TekRxTest Application Software
TekRxTest Application User Manual
Register your product at www.tek.com/register.
For product information, sales, service, and technical support, contact Tektronix, Inc. at 13725 SW Karl Braun Drive, P.O. Box 500, Beaverton, OR 97077 USA. In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to find contacts in your area.
Specifications
PCIe Rx Application
Software Compatibility
- Microsoft Visual C++ Redistributable: Required for Sigtest version 6.1.xx (PCIe 6.0 compliance testing). Download from Latest supported Visual C++ Redistributable downloads | Microsoft Learn.
- MATLAB Runtime: R2021a (version 9.10) required for Sigtest versions 5.1.xx and 6.1.xx.
- Sigtest: Refer to the provided link for the latest version and dependencies.
- TekRxTest: Version 6.7.x.x or above.
- TekRxService: 3.0.xx (for communication between TekRxTest and Sigtest/Seasim).
- Sigtest Versions Supported:
- 3.2.xx for PCIe Gen3
- 4.0.x for PCIe Gen4
- Phoenix 5.1.x for PCIe Gen5
- 6.1.xx for PCIe Gen6 (requires latest Microsoft Visual C++ Redistributable)
- Seasim:
- Version 0.77 for PCIe Gen4 and Gen5
- Version 2.0.104d for PCIe Gen6
Hardware Compatibility
- Anritsu MP1900a Firmware and Software:
- MX190000A Signal Quality Analyzer-R Control Software: V10.07.06 or above
- MX183000A High-Speed Serial Data Test Software: V10.07.01 or above
- Oscilloscope:
- DPO77002SX: Version 10.15.0.3 or above
- DPO70000DX, DPO70000SX: Version 10.15.1.3 or above
- PAMJET: Version 10.10.1 or above (Required for PCIe 6.0)
USB Rx Application
Software Compatibility
- SigTest Version: v0.87 only for USB4 Gen 2/3 (Legacy and Rounded) calibration.
Usage Instructions
Installing TekRxTest Application (PCIe Rx)
- Install Prerequisites:
- Microsoft Visual C++ Redistributable: Download and install the latest supported version from the Microsoft website. This is crucial for Sigtest version 6.1.xx (PCIe 6.0).
- MATLAB Runtime: Install MATLAB Runtime Compiler R2021a (version 9.10).
- Sigtest: Download the latest Sigtest version and its dependencies from the provided link.
- Run TekRxTest Installer: Execute the TekRxTest application installer (V6.7.x.x or above).
- Select Configuration Options: During installation, you will be prompted to choose components. Ensure you select the appropriate Sigtest version for your PCIe generation (e.g., Sigtest 6.1.xx for PCIe Gen6). TekRxService and Seasim components are bundled and will install automatically.
- Oscilloscope Setup: Ensure the correct oscilloscope software version is installed on your device (e.g., DPO77002SX: Version 10.15.0.3 or above).
Installing TekRxTest Application (USB Rx)
- Download SigTest v0.87: Obtain the SigTest v0.87 executable for USB4 Gen 2/3 calibration. If needed, contact your Tektronix Account Manager for assistance.
- Place SigTest Executable: Place the SigTest v0.87 executable in a designated folder.
- Configure SigTest Path: Before running USB4v1 calibration, navigate to Settings > Analysis Tool section of USB4v1 Settings within the TekRxTest application and configure the path to the SigTest executable.
Troubleshooting
- MATLAB Runtime Pop-up During Installation: If you see a pop-up prompting to install MATLAB during TekRxTest installation, it indicates that MATLAB Runtime (R2021a, version 9.10) was not detected. Please install it as per the prerequisites. If MATLAB Runtime is already installed, the installation should proceed without interruption.
- PCIe 6.0 Compliance Testing Issues: Ensure the latest Microsoft Visual C++ Redistributable is installed, as it is required for Sigtest version 6.1.xx.
- USB4 Calibration Issues: Verify that SigTest v0.87 is correctly placed in a folder and its path is configured in the USB4v1 Settings within the TekRxTest application.
FAQ
Q: Where can I register my TekRxTest product?
A: You can register your product online at www.tek.com/register.
Q: What are the minimum requirements for PCIe Rx application?
A: You need Microsoft Visual C++ Redistributable, MATLAB Runtime R2021a (version 9.10), and the appropriate Sigtest version for your PCIe generation. Refer to the ‘Specifications’ section for detailed versions.
Q: Do I need to install Seasim separately?
A: No, Seasim components are bundled with the TekRxTest application and will be automatically installed during the wrapper setup.
Q: Which SigTest version is required for USB4 Gen 2/3 calibration?
A: Only SigTest v0.87 is supported for USB4 Gen 2/3 (Legacy and Rounded) calibration.
Q: What should I do if the installer prompts me to install MATLAB?
A: This means MATLAB Runtime R2021a (version 9.10) is not installed. Please install it from the provided download link in the ‘Specifications’ section before proceeding with the TekRxTest installation.
TekRxTest Application
Register now! Click the following link to protect your product. www.tek.com/register
*** Set barcode variable ***
077-1945-00 April 2026
Copyright 2026, Tektronix, Inc. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. These are the original instructions in English. Go to tek.com/eula to read the Tektronix End User License Agreement.
Contacting Tektronix Tektronix, Inc. 13725 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA For product information, sales, service, and technical support: In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to find contacts in your area.
Contents
Prerequisites and installer instructions for PCIe Rx application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Prerequisites and installer instructions for USB Rx application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installation Instructions
WWW.TEK.COM | 3
Prerequisites and installer instructions for PCIe Rx application
Prerequisites
These are the pre-requisites required for installing the TekRxTest application. 1. Microsoft Visual C++ Redistributable
a. Required for use with Sigtest version 6.1.xx, which supports PCIe 6.0 compliance testing. b. Download Link: Latest supported Visual C++ Redistributable downloads | Microsoft Learn 2. MATLAB Runtime a. Install MATLAB Runtime Compiler R2021a (version 9.10). b. This is required for Sigtest versions 5.1.xx and 6.1.xx. c. Download Link: Link 3. Refer to the Link to download latest Sigtest version and its dependencies.
Install TekRxTest application
During installation of the TekRxTest application installer (V6.7.x.x or above), you will be prompted to choose from the following configuration options:
Installation Instructions
WWW.TEK.COM | 4
TekRxTest
Prerequisites and installer instructions for PCIe Rx application
TekRxService 3.0.xx (To communicate between TekRxTest and Sigtest/Seasim) Sigtest 3.2.xx for PCIe Gen3 Sigtest 4.0.x for PCIe Gen4 Sigtest Phoenix 5.1.x for PCIe Gen5 Sigtest 6.1.xx for PCIe Gen6 (Install latest Microsoft Visual C++ Redistributable: refer to pre-requisites (on page 4)) TekRxTest 6.7.x.x (Rx Application)
NOTE: If you did not install the prerequisite MATLAB Runtime (on page 4), the installer will display a pop-up prompting to install MATLAB. If MATLAB Runtime is already installed, the installation will continue without interruption.
Installation Instructions
WWW.TEK.COM | 5
TekRxTest
Prerequisites and installer instructions for PCIe Rx application
Anritsu MP1900a firmware and software compatibility
MX190000A Signal Quality Analyzer-R Control Software: V10.07.06 or above MX183000A High-Speed Serial Data Test Software: V10.07.01 or above.
Seasim
These components are bundled with the TekRxTest application and will be automatically installed during the wrapper setup: Version 0.77 for PCIe Gen4 and Gen5 Version 2.0.104d for PCIe Gen6
Oscilloscope
Ensure the correct version is installed on the oscilloscope before running TekRxTest. DPO77002SX: Version 10.15.0.3 or above DPO70000DX, DPO70000SX: Version 10.15.1.3 or above PAMJET: Version 10.10.1 or above (Required for PCIe 6.0)
Installation Instructions
WWW.TEK.COM | 6
Prerequisites and installer instructions for USB Rx application
Prerequisites
These are the pre-requisites required for installing TekRxTest application. 1. USB4
1. SigTest Version The application supports SigTest v0.87 only for USB4 Gen 2/3 (Legacy and Rounded) calibration. Download the SigTest v0.87 executable or contact Tektronix Account Manager for assistance. Place the SigTest v0.87 executable in a folder as shown in the below image.
Installation Instructions
WWW.TEK.COM | 7
TekRxTest
Prerequisites and installer instructions for USB Rx application
Before running USB4v1 calibration, the path to the SigTest executable must be configured in Settings > Analysis Tool section of USB4v1 Settings.
SigTest configuration for USB4v1 calibration
2. MATLAB Runtime Dependency
Installation Instructions
WWW.TEK.COM | 8
TekRxTest
Prerequisites and installer instructions for USB Rx application
SigTest v0.87 requires MATLAB Runtime Compiler R2021b (9.11). This must be installed on oscilloscope along with SigTest.
Install MATLAB Runtime Compiler R2021b (9.11) Download Link: Link
3. ETT Version Support The application only supports ETT Version 1.1.2 and above for USB4 Gen 2/3 (Legacy and Rounded) tests. ETT should be installed on the oscilloscope which will be used for running receiver tests. Download Link:Link
Installing DMC Drivers for USB4v1 Wilder Microcontroller: 1. DMC drivers must be installed on the oscilloscope to support USB4v1 Wilder Microcontroller. 2. Download the DMC drivers or contact Tektronix account manager for assistance. Place the DMC drivers in a folder (for example: C:USB4ETT)
3. The steps are as follows: a. Connect the USB4v1 wilder microcontroller to 20 V power supply b. Connect the Type A to Type B cable from scope to microcontroller c. Navigate to the Device Manager on scope d. Expand the Other Devices and right click on DMC Device e. Select Update Drivers f. Select Browse my computer for driver software as per the below image.
g. Select Let me pick from a list of available drivers on my computer and click on Next. Installation Instructions
WWW.TEK.COM | 9
TekRxTest
Prerequisites and installer instructions for USB Rx application
h. Select Show All Devices and click on Next.
Installation Instructions
WWW.TEK.COM | 10
TekRxTest
Prerequisites and installer instructions for USB Rx application
i. Select Have Disk…
Installation Instructions
WWW.TEK.COM | 11
TekRxTest
Prerequisites and installer instructions for USB Rx application
j. Click on Browse button in Install From Disk window.
k. Navigate to C:USB4ETTWilder gen 3 uCOntroller Kit Flash drive contentsVendor_DriverbinWin10x64 and select “cyusb3.inf” file.
Installation Instructions
WWW.TEK.COM | 12
TekRxTest
Prerequisites and installer instructions for USB Rx application
l. Select Billboard Bridge Vendor MFG from the list and click on Next.
Installation Instructions
WWW.TEK.COM | 13
TekRxTest
Prerequisites and installer instructions for USB Rx application
m. Click Yes on the warning pop-up.
Steps of installing and configuring ETT v1.1.2: 1. Download ETT v1.1.2 from USB-IF or contact Tektronix account manager for assistance. 2. Create USB4ETT folder in C drive (for example: C:USB4ETT) 3. Copy the following executables in the folder above:
USB4ElectricalTestTool.exe USB4ElectricalTestToolCLI.exe Wilder_USB4_Controller_GUI.exe
Installation Instructions
WWW.TEK.COM | 14
TekRxTest
Prerequisites and installer instructions for USB Rx application
4. Go to Environment Variables page in Windows OS. 5. Create a new variable in User variables section as follows:
Variable name: Wilder_GUI_path Variable value: Set the full path including the file name of Wilder_USB4_Controller_GUI.exe
6. Click on OK and apply the changes in Environment Variables page. Configuring ETT path in TekRxTest: While performing USB4v1 receiver testing using TekRxTest, configure the folder path as shown below:
Installation Instructions
WWW.TEK.COM | 15
TekRxTest
Prerequisites and installer instructions for USB Rx application
2. USB4v2
1. SigTest Version The application supports SigTest v0.99 only for USB4v2 Calibration. Download Link: Link Download the SigTest v0.99 executable or contact Tektronix Account Manager for assistance.
Installation Instructions
WWW.TEK.COM | 16
TekRxTest
Prerequisites and installer instructions for USB Rx application
Before running USB4v2 calibration, the path to SigTest v0.99 executable must be configured in Analysis Tool section of USB4v2 Settings.
SigTest configuration for USB4v2 calibration
2. MATLAB Runtime Dependency
Installation Instructions
WWW.TEK.COM | 17
TekRxTest
Prerequisites and installer instructions for USB Rx application
SigTest v0.99 requires MATLAB Runtime Compiler R2024b (24.2). MCR and SigTest v0.99 must be installed on an external computer that adheres to minimum Windows OS
requirement. Install MATLAB Runtime Compiler R2024b (24.2) Download Link: Link
NOTE: MATLAB Runtime R2024b (24.2) requires Windows 10 (version 22H2 or above). It is recommended to use an external computer running Windows 10 (version 22H2 or above) to install both MATLAB Runtime and SigTest v0.99.
3. Installation on External Computer Install MATLAB Runtime, SigTest v0.99, TekRxTest and TekRxService application on an external computer with Windows version 22H2 or higher. In case of USB4v2, the TekRxTest application must be run on external computer only, and not on oscilloscope.
4. ETT Version Support The application only supports ETT Version 2.0.0 and above for USB4v2 tests. Contact Tektronix Account Manager for assistance to get the required ETT version.
NOTE: Additional DMC driver installation is not required for Wilder USB4v2 microcontroller.
Steps of installing and configuring ETT v2.0.0: a. Create a folder in C drive (for example: C:USB4ETT) b. Copy the following executables in the folder
USB4ElectricalTestTool.exe USB4ElectricalTestToolCLI.exe EV-controller-WinApp_v_1_1.exe
Installation Instructions
WWW.TEK.COM | 18
TekRxTest
Prerequisites and installer instructions for USB Rx application
c. Go to Environment Variables page in Windows OS. d. Create a two new variable in User variables section as follows:
Variable name: Gen4_GUI_path Variable value: Set the full path including the file name of EV-controller-WinApp_v_1_1.exe.
e. Variable name: ControllerSelect_Wilder_Gen4 f. Variable value: Gen4
g. Click on OK and apply the changes in Environment Variables page. Configuring ETT path in TekRxTest: While performing USB4v2 receiver testing using TekRxTest, configure the folder path as shown below:
Installation Instructions
WWW.TEK.COM | 19
TekRxTest
Prerequisites and installer instructions for USB Rx application
Install TekRxTest Application
During installation of the TekRxTest Application installer (V6.7.x.x or above), you will be prompted to choose from the following configuration options:
Installation Instructions
WWW.TEK.COM | 20
TekRxTest
Prerequisites and installer instructions for USB Rx application
TekRxService 3.0.0.xx (To communicate between TekRxTest and Sigtest/Seasim) TekRxTest 6.7.x.x (Rx Application)
NOTE: Please use an external computer with Windows 10 (version 22H2 or above) for running the TekRxTest application. This is required for running USB4v2 SigTest.
TekRxTest_Application installation instructions in this case are as follows: Install the following application on external computer: TekRxService 3.0.0.xx (To communicate between TekRxTest and Sigtest) TekRxTest 6.7.x.x (Rx Application) Install the following application on Real Time Oscilloscope: TekRxService 3.0.0.xx (To communicate between TekRxTest and Sigtest)
Configuring SigTest in TekRxTest application
After installing the TekRxTest application, configure the SigTest executable path in the Analysis Tool section under the Settings tab for the USB4 Gen2/Gen3 or USB4v2 plugin.
Installation Instructions
WWW.TEK.COM | 21
TekRxTest
Prerequisites and installer instructions for USB Rx application
NOTE: Refer to the Anritsu MP1900a firmware and software compatibility (on page 4) section for supported Anritsu models, and the Oscilloscope (on page 4) section for supported Tektronix oscilloscopes.
Installation Instructions
WWW.TEK.COM | 22
Documents / Resources
![]() | TekRxTest Application Software |
References
- in.mathworks.com/products/compiler/matlab-runtime.htmlin.mathworks.com
- learn.microsoft.com/en-us/cpp/windows/latest-supported-vc-redist?view=msvc-170learn.microsoft.com
- END USER LICENSE AGREEMENT | Tektronixtek.com
- intel.com/content/www/us/en/collections/technology/high-speed-io-tools.htmlwww.intel.com
- Test and Measurement Equipment | Tektronixwww.tek.com
- Welcome to Product Registration | Tektronixwww.tek.com
- Test and Measurement Equipment | Tektronixwww.tek.com
- END USER LICENSE AGREEMENT | Tektronixwww.tek.com
- tek.com/registerwww.tek.com
- usb.org/document-library/usb4-sigtestwww.usb.org
- User Manualmanual.tools