REED Instruments TM-8811-NIST

REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge User Manual

Model: TM-8811-NIST

1. Introduction

This manual provides comprehensive instructions for the safe and effective operation, maintenance, and troubleshooting of the REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge. This device is designed for precise measurement of material thickness using ultrasonic technology. Please read this manual thoroughly before using the instrument.

2. Safety Information

Observe the following safety precautions to prevent injury and avoid damage to the instrument:

3. Product Overview

The REED Instruments TM-8811-NIST is an ultrasonic thickness gauge designed for quick and accurate measurements. It features a large LCD display and intuitive controls.

REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge with probes and a calibration certificate.

Figure 1: REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge and a sample calibration certificate.

This image displays the REED Instruments TM-8811-NIST Ultrasonic Thickness Gauge, a blue handheld device with a digital display showing "0.000". Two probes are connected to the top of the device via cables. Below the display are several control buttons: a red power button (circle with a vertical line), "SELECT", "CAL", "VEL", "mm/inch", and up/down arrow buttons. The text "Ultrasonic Thickness Gauge" and "COUPLING" are visible on the device face. Partially visible in the background is a "CALIBRATION CERTIFICATE" document, indicating the importance of calibration for such instruments.

Key Components:

4. Setup

  1. Battery Installation: Open the battery compartment (typically on the back) and insert the required batteries, observing correct polarity. Close the compartment securely.
  2. Probe Connection: Connect the ultrasonic probe(s) to the connectors at the top of the gauge. Ensure a firm connection.
  3. Power On: Press the Power button to turn on the device. The display will illuminate.

5. Operating Instructions

5.1 Basic Measurement

  1. Prepare Surface: Ensure the surface of the material to be measured is clean, smooth, and free of rust, paint, or debris.
  2. Apply Couplant: Apply a small amount of ultrasonic couplant (e.g., glycerin, gel) to the surface where the measurement will be taken. This ensures proper sound transmission.
  3. Place Probe: Press the probe firmly and flatly against the prepared surface. The "COUPLING" indicator on the display should appear, confirming good contact.
  4. Read Measurement: The thickness reading will be displayed on the LCD. If the reading is unstable, adjust the probe position or reapply couplant.
  5. Remove Probe: Lift the probe from the surface. The reading will typically hold or return to zero.

5.2 Unit Selection (mm/inch)

Press the mm/inch button to toggle between millimeters (mm) and inches (inch) for the displayed thickness measurement.

5.3 Sound Velocity Adjustment

The accuracy of ultrasonic thickness measurement depends on the correct sound velocity setting for the material being tested. Refer to a material velocity chart for common materials.

  1. Press the VEL button. The current sound velocity value will be displayed.
  2. Use the Up and Down arrow buttons to adjust the velocity to the desired value for your material.
  3. Press VEL again or SELECT to confirm and exit velocity adjustment mode.

6. Calibration

The TM-8811-NIST gauge is factory calibrated. However, periodic calibration checks and adjustments are crucial for maintaining measurement accuracy, especially when measuring different materials or after prolonged use. The device may feature a built-in calibration block or require external reference standards.

6.1 Zero Calibration

Zero calibration compensates for the probe delay line. This is typically performed by placing the probe on a known thickness standard or by coupling the probe to itself (if applicable to the probe type).

  1. Ensure the probe is clean and free of couplant.
  2. Press the CAL button. Follow the on-screen prompts.
  3. Place the probe on the zero calibration block or perform the specified zeroing procedure.
  4. Confirm the calibration.

6.2 Two-Point Calibration

For enhanced accuracy, especially over a range of thicknesses, a two-point calibration using two known thickness standards (e.g., thin and thick) is recommended.

  1. Press the CAL button and select the two-point calibration option.
  2. Measure the first known thickness standard and input its value using the arrow buttons.
  3. Measure the second known thickness standard and input its value.
  4. Confirm the calibration.

Always refer to the specific instructions in the device's internal menu for detailed calibration procedures.

7. Maintenance

8. Troubleshooting

ProblemPossible CauseSolution
No display/Device won't turn onDead or incorrectly installed batteries.Check battery polarity; replace batteries.
Unstable or "0.000" readingPoor probe coupling; incorrect sound velocity; damaged probe.Reapply couplant, ensure firm probe contact; adjust sound velocity; inspect/replace probe.
"COUPLING" indicator not appearingInsufficient couplant; rough surface; probe not flat.Apply more couplant; prepare surface; ensure probe is flat.
Inaccurate readingsIncorrect sound velocity; calibration required; surface conditions.Verify sound velocity setting; perform calibration; ensure proper surface preparation.

9. Specifications

10. Warranty and Support

REED Instruments products are designed for reliability and performance. For information regarding warranty coverage, technical support, or service, please refer to the official REED Instruments website or contact their customer service department. Keep your purchase receipt as proof of purchase for warranty claims.

Manufacturer: REED Instruments

Date First Available: July 10, 2017

Related Documents - TM-8811-NIST

Preview REED R2160 Thermal Imaging Camera Instruction Manual
Comprehensive instruction manual for the REED R2160 Thermal Imaging Camera, covering features, specifications, operation, safety, and product support.
Preview REED R1610 Thermo-Hygrometer Instruction Manual
This manual provides comprehensive instructions for the REED R1610 Thermo-Hygrometer, a Bluetooth® Smart Series instrument designed for accurate ambient temperature and humidity measurements. It covers features, specifications, setup, usage, care, and compliance information for safe and reliable operation.
Preview REED R1640 Thermocouple Thermometer Instruction Manual
Comprehensive instruction manual for the REED R1640 Thermocouple Thermometer, featuring Bluetooth Smart Series connectivity. Covers product features, specifications, operating procedures, safety guidelines, and support information.
Preview REED R8005 Laser Distance Meter Instruction Manual
Comprehensive instruction manual for the REED R8005 Laser Distance Meter, covering features, specifications, operating instructions, measurement modes, troubleshooting, and warranty information.
Preview REED R3530 Conductivity/TDS/Salinity Meter Instruction Manual
Instruction manual for the REED R3530 Conductivity/TDS/Salinity Meter, covering features, specifications, operation, calibration, maintenance, and product support.
Preview REED SW600 Digital Stopwatch User Manual and Features
Comprehensive guide to the REED SW600 Digital Stopwatch, covering its features, stopwatch functions, alarm and snooze settings, time and date configuration, and operational instructions.