free wearable solution designed for the effective relief of nausea and vomiting. It works by stimulating the median nerve on the underside of the wrist, sending signals to the brain that normalize nerve messaging to the stomach, thereby rapidly relieving symptoms without side effects.
This device is clinically proven to provide fast-acting relief for various conditions including motion sickness (car, air, train, sea), morning sickness, migraines, and post-operative nausea.

Image: The Reliefband Classic Wristband shown with two tubes of hypoallergenic conductivity gel.

Image: Conditions for which Reliefband provides FDA-cleared relief, such as motion sickness, morning sickness, and post-operative nausea.
What's Included
- 1 x Reliefband Classic Wristband (with pre-installed batteries)
- 2 x Hypoallergenic Conductivity Gel Tubes (7.5ml / 0.25 fl. oz. each)
- 1 x Adjustable Nylon Strap
Setup and Application
Proper placement and application of the conductivity gel are crucial for the Reliefband Classic to function effectively.
- Find the Area: On either wrist, locate the area between the two central tendons. This spot is approximately two finger-widths from the first crease of your wrist.
- Apply the Gel: Clean the identified area on your wrist. Apply a small drop of the hypoallergenic conductivity gel and spread it in a circular motion, covering an area about the size of a large coin. This gel ensures optimal conductivity between the device and your skin.
- Place and Fasten: Position the Reliefband Classic directly over the gelled area. Secure the adjustable nylon strap snugly around your wrist.

Image: Step-by-step guide for applying the Reliefband Classic.
Operating Instructions
Once the Reliefband Classic is properly positioned and the gel applied, follow these steps to activate and adjust the device:
- Power On: Press the power button located on the device. The device will start at intensity level 1.
- Adjust Intensity: Increase the stimulation level by pressing the intensity up button until you feel a comfortable tingling sensation in your palm and middle finger. The Reliefband Classic has 5 intensity settings, allowing you to customize the treatment to your precise needs.
- Continuous Use: The device is designed for continuous use and can provide relief for up to 150 hours with its pre-installed batteries.
- Power Off: To turn off the device, press and hold the power button for 3 seconds.

Image: The Reliefband Classic in use, demonstrating its placement and controls.
Maintenance
To ensure the longevity and optimal performance of your Reliefband Classic, follow these maintenance guidelines:
- Cleaning: After each use, gently wipe the electrodes and the surrounding area of the device with a clean, damp cloth to remove any residual conductivity gel. Ensure the device is dry before storage.
- Battery Replacement: The Reliefband Classic uses 2 Lithium Metal batteries (included). The device features a low battery indicator. When the battery needs replacement, refer to the detailed instructions provided in the full user manual for safe and proper battery exchange. Note: Some users have reported difficulty with battery replacement; exercise caution and follow instructions carefully.
- Storage: Store the device in a cool, dry place away from direct sunlight and extreme temperatures.
- Gel Storage: Keep the conductivity gel tubes sealed when not in use to prevent drying.

Image: Key features of the Reliefband Classic, including its replaceable battery and splash resistance.
Troubleshooting
If you encounter issues with your Reliefband Classic, consider the following:
- No Sensation or Weak Sensation:
- Ensure sufficient conductivity gel is applied to the skin and spread evenly.
- Verify the device is placed correctly over the gelled area, between the two central tendons.
- Increase the intensity level. The sensation should be a comfortable tingling, not painful.
- Check battery status. If the low battery indicator is on, replace the batteries.
- Skin Irritation:
- Ensure the conductivity gel is hypoallergenic.
- Clean the skin thoroughly before application.
- If irritation persists, discontinue use and consult a healthcare professional. Note: Some users with metal allergies (e.g., nickel) have reported skin reactions; the specific metal content of conductors is not always listed.
- Device Not Powering On:
- Check if the batteries are correctly installed.
- Replace batteries if they are old or depleted.
- Difficulty with Battery Replacement:
- The battery holder is designed to be secure. Follow the instructions precisely. If you encounter significant difficulty, contact customer support to avoid damaging the device.
Specifications
| Feature | Detail |
|---|---|
| Model | Reliefband Classic Wristband |
| Dimensions (L x W x H) | 1.5 x 0.5 x 2 inches |
| Item Weight | 3.84 ounces |
| Power Source | 2 Lithium Metal batteries (included) |
| Battery Life | Up to 150 hours of continuous use |
| Intensity Levels | 5 adjustable settings |
| Water Resistance | Splash & Spray Proof |
| Wrist Sizes | Fits 4.5" - 9.25" |
| Special Features | Clinically Proven, Targeted, Fast-Acting, Drug-Free |

Image: Comparison of Reliefband technology with other nausea relief methods.
Warranty and Support
For information regarding product warranty, returns, or technical support, please contact Reliefband customer service directly. As a product from a small business brand, your support is valued.
Visit the official Reliefband Store for more information and contact details: Reliefband Store
Ask a question about this manual
Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.