DLAB 7030301006

DLAB Micropette Plus Single-Channel Adjustable Volume Pipette User Manual

Model: 7030301006

1. Introduction

This user manual provides comprehensive instructions for the safe and effective operation, maintenance, and troubleshooting of the DLAB Micropette Plus Single-Channel Adjustable Volume Pipette, Model 7030301006. This precision instrument is designed for accurate and reliable liquid handling in laboratory settings, specifically for volumes ranging from 5 to 50 microliters (ul).

Please read this manual thoroughly before using the pipette to ensure proper handling and to maximize the lifespan and accuracy of the device.

2. Safety Information

Always adhere to standard laboratory safety procedures when using this pipette. This includes wearing appropriate personal protective equipment (PPE) such as gloves, lab coats, and eye protection.

  • Do not pipette corrosive, highly volatile, or highly viscous liquids without proper precautions and tip selection.
  • Ensure tips are securely attached before pipetting to prevent liquid leakage.
  • Dispose of contaminated tips and liquids according to laboratory waste disposal protocols.
  • Avoid dropping the pipette or subjecting it to strong impacts, as this can affect its calibration and performance.
  • Only use the pipette within its specified volume range (5-50 ul).

3. Product Overview and Components

The DLAB Micropette Plus is a robust and user-friendly single-channel pipette. Key features include its fully autoclavable design and ease of calibration and maintenance.

DLAB Micropette Plus Single-Channel Adjustable Volume Pipette

Figure 1: DLAB Micropette Plus Single-Channel Adjustable Volume Pipette. The image shows the full length of the pipette, from the plunger button at the top to the tip ejector and the tip cone at the bottom. The DLAB logo and volume display are visible on the main body.

Key Components:

  • Plunger Button: Used for aspiration and dispensing of liquids.
  • Volume Adjustment Knob: Rotated to set the desired volume.
  • Volume Display: Shows the currently set volume.
  • Tip Ejector Button: Depressed to eject the used tip.
  • Tip Cone: Where the disposable pipette tip is attached.

4. Setup

Before first use, ensure the pipette is clean and free from any packaging debris.

  1. Unpacking: Carefully remove the pipette from its packaging. Inspect for any visible damage.
  2. Tip Attachment: Firmly press a compatible disposable pipette tip onto the tip cone. Ensure a tight seal to prevent air leaks, which can affect accuracy.
  3. Volume Setting: Rotate the volume adjustment knob to set the desired volume within the 5-50 ul range. Ensure the volume display aligns precisely with the desired setting.

5. Operating Instructions

5.1. Aspiration (Drawing Liquid)

  1. Press the plunger button down to the first stop.
  2. Immerse the tip into the liquid to the appropriate depth (typically 1-2 mm for small volumes).
  3. Slowly release the plunger button to aspirate the liquid. Wait a moment for the liquid to fully enter the tip.
  4. Remove the tip from the liquid, touching it against the edge of the container to remove any excess droplets.

5.2. Dispensing (Releasing Liquid)

  1. Place the tip against the inner wall of the receiving vessel at a slight angle.
  2. Press the plunger button down slowly and smoothly to the first stop to dispense the liquid.
  3. Continue pressing the plunger button to the second stop (blow-out) to ensure complete expulsion of the liquid.
  4. While holding the plunger button at the second stop, withdraw the tip from the vessel.
  5. Release the plunger button.

5.3. Tip Ejection

After dispensing, press the tip ejector button to discard the used tip into an appropriate waste container.

6. Maintenance and Cleaning

Regular maintenance ensures the longevity and accuracy of your DLAB Micropette Plus pipette. This model is fully autoclavable, simplifying sterilization.

6.1. Daily Cleaning

  • Wipe the exterior of the pipette with a soft cloth dampened with 70% ethanol or a mild detergent solution.
  • Do not immerse the entire pipette in liquid unless performing autoclaving as per instructions.

6.2. Autoclaving

The DLAB Micropette Plus is fully autoclavable at 121°C (250°F) for 20 minutes at 1 bar (15 psi) pressure. Ensure the pipette is disassembled according to the manufacturer's guidelines (refer to specific disassembly instructions if provided with the product, or consult DLAB's official resources) before autoclaving. Allow the pipette to cool completely and dry thoroughly before reassembly and use.

6.3. Calibration

The pipette features easy calibration. Regular calibration checks are recommended, especially after extensive use, autoclaving, or if accuracy is suspected to be compromised. Refer to a certified calibration service or DLAB's official calibration procedures for detailed instructions. Calibration typically involves gravimetric testing using distilled water at a controlled temperature.

7. Troubleshooting

ProblemPossible CauseSolution
Inaccurate pipetting / Volume error
  • Loose or incompatible tip
  • Improper pipetting technique (too fast, wrong immersion depth)
  • Pipette out of calibration
  • Internal contamination or damage
  • Ensure tip is securely attached and compatible.
  • Review operating instructions for proper technique.
  • Perform calibration check and recalibrate if necessary.
  • Clean and inspect internal components; contact service if damaged.
Liquid leakage from tip
  • Damaged or incorrect tip
  • Tip cone damage
  • Seal issues within the pipette
  • Replace with a new, correct tip.
  • Inspect tip cone for cracks or deformities.
  • Contact DLAB support for seal replacement or repair.
Plunger sticky or difficult to move
  • Contamination
  • Lack of lubrication
  • Clean the pipette thoroughly.
  • Apply a small amount of silicone grease (specified for pipettes) to the plunger shaft if accessible and recommended by DLAB.

8. Specifications

AttributeDetail
Model Number7030301006
Volume Range5-50 ul (microliters)
TypeSingle-Channel, Adjustable Volume
AutoclavableYes, Fully Autoclavable
ManufacturerDLAB
Package Dimensions11 x 4.15 x 2.57 inches
Item Weight6.4 ounces
Date First AvailableNovember 3, 2016

9. Warranty and Support

For warranty information, technical support, or service inquiries, please contact DLAB customer service or your authorized DLAB distributor. Keep your purchase receipt and product model number (7030301006) handy when contacting support.

Manufacturer: DLAB

ASIN: B01M9K3OPS

© 2023 DLAB. All rights reserved.

Related Documents - 7030301006

Preview Manuel de la glacière thermostatique DLAB CT4-Pro
Manuel d'utilisation et d'entretien pour la glacière thermostatique DLAB CT4-Pro, incluant les précautions de sécurité, les spécifications techniques, la description des fonctions et les informations de certification et de garantie.
Preview DLAB DPCD10 Portable pH/Conductivity/DO Meter - User Guide and Specifications
Comprehensive guide to the DLAB DPCD10 portable meter, covering its features, specifications, and operation for pH, conductivity, and dissolved oxygen measurements. Includes detailed technical data and product overview.
Preview DLAB Stirring & Heating LCD Synchronization Mixer: User Manual and Technical Specifications
Comprehensive user manual for DLAB Stirring & Heating LCD Synchronization Mixers (Models OS-T40-Plus, OS-T60-Plus, OS-T400-Plus), covering safety, operation, maintenance, and technical specifications.
Preview DLAB Laboratory Equipment Product Guide
Explore the comprehensive range of DLAB laboratory instruments, including orbital shakers, bottle-top dispensers, hotplate magnetic stirrers, pipettes, centrifuges, evaporators, and more. Discover high-quality scientific equipment designed for research, analysis, and laboratory applications.
Preview Agitateurs à Suspension DLAB OS-T40-Plus & OS-T60-Plus : Caractéristiques et Spécifications
Présentation détaillée des agitateurs à suspension DLAB OS-T40-Plus et OS-T60-Plus, incluant les caractéristiques, les spécifications techniques et les applications dans la synthèse chimique, l'analyse alimentaire, les produits biopharmaceutiques, et plus encore.
Preview DLAB Digital Stirrers with Heating Mantle - LBD Series Specifications
Detailed specifications for the DLAB LBD series digital stirrers with heating mantles, including models for round-bottom flasks and flat-bottom flasks. Features include precise temperature control, uniform heating, and safety features.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.