Extech FL700

Extech FL700 Fluoride Meter User Manual

Model: FL700

1. Introduction

The Extech FL700 Fluoride Meter is a precision instrument designed for the rapid and accurate measurement of fluoride ions in various aqueous samples, including drinking water. This device integrates a sensing electrode, measurement electronics, and a digital display into a single, portable unit, making it suitable for both field and laboratory applications.

The FL700 provides direct readings of total fluoride concentrations ranging from 9.99 ppm or mg/L down to 0.1 ppm. It features automatic temperature compensation to ensure measurement accuracy across varying environmental conditions.

2. Safety Information

Please read and understand all safety precautions before operating the Extech FL700 Fluoride Meter. Failure to follow these instructions may result in injury or damage to the instrument.

  • Always handle the fluoride electrode with care. Avoid touching the sensing element directly.
  • When working with TISAB reagent tablets or any chemical solutions, wear appropriate personal protective equipment, including gloves and eye protection.
  • Do not immerse the entire meter in liquid. The meter is designed for electrode immersion only.
  • Store the meter and reagents in a cool, dry place, away from direct sunlight and extreme temperatures.
  • Keep out of reach of children.

3. Product Overview

The Extech FL700 Fluoride Meter is a compact and user-friendly device. Key components and features are outlined below:

Extech FL700 Fluoride Meter front view
Front view of the Extech FL700 Fluoride Meter, showing the display and control buttons.
Hand holding Extech FL700 Fluoride Meter
The Extech FL700 Fluoride Meter held in hand, demonstrating its ergonomic design and clear display.
Close-up of Extech FL700 electrode tip
Close-up view of the fluoride electrode tip, which is integrated into the meter.
Extech FL700 Fluoride Meter in use with sample
The Extech FL700 Fluoride Meter being used to measure a liquid sample, showing the meter in its stand.
Extech FL700 Fluoride Meter and included accessories
All components included with the Extech FL700: the meter, sensor cap, TISAB reagent tablets, and neckstrap.

Key Features:

  • Two-point calibration: Ensures accurate measurements across the fluoride range.
  • Memory storage: Stores up to 25 labeled readings for later review.
  • Field or lab testing: Designed for versatile use in various environments.
  • Fast response: Provides fluoride measurements in less than 1 minute.
  • Large dual LCD: Simultaneously displays ppm (or mg/L) and temperature with a bargraph.

4. Setup

4.1. Battery Installation

The Extech FL700 requires four 3V button batteries (included). To install or replace batteries:

  1. Twist off the battery compartment cap located at the top of the meter.
  2. Insert the four 3V button batteries, ensuring correct polarity as indicated inside the compartment.
  3. Replace the battery compartment cap and twist to secure.

4.2. Initial Preparation

Before first use or after prolonged storage, prepare the electrode:

  1. Remove the sensor cap from the electrode tip.
  2. Rinse the electrode tip with distilled water.
  3. Condition the electrode by immersing it in a fluoride standard solution (e.g., 1 ppm or 10 ppm) for approximately 15-30 minutes.

5. Operating Instructions

5.1. Powering On/Off

Press the ON/OFF button to turn the meter on or off.

5.2. Calibration (Two-Point)

Accurate calibration is crucial for reliable measurements. The FL700 supports two-point calibration.

  1. Prepare two fluoride standard solutions, typically one low concentration (e.g., 1 ppm) and one high concentration (e.g., 10 ppm). Ensure the solutions are at a stable temperature.
  2. Rinse the electrode with distilled water and blot dry.
  3. Immerse the electrode into the first standard solution.
  4. Press and hold the CAL/RECALL button until 'CAL' appears on the display. The meter will automatically recognize the standard and calibrate. Wait for the reading to stabilize.
  5. Rinse the electrode again and immerse it into the second standard solution.
  6. Press and hold the CAL/RECALL button again. The meter will complete the two-point calibration.
  7. The meter is now calibrated and ready for measurement.

5.3. Taking a Measurement

To measure fluoride concentration in a sample:

  1. Ensure the meter is calibrated.
  2. Rinse the electrode with distilled water and blot dry.
  3. Add TISAB (Total Ionic Strength Adjustment Buffer) reagent to your sample according to the reagent instructions. TISAB ensures that the fluoride ions are in a measurable form and minimizes interference.
  4. Immerse the electrode tip into the prepared sample solution. Ensure the tip is fully submerged.
  5. Allow the reading to stabilize on the display. The large dual LCD will show the fluoride concentration in ppm (or mg/L) and the temperature.
  6. To hold the current reading, press the MODE/HOLD button. Press again to release.

5.4. Recalling Stored Readings

The meter can store up to 25 labeled readings.

  • To recall stored readings, press the CAL/RECALL button briefly when not in calibration mode.
  • Use the MODE/HOLD button to scroll through the stored readings.

6. Maintenance

6.1. Electrode Cleaning

Regular cleaning of the electrode ensures optimal performance.

  • After each use, rinse the electrode thoroughly with distilled water.
  • If the electrode becomes fouled or readings become erratic, clean it with a mild detergent solution or a specialized electrode cleaning solution. Rinse thoroughly with distilled water afterward.

6.2. Storage

Proper storage extends the life of the electrode and meter.

  • Always replace the sensor cap on the electrode tip when not in use. The cap should contain a small amount of electrode storage solution or a fluoride standard solution to keep the membrane hydrated. Do not store dry.
  • Store the meter in a cool, dry place.

6.3. Battery Replacement

Replace batteries when the low battery indicator appears on the display or when the meter fails to power on. Refer to Section 4.1 for battery installation instructions.

7. Troubleshooting

ProblemPossible CauseSolution
Meter does not power onDead or incorrectly installed batteriesCheck battery polarity; replace batteries.
Erratic or unstable readingsDirty electrode; air bubbles on electrode; insufficient sample volume; expired or incorrect TISAB.Clean electrode; gently tap electrode to remove bubbles; ensure electrode is fully submerged; check TISAB.
Meter fails to calibrateIncorrect standard solutions; dirty electrode; electrode not conditioned; damaged electrode.Verify standard concentrations; clean and condition electrode; replace electrode if damaged.
Low battery indicatorBatteries are low.Replace all four 3V button batteries.

8. Specifications

SpecificationDetail
ManufacturerExtech
Part NumberFL700
Item Weight7.7 ounces
Product Dimensions6.8 x 1.6 x 1.4 inches
Item Model NumberFL700
Batteries4 Lithium Metal batteries required (included)
StyleFL700
Item Package Quantity1
Number Of Pieces1
Included ComponentsUnit, Instruction Guide
Battery Cell TypeAlkaline
First Available DateJuly 15, 2008

9. Warranty Information

The Extech FL700 Fluoride Meter comes with a full one-year warranty on parts and workmanship, unless otherwise specified. This warranty does not cover defects resulting from user actions such as misuse, improper wiring, operation outside of specified parameters, improper maintenance or repair, or unauthorized modification.

10. Support

For technical support, service, or additional information, please visit the official Extech website or contact their customer service department. You can also find more information and products at the Extech Store on Amazon.

Related Documents - FL700

Preview EXTECH 45170 4-in-1 Humidity, Temperature, Airflow, and Light Meter User Manual
User manual for the EXTECH 45170, a 4-in-1 meter measuring humidity, temperature, airflow, and light. Includes operation, specifications, and warranty information.
Preview Manuale Utente Misuratore di Radiazione UV-AB Extech UV505
Manuale utente completo per il Misuratore di Radiazione UV-AB Extech UV505. Include istruzioni operative, specifiche tecniche, informazioni di sicurezza e manutenzione.
Preview Extech UV505 UV-AB Light Meter User Manual
User manual for the Extech UV505 UV-AB Light Meter. This document provides detailed information on the device's features, operation, specifications, safety precautions, and maintenance for measuring UVA and UVB light.
Preview Extech EX810 1000 Amp Klemmeter met IR Thermometer Gebruikershandleiding
Gedetailleerde gebruikershandleiding voor de Extech EX810 1000 Amp klemmeter met ingebouwde IR thermometer. Bevat instructies, veiligheidsrichtlijnen en technische specificaties voor nauwkeurige metingen van spanning, stroom, weerstand, capaciteit, frequentie en temperatuur.
Preview Extech EX650 Series User Guide: 600A True RMS Digital Clamp Meters
Comprehensive user guide for the Extech EX650 Series of 600A True RMS digital clamp meters, covering models EX650 and EX655. Learn about operation, features, safety, and specifications for accurate AC/DC voltage, current, resistance, and more measurements.
Preview Extech DM100 Pocket Multimeter User's Guide
User's guide for the Extech DM100 4000 Count Autoranging Pocket Multimeter, covering safety, specifications, operation, features, and maintenance.