BEAR™ Devices (BEAR™ 3 Series) — Overview
The BEAR™ 3, BEAR™ 3 plus, BEAR™ 3 go, and BEAR™ 3 eyes & lips devices are intended for facial and neck stimulation for over-the-counter cosmetic use. They use microcurrent spheres and T-Sonic™ pulsations to deliver microcurrent while an app can provide additional intensity options.
Specifications
| Device battery (noted in manual) | BEAR™ 3 & BEAR™ 3 plus: 520 mAh; BEAR™ 3 eyes & lips: 200 mAh; BEAR™ 3 go: 200 mAh |
|---|---|
| Charging time | Up to 1.5 hours (until fully charged) |
| Uses per charge | Up to 90 uses (based on 1-minute treatment time per use) |
| Wireless technology | IEEE 802.15.1 Bluetooth Low Energy (BLE) |
| Wireless modulation | GFSK |
| Operating frequency band | 2.4 GHz to 2.4835 GHz (ISM band) |
| Wireless security | LE Encryption of BLE |
| Wireless functions compliance | Conform to the EMC part of AAMI TIR69 |
| Operating range (phone-to-device) | Less than 1 meter between phone and BEAR™ device |
| Environmental temperature | Operation: 5 to 40 °C; Storage & transport: -10 to 50 °C |
| Environmental humidity | 15 to 93% RH (non-condensing) |
Quick Setup & Use
- Charge: Charge the device using the USB charging cable provided for up to 1.5 hours or until fully charged. During charging, the intensity indicator lights flash; when all lights are lit, it’s fully charged.
- First-time unlock/register (optional but required for first-time use): Download the FOREO For You app, connect via Bluetooth on your smartphone, then press the device universal button to sync.
- Prep skin: Clean and dry your face and neck carefully so there is no residue.
- Apply conductive primer: Apply the primer evenly across the areas you want to treat; massage until fully absorbed.
- Start treatment: Press the universal button to activate microcurrent. Use single quick presses to change intensity levels. Double press the universal button to turn off T-Sonic™ pulsations. (Use the app for a wider range of microcurrent intensities.)
- Glide device: Lightly press both metallic spheres to your skin and gently glide across the treatment area (slow, upward motion, light pressure), keeping both spheres in contact with skin at all times.
- Finish: Press and hold the universal button for 3 seconds to turn off the device, then apply any other skincare products if desired.
BEAR™ Devices
USER MANUAL
BEAR™ 3 BEAR™ 3 plus BEAR™ 3 go BEAR™ 3 eyes & lips
GETTING STARTED: This booklet contains all essential instructions. If you have any further questions regarding the device’s operation, or would like to access additional product resources, visit foreo.com/support.
WARNING: No modification of this equipment is allowed. Please READ ALL INSTRUCTIONS BEFORE USE and utilize this device only for its intended use as described in this manual.
INDICATIONS FOR USE: BEAR™ 3, BEAR™ 3 plus, BEAR™ 3 go and BEAR™ 3 eyes & lips devices are intended for facial and neck stimulation, and are indicated for over-the-counter cosmetic use.
BEAR™ DEVICES – FEATURES
- Microcurrent Spheres
- Silicone
- Power Button / Bluetooth Indicator Light
- Charging Port
- Intensity Indicator Lights
6. Refillable Capsule
Remark: They have different appearances and sizes, and the sizes of their internal batteries are also different.

BEAR™ 3 & BEAR™ 3 plus with 520mAh battery

BEAR™ 3 eyes & lips with 200mAh battery

BEAR™ 3 go with 200mAh battery
- Microcurrent Spheres: Deliver microcurrent directly to skin for a more toned complexion.
- T-Sonic™ Pulsations
- Smart Swedish Design: Gently curved to fit facial contours, and lightweight for skincare on-the-go.
- Anti-Shock System™: Scans skin’s resistance & adjusts microcurrent intensity.
- Bluetooth Indicator Light: Flashes to indicate when device is in Bluetooth pairing mode.
- Universal Power Button: Turns device on and off, and puts device in Bluetooth pairing mode.
- Intensity Indicator Lights: Light up to indicate the microcurrent intensity levels of your device.
- Charging Port: Up to 90 uses per charge (based on a 1-minute treatment time per use).
- Refillable Capsule: Detaches from device, and is designed to easily store your conductive primer.
- Wireless Technology & Functions: IEEE 802.15.1 Bluetooth Low Energy technology (BLE). GFSK Modulation. The bandwidth is 2.4GHz to 2.4835GHz which is in the ISM of the frequency band. The wireless security measure is the LE Encryption of BLE. The wireless functions of BEAR™ devices conform to the EMC part of AAMI TIR69.
CONTRAINDICATIONS FOR USE
Do NOT use BEAR™ devices:
- On the chest / breast area, eye area (circular muscle within the orbital rim), mid-line (bone) of neck, or the genitals / groin area.
- Over red skin, raised moles, major arteries (eg. carotid), enlarged capillaries, metal implants, infected areas, or areas with no feeling in them.
- These devices to treat rosacea, moles, warts, open sores, cancerous lesions, or any skin condition.
- If you have any existing medical conditions such as epilepsy, a hemorrhagic disease, cancer, tumors, or perceptual disturbance.
- If you have any implanted medical device or any other electronic instrument or body aids.
- If you have had a laser treatment, chemical peel, or have any other form of injured or damaged skin.
- If you have had cosmetic surgery on your face.
- On or near children or those with reduced physical and mental abilities.
- The safety of these devices for use during pregnancy has not been established. If you are pregnant, consult your physician before using this device.
- If you have a skin condition or any medical concerns, please consult a physician before use.
- If you have suspected or diagnosed heart disease, follow precautions recommended by your physician.
- These devices should not be used by children or those with reduced physical and mental abilities.
HOW TO USE BEAR™ DEVICES
Before using your device:
- Charge your BEAR™ device using the USB charging cable provided for up to 1.5 hours or until fully charged. A fully charged BEAR™ device should last for up to 90 uses, based on a 1-minute treatment time per use. The intensity indicator lights under the universal power button flash while the device is charging to indicate how much battery your device has. When all of the indicator lights are lit, it means your device is fully charged.
- To unlock and register your device for first time use, download the FOREO For You app. Connect to the app via Bluetooth on your smartphone, and press the universal button on your BEAR™ device to sync your device to the app.
Safety & Security
To ensure the safety and security of your device:
- Ensure the network to which your mobile phone is connected is safe.
- Ensure the BEAR™ device firmware is up to date with the latest version.
- Do not connect the BEAR™ devices to an unofficial application through Bluetooth.
- Ensure optimal Quality of Service (QoS) of wireless function, by making sure your phone supports Bluetooth Low Energy (BLE) with a version above 4.0.
- Ensure that the operating range is less than 1 meter between your phone and your BEAR™ device.
Caution
- BEAR™ devices must be completely dry before starting your microcurrent treatment.
- These devices must be used with a legally marketed conductive primer.
Directions for Use
- Clean and dry your face and neck carefully so there is no remaining residue of any kind.
- Apply the primer to your skin, distributing it evenly across all the areas you wish to treat. Massage until fully absorbed.
- Treating the skin:
- i) Press the universal button to activate microcurrent. The lights underneath the universal button indicate the microcurrent intensity setting. You can change the microcurrent intensity by quick pressing the universal button once for each level. To turn off T-Sonic™ pulsations, double press the universal button. Use the app to access a wider range of microcurrent intensities.
- ii) Lightly press both metallic spheres to your face, and gently glide the device across your cheekbones and forehead, around your lips, and over your jawline and neck as desired. Make sure that both microcurrent spheres are touching your skin at all times. Always glide the device slowly, in an upward motion, and with light pressure for enhanced results.
- iii) Once you have finished your treatment, press and hold the universal button for 3 seconds to turn off device.
- Apply any other skincare products of your choice, if desired.
TREATMENT AREAS

Cheeks
- Hold your device vertically and at a slight angle so that the top sphere is at the corner of your nose, and the bottom sphere is at the corner of your mouth.
- Slowly glide the microcurrent spheres under the natural contour of your cheek, and upward toward your ear.
- Repeat movement evenly across both cheeks.
Forehead
- Hold your device horizontally at the top of your brow, and slowly glide the microcurrent spheres upward toward your hairline.
- Repeat movement evenly across your whole forehead.
Jawline
- Hold your device vertically at the center of your chin, with one sphere above the jawline and one sphere below the jawline.
- Glide the microcurrent spheres slowly along the right side of your jawline toward your right ear.
- Move back to the center of your chin, and glide the microcurrent spheres slowly along the left side of your jawline toward your left ear.
- Repeat both movements evenly.
Neck
- Hold your device horizontally at the base of your neck (slightly off-center), and slowly glide the microcurrent spheres up, ending on the edge of the jawline.
- Repeat movement evenly across your whole neck, always avoiding the mid-line (bone) of the neck.
WARNINGS
- Toning with BEAR™ devices should be comfortable – if you experience any discomfort or irritation, discontinue use immediately and consult a physician.
- Electronic monitoring equipment, such as ECG monitors and ECG alarms, may not operate properly when these devices are in use.
- Close supervision is necessary when these devices are used, cleaned or stored near children or those with reduced physical or mental abilities.
- The long-term effects of microcurrent are unknown.
- Exercise particular care when toning the under-eye regions and do not bring the devices into contact with the eyelids or eyes themselves.
- There may be a perception of flashing lights during the use of your device, resulting from the stimulation of your optic nerve. If you notice this condition constantly without the use of your device, consult your physician.
- A slight tingling sensation may occur while using these devices, which is normal and not cause for any concern. Lowering the intensity may reduce or eliminate this sensation.
- Given the efficiency of the FOREO toning routines, we recommend that you do not use BEAR™ devices for more than 3 minutes at a time.
- For reasons of hygiene, we do not recommend sharing your BEAR™ devices with anyone else.
- Do not use while driving or operating heavy machinery.
- Do not use the devices when they are plugged in.
- Do not insert any object into any opening of the devices.
- Do not use the devices if they are overheating or you suspect that they are malfunctioning.
- Avoid leaving your BEAR™ device in direct sunlight and never expose it to extreme heat or boiling water.
- BEAR™ devices should be completely dry before activating the microcurrent function. Do not use the devices if they have been submerged in water, and do not use with wet hands.
- These devices should only be used with a SELV power adapter.
- It is recommended that IEC60950 standard power suppliers are used to charge the device.
- Before charging, make sure that the plug and socket are completely dry. Failure to do so may result in electric shock, short circuit, or fire.
- Do not use these devices while charging. Discontinue use if the device or charger is not working properly or appears damaged in any way. Use the power cord supplied with your device only.
- These devices contain a non-replaceable battery. The battery must be removed from the appliance before it is disposed of. The appliance must be disconnected from the supply mains when removing the battery, and the battery is to be disposed of safely.
- These products contain no serviceable parts.
- These devices are for cosmetic use on the face and neck. Any harmful consequences resulting from misuse, application to other body areas, connection to improper voltage sources, dirty conductive solution or spheres, or any other improper applications are not the responsibility of FOREO.
MAINTENANCE & SAFETY
CLEANING: Always clean BEAR™ devices thoroughly after use. Wash the metallic spheres and silicone surface with water and soap, then rinse with warm water. Pat dry with a lint-free, non-abrasive cloth or towel.
NOTE: Never use cleaning products containing alcohol, petrol or acetone, as they may irritate the skin and damage the silicone.
Environmental Requirements
| Emission Test | Operation | Storage and Transport |
|---|---|---|
| Temperature | 5 to 40 °C | -10 to 50 °C |
| Humidity | 15 to 93% RH, non-condensing | <=93% RH, non-condensing |
| Pressure | 700 – 1060hPa | 700 – 1060hPa |
Troubleshooting
If BEAR™ devices are not activated when pressing the universal button:
- Battery is empty. Recharge using USB charging cable for up to 1.5 hours until fully charged and then restart your device by holding down the universal button.
If BEAR™ devices cannot be switched off and/or the universal button does not respond:
- Microprocessor is temporarily malfunctioning. Press and hold the universal button to restart the device.
If BEAR™ devices won’t sync to the FOREO For You app:
- Switch your Bluetooth off and then on again to try reconnecting.
- Close the FOREO For You app and then reopen it to start the process over.
- Check to see if the app needs to be updated in your mobile device’s app store.
WARRANTY TERMS & CONDITIONS
REGISTER WARRANTY: To activate your 2-Year Limited Warranty, register through the FOREO For You app, or visit foreo.com/product-registration for more information.
2-YEAR LIMITED WARRANTY: FOREO warrants this device for a period of TWO (2) YEARS after the original date of purchase against defects due to faulty workmanship or materials arising from Normal Use of the device. The warranty covers working parts that affect the function of the device. It does NOT cover cosmetic deterioration caused by fair wear and tear, or damage caused by accident, misuse or neglect. Any attempt to open or take apart the device (or its accessories) will void the warranty. If you discover a defect and notify FOREO during the warranty period, FOREO will, at its discretion, replace the device free of charge. Claims under warranty must be supported by reasonable evidence that the date of the claim is within the warranty period. To validate your warranty, please keep your original purchase receipt together with these warranty conditions for the duration of the warranty period. To claim your warranty, you must log in to your account at www.foreo.com and then select the option to make a warranty claim. Shipping costs are nonrefundable. This undertaking is in addition to your statutory rights as a consumer and does not affect those rights in any way.
DISPOSAL INFORMATION: Battery must be recycled or disposed of properly. Disposal of old electronic equipment (applicable in the EU and other European countries with separate waste collection systems). The crossed-out dustbin symbol indicates that this device should not be treated as household waste, but rather be brought to the appropriate collection point for recycling of electrical and electronic equipment. For more information about the recycling of your device, please contact your local household waste disposal service or your place of purchase. Please see instructions below on how to remove the battery.
SPECIFICATIONS
PRODUCT COMPOSITION: BEAR™ devices are composed of 2 chrome plated zinc alloy metallic spheres, a silicone outer layer, a plastic shell, a PCBA, a motor and a Li-ion battery.
WAVEFORMS SPECIFICATIONS: Pulse durations: 0.34mS, Pulse repetition frequencies: 100PPS/50Hz, Max current: 680uA, Impedance: 100ohm~100Kohm.
Electromagnetic Compatibility (EMC) Information
BEAR™ devices conform to IEC60601-1-2:2014 Electromagnetic Compatibility (EMC) standard. Nevertheless, special precautions need to be observed:
WARNING:
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Portable RF communications device (including peripherals such as antenna cables and external antennas) should be used no closer than 12 inches (30 cm) to any part of the device, including cables specified by FOREO. Otherwise, degradation of the performance of the monitor could result.
- Use of accessories and cables other than those specified or provided by FOREO could result in increased electromagnetic emission or decreased electromagnetic immunity of the equipment and result in improper operation.
- Refer to further guidance below regarding the EMC environment in which the device should be used.
Table 1: EMISSION Limits and Compliance:
| Phenomenon | Emission Limits | Compliance |
|---|---|---|
| Conducted and radiated RF EMISSIONS (CISPR 11) | Group 1, Class B | Compliant |
NOTE: EMISSION tests for Harmonic distortion, Voltage fluctuations and flicker are not applicable.
Table 2: IMMUNITY TEST LEVELS:
| Phenomenon | Basic EMC standard | Immunity Test Levels |
|---|---|---|
| Electrostatic discharge | IEC 61000-4-2 | ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air |
| Radiated RF electromagnetic fields | IEC 61000-4-3 | 10 V/m 80 MHz to 2.7 GHz 80 % AM at 1 kHz |
| Proximity fields from RF wireless communications equipment | IEC 61000-4-3 | See Table 3 |
| Rated power frequency magnetic fields | IEC 61000-4-8 | 30 A/m 50Hz |
NOTE: IMMUNITY tests for Electrical fast transients / bursts, Surges Line-to-line, Surges Line-to-ground, Conducted disturbances induced by RF fields, Voltage dips and Voltage interruptions are not applicable.
Table 3: Test specifications for Proximity fields from RF wireless communication equipment
| Test Frequency (MHz) | Modulation | Immunity Level Applied (V/m) |
|---|---|---|
| 385 | Pulse Modulation: 18 Hz | 27 |
| 450 | Pulse Modulation: 18 Hz | 28 |
| 710, 715, 745 | Pulse Modulation: 217 Hz | 9 |
| 810, 870, 930 | Pulse Modulation: 18 Hz | 28 |
| 1720, 1845, 1970 | Pulse Modulation: 217 Hz | 28 |
| 2450 | Pulse Modulation: 217 Hz | 28 |
| 5240, 5500, 5785 | Pulse Modulation: 217 Hz | 9 |
COMPLIANCE & REGULATORY
DISCLAIMER: Users of these devices do so at their own risk. Neither FOREO nor its retailers assume any responsibility or liability for any injuries or damages, physical or otherwise, resulting, directly or indirectly, from the use of these devices. Furthermore, FOREO reserves the right to revise this publication and to make changes from time to time in the contents thereof without obligation to notify any person of such revision or changes.
CAUTION: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
NOTE: This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
- (1) This device may not cause harmful interference, and
- (2) This device must accept any interference received, including interference that may cause undesired operation.
This device contains licence-exempt transmitter(s)/receiver(s) that comply with Innovation, Science and Economic Development Canada’s licence-exempt RSS(s). Operation is subject to the following two conditions:
- (1) This device may not cause interference.
- (2) This device must accept any interference, including interference that may cause undesired operation of the device.
L’appareil contient des émetteurs/récepteurs exempts de licence qui sont conformes aux CNR exempts de licence d’Innovation, Sciences et Développement économique Canada. L’exploitation est soumise aux deux conditions suivantes :
- (1) l’appareil ne doit pas produire de brouillage,
- (2) l’utilisateur de l’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d’en compromettre le fonctionnement.

Battery removal sequence.
These devices comply with the latest Safety and EMC Standards and Amendments applicable to Medical Devices. These are: IEC 60601-1:2005+A1:2012 +A2:2020, IEC 60601-1-2:2014+AMD1:2020, IEC 60601-2-10:2012+Ams, IEC 60601-1-6:2010 + Ams, IEC 62366-1:2015, IEC 60601-1-11: 2015+ Ams, IEC 62304:2006 + Ams, ANSI/AAMI ES 60601-1:2005 + Ams






©2026 FOREO. ALL RIGHTS RESERVED. MANUFACTURED FOR: FOREO. EU IMP. & RP.: FAQ EUROPE D.O.O., RADNICKA CESTA 202, 10000 ZAGREB, CROATIA. US DISTR.: FOREO INC., 1525 E PAMA LANE, LAS VEGAS, NV 89119, USA. AUS DISTR.: FOREO AUS., 6 PROHASKY STREET, PORT MELBOURNE, VIC, 3207 AUSTRALIA. UK RP: FOREO LTD., LAWFORD HOUSE, ALBERT PLACE, N3 1QA LONDON. MANUFACTURED BY: VIS (SHANGHAI) TECHNOLOGY CO. LTD., NO.58 LINSHENG ROAD, TINGLIN TOWN, JINSHAN DISTRICT, SHANGHAI, CHINA, 201505. ML-P1-20261110
Troubleshooting
- Device won’t charge / charging indicators unclear: Charge using the USB charging cable provided. If the device or charger is not working properly or appears damaged, discontinue use.
- Microcurrent shouldn’t start if the device is wet: Ensure the BEAR™ device is completely dry before starting microcurrent treatment; do not use if submerged or with wet hands.
- Bluetooth pairing not working: Verify your phone supports Bluetooth Low Energy (BLE) version above 4.0, keep the device within less than 1 meter of your phone, and avoid connecting to an unofficial application via Bluetooth.
- Unexpected discomfort or irritation during toning: Discontinue use immediately and consult a physician.
- Sensation of flashing lights or persistent flashing without using the device: If this occurs constantly without using the device, consult a physician.
- Tingling sensation: A slight tingling sensation may occur and is stated as normal; lowering intensity may reduce or eliminate it.
- Intensity/controls behave unexpectedly: Use the universal button as described (single quick press changes intensity level; double press turns off T-Sonic™ pulsations; press-and-hold for 3 seconds turns off).